Gamma-ray efficiency of a HPGe detector as a function of energy and geometry

被引:17
作者
Challan, Mohsen B. [1 ,2 ]
机构
[1] AEA, Nucl Res Ctr, Dept Nucl Phys, Cairo 13759, Egypt
[2] Majmaah Univ, Fac Educ Zulfi, Dept Phys, Al Zulfi 159422588, Saudi Arabia
关键词
HPGe detector; Efficiency; Gamma-ray sources; Spectrometry; VIRTUAL POINT-DETECTOR; SPECTRUM ANALYSIS; PEAK EFFICIENCY; CALIBRATION; DEPENDENCE;
D O I
10.1016/j.apradiso.2013.07.010
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The concept of double points detector model approach (DPDM) is developed as a procedure to find the full energy peak efficiency of the coaxial 120 cm(3) closed hyperpure germanium (HPGe) detectors. Usually in the experimental nuclear physics work, which involves using HPGe detector for gamma-ray spectrometry, the full energy peak efficiency function must represent adequately the HPGe detector response. In the current work the gamma-ray energy in the range from 60 key to 1332 keV and gamma-ray intensity changes by changing source to detector distance from 10 mm to 800 mm. The detector was characterized using a number of point-like standard sources. The calculated efficiencies obtained by (DPDM) are in good agreement with experimental data. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:166 / 169
页数:4
相关论文
共 24 条
  • [1] Full energy peak efficiency calibration of HPGe detector for point and extended sources using Monte Carlo code
    Agarwal, Chhavi
    Chaudhury, Sanhita
    Goswami, A.
    Gathibandhe, M.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2011, 287 (03) : 701 - 708
  • [2] HPGe virtual point detector for radioactive disk sources
    Alfassi, Z. B.
    Lavi, N.
    Presler, O.
    Pushkarski, V.
    [J]. APPLIED RADIATION AND ISOTOPES, 2007, 65 (02) : 253 - 258
  • [3] New analytical approach to calibrate the co-axial HPGe detectors including correction for source matrix self-attenuation
    Badawi, Mohamed S.
    Gouda, Mona M.
    Nafee, Sherif S.
    El-Khatib, Ahmed M.
    El-Mallah, Ekram A.
    [J]. APPLIED RADIATION AND ISOTOPES, 2012, 70 (12) : 2661 - 2668
  • [4] An investigation of HPGe gamma efficiency calibration software (ANGLE V.3) for applications in nuclear decommissioning
    Bell, S. J.
    Judge, S. M.
    Regan, P. H.
    [J]. APPLIED RADIATION AND ISOTOPES, 2012, 70 (12) : 2737 - 2741
  • [5] Bevington PhilipR., 2010, DATA REDUCTION ERROR
  • [6] Determination of the dependence of HPGe virtual point detector location on source dimensions in 60 keV-2MeV range using Monte Carlo simulation
    Celik, N.
    [J]. JOURNAL OF INSTRUMENTATION, 2012, 7
  • [7] Challan M., 2007, Nucl. Rad. Phys, V2, P1
  • [8] Debertin K., 2001, Gamma and X-Ray Spectrometry with Semiconductor Detectors
  • [9] DEPENDENCE OF GE(LI) EFFICIENCY SLOPE ON SOURCE-TO-DETECTOR DISTANCE
    GRANT, PM
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1975, 127 (03): : 371 - 372
  • [10] ANALYTICAL PEAK FITTING FOR GAMMA-RAY SPECTRUM ANALYSIS WITH GE DETECTORS
    HAMMED, MA
    GRAY, PW
    NABOULSI, AH
    MACMAHON, TD
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 334 (2-3) : 543 - 550