Coronary Plaque Quantification by Voxel Analysis: Dual-Source MDCT Angiography Versus Intravascular Sonography

被引:32
作者
Brodoefel, Harald [1 ,2 ]
Burgstahler, Christof [3 ]
Sabir, Adeel [1 ]
Yam, Chun-Shan [1 ]
Khosa, Faisal [1 ]
Claussen, Claus D. [2 ]
Clouse, Melvin E. [1 ]
机构
[1] Harvard Univ, Dept Radiol, Beth Israel Deaconess Med Ctr, Sch Med, Boston, MA 02215 USA
[2] Univ Tubingen, Dept Diagnost Radiol, Tubingen, Germany
[3] Univ Tubingen, Dept Cardiol, Tubingen, Germany
关键词
dual-source CT; intravascular sonography; MDCT angiography; plaque burden; voxel analysis; MULTIDETECTOR COMPUTED-TOMOGRAPHY; ATHEROSCLEROTIC PLAQUE; ULTRASOUND; PATHOGENESIS; MORPHOLOGY; ACCURACY; SEGMENTS; LUMEN;
D O I
10.2214/AJR.08.1381
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate a voxel-based analytic technique for quantification of noncalcified coronary artery plaque with intravascular sonography as a standard of reference. SUBJECTS AND METHODS. Intravascular sonography and dual-source MDCT angiography prospectively performed on 12 patients resulted in identification of 20 segments containing noncalcified plaque. Four of these segments were used to establish reference measurements of 0.6-mm proximal wall thickness with a 0-HU cutoff between the epicardial fat and outer wall and an individually adjusted threshold for the interface between the wall and lumen. With these data, consecutive circular layers of the outer wall were subtracted from a 3D volume to determine the plaque plus medial layer and the actual plaque volume in the other 16 segments. Accuracy of the voxel technique was assessed by comparing the results with intravascular sonographic findings. RESULTS. Both the total plaque burden (plaque plus medial layer) and the actual plaque volume had good concordance with intravascular sonographic findings (49.6 +/- 20 mm(3) vs 56.7 +/- 23.6 mm(3), p = 0.076; 26.5 +/- 14.8 mm(3) vs 30.9 +/- 15.3 mm(3), p = 0.09). Corresponding correlation coefficients were r = 0.76 and r = 0.79. The method had good reproducibility, the an intraclass correlation coefficients being 0.93 for total plaque burden and 0.90 for actual plaque volume. CONCLUSION. Voxel analysis can be used for accurate and reproducible quantification not only of plaque burden but also of actual plaque volume.
引用
收藏
页码:W84 / W89
页数:6
相关论文
共 28 条
  • [1] Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography - A segment-based comparison with intravascular ultrasound
    Achenbach, S
    Moselewski, F
    Ropers, D
    Ferencik, M
    Hoffmann, U
    MacNeill, B
    Pohle, K
    Baum, U
    Anders, K
    Jang, I
    Daniel, WG
    Brady, TJ
    [J]. CIRCULATION, 2004, 109 (01) : 14 - 17
  • [2] Characterization of coronary atherosclerotic plaques by multidetector computed tomography
    Carrascosa, PM
    Capuñay, CM
    Garcia-Merletti, P
    Carrascosa, J
    Garcia, MJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (05) : 598 - 602
  • [3] Measuring noncalcified coronary atherosclerotic plaque using voxel analysis with MDCT angiography: A pilot clinical study
    Clouse, Melvin E.
    Sabir, Adeel
    Yam, Chun-Shan
    Yoshimura, Norihiko
    Lin, Shezhang
    Welty, Francine
    Martinez-Clark, Pedro
    Raptopoulos, Vassilios
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (06) : 1553 - 1560
  • [4] Pathogenesis of atherosclerosis
    Falk, E
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (08) : C7 - C12
  • [5] Computed tomography and magnetic resonance imaging for noninvasive coronary angiography and plaque imaging - Current and potential future concepts
    Fayad, ZA
    Fuster, V
    Nikolaou, K
    Becker, C
    [J]. CIRCULATION, 2002, 106 (15) : 2026 - 2034
  • [6] Fayad ZA, 2000, CIRCULATION, V102, P506
  • [7] FUSTER V, 1992, NEW ENGL J MED, V326, P310
  • [8] Early characterization of atherosclerotic coronary plaques with multidetector computed tomography in patients with acute coronary syndrome
    Iriart, Xavier
    Brunot, Sebastien
    Coste, Pierre
    Montaudon, Michel
    Dos-Santos, Pierre
    Leroux, Lionel
    Labeque, Jean-Noel
    Jais, Catherine
    Laurent, Francois
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (10) : 2581 - 2588
  • [9] Accuracy of 64-slice computed tomography to classify and quantify plaque volumes in the proximal coronary system - A comparative study using intravascular ultrasound
    Leber, AW
    Becker, A
    Knez, A
    von Ziegler, F
    Sirol, M
    Nikolaou, K
    Ohnesorge, B
    Fayad, ZA
    Becker, CR
    Reiser, M
    Steinbeck, G
    Boekstegers, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (03) : 672 - 677
  • [10] Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques - A comparative study with intracoronary ultrasound
    Leber, AW
    Knez, A
    Becker, A
    Becker, C
    von Ziegler, F
    Nikolaou, K
    Rist, C
    Reiser, M
    White, C
    Steinbeck, G
    Boekstegers, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) : 1241 - 1247