Intensity-modulated tangential beam irradiation of the intact breast

被引:224
作者
Hong, L
Hunt, M
Chui, C
Spirou, S
Forster, K
Lee, H
Yahalom, J
Kutcher, GJ
McCormick, B
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10021 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 44卷 / 05期
关键词
breast cancer; radiation therapy; treatment planning; inverse planning; intensity modulation; tangential fields;
D O I
10.1016/S0360-3016(99)00132-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the potential benefits of intensity modulated tangential beams in the irradiation of the intact breast. Methods and Materials: Three-dimensional treatment planning was performed on five left and five right breasts using standard wedged and intensity modulated (IRI) tangential beams. Optimal beam parameters mere chosen using beams-eye-view display. For the standard plans, the optimal wedge angles were chosen based on dose distributions in the central plane calculated without inhomogeneity corrections, according to our standard protocol. Intensity-modulated plans were generated using an inverse planning algorithm and a standard set of target and critical structure optimization criteria. Plans were compared using multiple dose distributions and dose volume histograms for the planning target volume (PTV), ipsilateral lung, coronary arteries, and contralateral breast. Results: Significant improvements in the doses to critical structures were achieved using intensity modulation. Compared with a standard-wedged plan prescribed to 46 Gy, the dose from the IM plan encompassing 20% of the coronary artery region decreased by 25% (from 36 to 27 GS) for patients treated to the left breast; the mean dose to the contralateral breast decreased by 32% (from 1.2 to 0.7 Gy); the ipsilateral lung volume receiving more than 36 Gy decreased by 30% (from 10% to 7%); the volume of surrounding soft tissue receiving more than 46 Gy decreased by 31% (from 48% to 33%). Dose homogeneity within the target volume improved greatest in the superior and inferior regions of the breast (approximately 8%), although some decrease in the medial and lateral high-dose regions (approximately 4%) was also observed. Conclusion: Intensity modulation with a standard tangential beam arrangement significantly reduces the dose to the coronary arteries, ipsilateral lung, contralateral breast, and surrounding soft tissues. Improvements in dose homogeneity throughout the target volume can also be achieved, particularly in the superior and inferior regions of the breast. It remains to be seen whether the dosimetric improvements achievable with IR-IRT will lead to significant clinical outcome improvements. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:1155 / 1164
页数:10
相关论文
共 45 条
  • [1] [Anonymous], 1993, ICRU report 50, DOI 10.1118/1.597396
  • [2] QUALITY ASSURANCE IN CONSERVATIVE TREATMENT OF EARLY BREAST-CANCER - REPORT ON A CONSENSUS MEETING OF THE EORTC RADIOTHERAPY AND BREAST-CANCER COOPERATIVE GROUPS AND THE EUSOMA (EUROPEAN-SOCIETY-OF-MASTOLOGY)
    BARTELINK, H
    GARAVAGLIA, G
    JOHANSSON, KA
    MIJNHEER, BJ
    VANDENBOGAERT, W
    VANTIENHOVEN, G
    YARNOLD, J
    [J]. RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) : 323 - 326
  • [3] CANCER IN THE CONTRALATERAL BREAST AFTER RADIOTHERAPY FOR BREAST-CANCER
    BOICE, JD
    HARVEY, EB
    BLETTNER, M
    STOVALL, M
    FLANNERY, JT
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (12) : 781 - 785
  • [4] Dosimetric analysis of intact breast irradiation in off-axis planes
    Buchholz, TA
    Gurgoze, E
    Bice, WS
    Prestidge, BR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (01): : 261 - 267
  • [5] THE EFFECT OF THE NUMBER OF COMPUTED TOMOGRAPHIC SLICES ON DOSE DISTRIBUTIONS AND EVALUATION OF TREATMENT PLANNING SYSTEMS FOR RADIATION-THERAPY OF INTACT BREAST
    CHENG, CW
    DAS, IJ
    STEA, B
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (01): : 183 - 195
  • [6] 3-DIMENSIONAL PHOTON DOSE DISTRIBUTIONS WITH AND WITHOUT LUNG CORRECTIONS FOR TANGENTIAL BREAST INTACT TREATMENTS
    CHIN, LM
    CHENG, CW
    SIDDON, RL
    RICE, RK
    MIJNHEER, BJ
    HARRIS, JR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (06): : 1327 - 1335
  • [7] DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS
    CHUI, CS
    LOSASSO, T
    SPIROU, S
    [J]. MEDICAL PHYSICS, 1994, 21 (08) : 1237 - 1244
  • [8] CARCINOGENIC EFFECTS OF SCATTERED DOSE ASSOCIATED WITH RADIATION-THERAPY
    CUMBERLIN, RL
    DRITSCHILO, A
    MOSSMAN, KL
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (03): : 623 - 629
  • [9] TOLERANCES IN SETUP AND DOSIMETRIC ERRORS IN THE RADIATION TREATMENT OF BREAST-CANCER
    DAS, IJ
    CHENG, CW
    FOSMIRE, H
    KASE, KR
    FITZGERALD, TJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (05): : 883 - 890
  • [10] Active minimisation of radiation scatter during breast radiotherapy: Management implications for young patients with good-prognosis primary neoplasms
    Epstein, RJ
    Kelly, SA
    Cook, M
    Bateman, A
    Paddick, I
    Kam, KC
    Dunn, P
    Hanham, IW
    Dale, RG
    Price, PM
    [J]. RADIOTHERAPY AND ONCOLOGY, 1996, 40 (01) : 69 - 74