Note: Laser frequency shifting by using two novel triple-pass acousto-optic modulator configurations

被引:4
作者
de Carlos-Lopez, E. [1 ]
Lopez, J. M. [1 ]
Lopez, S. [1 ]
Espinosa, M. G. [1 ]
Lizama, L. A. [1 ]
机构
[1] Ctr Nacl Metrol, Div Tiempo & Frecuencia, El Marques 76246, Queretaro, Mexico
关键词
STANDARD;
D O I
10.1063/1.4758998
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report the design of two novel triple-pass acousto-optic modulator systems. These designs are extensions of the well known acousto-optic modulator (AOM) double-pass configuration, which eliminates the angle dependence of the diffracted beam with respect to the modulation frequency. In a triple-pass system, however, the frequency dependence of the angle does not disappear but the frequency shift is larger, spanning 3 times the AOM central frequency. In some applications, such as optically pumped Cesium-beam frequency standards, the frequencies of the two laser beams remain fixed and a triple-pass optical system can be used to reduce to one the number of lasers used in such atomic clocks. The two triple-pass configurations use either a retro-reflecting mirror, or a right angle prism to pass for third time the laser beam through the AOM, obtaining diffraction efficiencies of about 27% and 44%, respectively. (C) 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4758998]
引用
收藏
页数:3
相关论文
共 8 条
[1]   Double-pass acousto-optic modulator system [J].
Donley, EA ;
Heavner, TP ;
Levi, F ;
Tataw, MO ;
Jefferts, SR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (06)
[2]   ERROR ANALYSIS OF THE NIST OPTICALLY PUMPED PRIMARY FREQUENCY STANDARD [J].
DRULLINGER, RE ;
SHIRLEY, JH ;
LOWE, JP ;
GLAZE, DJ .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (02) :453-456
[3]  
DRULLINGER RE, 1990, PROCEEDINGS OF THE 44TH ANNUAL SYMPOSIUM ON FREQUENCY CONTROL 1990, P76, DOI 10.1109/FREQ.1990.177483
[4]   ACCURATE MEASUREMENT OF TIME [J].
ITANO, WM ;
RAMSEY, NF .
SCIENTIFIC AMERICAN, 1993, 269 (01) :56-&
[5]  
López ED, 2004, REV MEX FIS, V50, P569
[6]  
Lopez-Romero J. M., 1998, P 12 EUR FREQ TIM FO, P557
[7]   Evaluation of the accuracy of the optically pumped caesium beam primary frequency standard of the BNM-LPTF [J].
Makdissi, A ;
de Clercq, E .
METROLOGIA, 2001, 38 (05) :409-425
[8]   Atomic fountain clocks [J].
Wynands, R ;
Weyers, S .
METROLOGIA, 2005, 42 (03) :S64-S79