Solid phase extraction for multiresidue analysis of anabolic steroids and related substances from calf urine using C18 and alumina columns

被引:6
|
作者
Koole, A [1 ]
Franke, JP [1 ]
de Zeeuw, RA [1 ]
机构
[1] Univ Groningen, Ctr Pharm, Dept Analyt Chem & Toxicol, NL-9713 AV Groningen, Netherlands
关键词
D O I
10.1081/JLC-100102048
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A solid phase extraction method for anabolic steroids and related substances in calf urine is reported, that is suitable as a screening method for illegal growth promoters. Two types of sorbent were used: a reversed phase C18 material and a polar alumina material. After overnight enzymatic deconjugation, the 5-mL sample was first brought on the C18 column. This column was washed with 55% methanol in water and was then eluted with 95% acetone in water. The extract was directly brought on the alumina column. The run-through and an additional elution with 95% acetone in water were collected. The final extracts were analysed with an HPLC-DAD method described previously. The steroids are separated on a Superspher RP-Select B column with a gradient mobile phase consisting of acetonitrile and water. The method was found to be suitable for at least 19 illegally used anabolic steroids, having recoveries ranging from 65-110% at a spiking level of 25 ng/mL. Detection limits ranged from 0.6-80 ng injected amount or 1-160 ng/mL urine.
引用
收藏
页码:2627 / 2650