Elastomeric Pump Reliability in Postoperative Regional Anesthesia: A Survey of 430 Consecutive Devices

被引:37
作者
Remerand, Francis [1 ]
Vuitton, Anne Sophie [1 ]
Palud, Michel [1 ]
Buchet, Sylvie [1 ]
Pourrat, Xavier [1 ]
Baud, Annick [1 ]
Laffon, Marc [1 ]
Fusciardi, Jacques [1 ]
机构
[1] Univ Tours, CHU Tours, Hop Trousseau, Groupement Anesthesie Reanimat, F-37041 Tours 1, France
关键词
D O I
10.1213/ane.0b013e318187c9bb
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Postoperative analgesia via continuous perineural infusion of local anesthetics compares favorably with systemic analgesia. Elastomeric pumps increase patient satisfaction compared with electronic models. In in vitro investigations, infusions remained within 15% of their designated set rates. We assessed in vivo the infusion rate of elastomeric pumps in regional analgesia after orthopedic surgery. METHODS: All consecutive elastomeric pumps were retrospectively studied during a 10-mo period. perineural catheters were inserted preoperatively and connected postoperatively to elastomeric pumps filled with ropivacaine 0.2%. Before infusion, elastomeric pumps and ropivacaine were stored at room temperature. Two models of pumps were randomly used: Infusor (TM) LV5 (Baxter, France) or Easypump (TM) (Braun, Germany), both set at 5 mL/h. Nurses weighed the devices at the bedside using a portable electronic scale several times a day until catheter removal. Weights over time allowed accurate deflation profile assessment and flow rate calculation. An unchanged weight over time indicated either an obstructed catheter or an ineffective device. RESULTS: After connection to the catheter, 88 devices did not deflate (80 Easypump of 300 and 8 Infusor of 130, P < 0.0001). One Easypump was impossible to deflate, even after disconnection from its catheter. In two cases, catheters were obstructed. In 21 cases, catheters were removed 11 to 72 h later without being tested for patency. In 24 cases, pumps correctly deflated after catheters were injected without difficulty with a local anesthetic bolus. The remaining 40 devices spontaneously started to deflate 6 to 43 h after their connection. These 88 elastomeric Pumps were associated with higher maximal Visual analog scale scores during the first postoperative night than devices showing immediate deflation after connection (34 +/- 21 mm vs 26 +/- 19 mm, P = 0.006). Flow rates were calculated over a mean period of 54 +/- 18 h (Easypump) and 49 +/- 19 h (Infusor). The flow rates differed from those set by manufacturers (5 mL/h +/- 15%) in 47% of Easypump and in 34% of Infusor devices (P = 0.01). CONCLUSIONS: In vivo reliability of elastomeric pumps is different than in vitro. In the event of early insufficient postoperative perineural analgesia, an absence of deflation of the elastomeric pump must be considered. We recommend weighing these devices every 3 h during the first 24 h of infusion. (Anesth Analg 2008;107:2079-84)
引用
收藏
页码:2079 / 2084
页数:6
相关论文
共 18 条
[1]  
Ackermann Monique, 2007, J Oncol Pharm Pract, V13, P77, DOI 10.1177/1078155207078349
[2]   Patient-controlled perineural analgesia after ambulatory orthopedic surgery: A comparison of electronic versus elastomeric pumps [J].
Capdevila, X ;
Macaire, P ;
Aknin, P ;
Dadure, C ;
Bernard, N ;
Lopez, S .
ANESTHESIA AND ANALGESIA, 2003, 96 (02) :414-417
[3]   Effect of patient-controlled perineural analgesia on rehabilitation and pain after ambulatory orthopedic surgery - A multicenter randomized trial [J].
Capdevila, Xavier ;
Dadure, Christophe ;
Bringuier, Sophie ;
Bernard, Nathalie ;
Biboulet, Philippe ;
Gaertner, Elisabeth ;
Macaire, Philippe .
ANESTHESIOLOGY, 2006, 105 (03) :566-573
[4]   The flow rate of the elastomeric balloon infusor is influenced by the internal pressure of the infusor [J].
Chung, IS ;
Cho, HS ;
Kim, JA ;
Lee, KH .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2001, 16 (06) :702-706
[5]   Continuous psoas compartment blocks after major orthopedic surgery in children: A prospective computed tomographic scan and clinical studies [J].
Dadure, C ;
Raux, O ;
Gaudard, P ;
Sagintaah, M ;
Troncin, R ;
Rochette, A ;
Capdevila, X .
ANESTHESIA AND ANALGESIA, 2004, 98 (03) :623-628
[6]   Elastomeric pumps for ambulatory patient controlled regional analgesia. [J].
Ganapathy, S ;
Amendola, A ;
Lichfield, R ;
Fowler, PJ ;
Ling, E .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2000, 47 (09) :897-902
[7]   Portable infusion pumps used for continuous regional analgesia: Delivery rate accuracy and consistency [J].
Ilfeld, BM ;
Morey, TE ;
Enneking, FK .
REGIONAL ANESTHESIA AND PAIN MEDICINE, 2003, 28 (05) :424-432
[8]   Delivery rate accuracy of portable, bolus-capable infusion pumps used for patient-controlled continuous regional analgesia [J].
Ilfeld, BM ;
Morey, TE ;
Enneking, FK .
REGIONAL ANESTHESIA AND PAIN MEDICINE, 2003, 28 (01) :17-23
[9]   The delivery rate accuracy of portable infusion pumps used for continuous regional analgesia [J].
Ilfeld, BM ;
Morey, TE ;
Enneking, FK .
ANESTHESIA AND ANALGESIA, 2002, 95 (05) :1331-1336
[10]   PROLONGED INFUSION TIMES WITH DISPOSABLE ELASTOMERIC INFUSION DEVICES [J].
KAYE, T .
AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1994, 51 (04) :533-534