Development of a screen-printed carbon electrode based disposable enzyme sensor strip for the measurement of glycated albumin

被引:28
作者
Hatada, Mika [1 ]
Tsugawa, Wakako [1 ,2 ]
Kamio, Eri [1 ]
Loew, Noya [2 ]
Klonoff, David C. [2 ,3 ]
Sode, Koji [1 ,2 ]
机构
[1] Tokyo Univ Agr & Technol, Dept Biotechnol & Life Sci, Grad Sch Engn, 2-24-16 Naka Cho, Koganei, Tokyo 1848588, Japan
[2] Tokyo Univ Agr & Technol, Inst Global Innovat Res, 2-24-16 Naka Cho, Koganei, Tokyo 1848588, Japan
[3] Mills Peninsula Hlth Serv, Diabet Res Inst, 100 South San Mateo Dr, San Mateo, CA 94401 USA
关键词
Diabetes mellitus; Glycated albumin; Fructosyl amino acid oxidase; Hemoglobin A1c; Point-of-care testing; Screen-printed carbon electrode; FRUCTOSYL-AMINE OXIDASE; HUMAN-SERUM-ALBUMIN; ELECTROCHEMICAL SENSOR; GLUCOSE BIOSENSORS; HEMOGLOBIN; ASSAY; CONSTRUCTION; GLYCEMIA; PROTEIN; SYSTEM;
D O I
10.1016/j.bios.2016.08.005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Glycated proteins, such as glycated hemoglobin (HbA1c) or glycated albumin (GA) in the blood, are essential indicators of glycemic control for diabetes mellitus. Since GA, compared to HbA1c, is more sensitive to short term changes in glycemic levels, GA is expected to be used as an alternative or together with HbA1c as a surrogate marker indicator for glycemic control. In this paper we report the development of a sensing system for measuring GA by combining an enzyme analysis method, which is already used in clinical practice, with electrochemical principles. We used fructosyl amino acid oxidase, hexaammineruthenium(III) chloride as the electron mediator, and an inexpensive and economically attractive screen-printed carbon electrode. We used chronoamperometry to measure protease-digested GA samples. The developed sensor strips were able to measure protease-digested samples containing GA in very small sample volumes (1.3 mu L) within about 1 min. We also prepared enzyme sensor strips suitable for clinical use in which the enzyme and the mediator were deposited and dried on. This sensor system showed a clear correlation between the GA concentration and the resulting current. The strips were stable following 3 months of storage at 37 degrees C. We conclude that this disposable enzyme sensor strip system for measuring GA is suitable for point-of-care test (POCT) applications. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:167 / 173
页数:7
相关论文
共 56 条
  • [1] An improved enzymatic assay for glycated serum protein
    Abidin, Dewi
    Liu, Limin
    Dou, Chao
    Datta, Abhijit
    Yuan, Chong
    [J]. ANALYTICAL METHODS, 2013, 5 (10) : 2461 - 2469
  • [2] Extending Serum Half-life of Albumin by Engineering Neonatal Fc Receptor ( FcRn) Binding
    Andersen, Jan Terje
    Dalhus, Bjorn
    Viuff, Dorthe
    Ravn, Birgitte Thue
    Gunnarsen, Kristin Stoen
    Plumridge, Andrew
    Bunting, Karen
    Antunes, Filipa
    Williamson, Rebecca
    Athwal, Steven
    Allan, Elizabeth
    Evans, Leslie
    Bjoras, Magnar
    Kjaerulff, Soren
    Sleep, Darrell
    Sandlie, Inger
    Cameron, Jason
    [J]. JOURNAL OF BIOLOGICAL CHEMISTRY, 2014, 289 (19) : 13492 - 13502
  • [3] [Anonymous], AGR FOOD CHEM
  • [4] [Anonymous], BIOTECHNOL J
  • [5] [Anonymous], ONETOUCH ULTR STRIPS
  • [6] Graphene based aptasensor for glycated albumin in diabetes mellitus diagnosis and monitoring
    Apiwat, Chayachon
    Luksirikul, Patraporn
    Kankla, Pacharapon
    Pongprayoon, Prapasiri
    Treerattrakoon, Kiatnida
    Paiboonsukwong, Kittiphong
    Fucharoen, Suthat
    Dharakul, Tararaj
    Japrung, Deanpen
    [J]. BIOSENSORS & BIOELECTRONICS, 2016, 82 : 140 - 145
  • [7] Characterisation and analytical use of a polypyrrole electrode containing anti-human serum albumin
    Barisci, JN
    Hughes, D
    Minett, A
    Wallace, GG
    [J]. ANALYTICA CHIMICA ACTA, 1998, 371 (01) : 39 - 48
  • [8] GLYCOSYLATION OF HEMOGLOBIN - RELEVANCE TO DIABETES-MELLITUS
    BUNN, HF
    GABBAY, KH
    GALLOP, PM
    [J]. SCIENCE, 1978, 200 (4337) : 21 - 27
  • [9] An amperometric hemoglobin A1c biosensor based on immobilization of fructosyl amino acid oxidase onto zinc oxide nanoparticles-polypyrrole film
    Chawla, Sheetal
    Pundir, Chandra Shekhar
    [J]. ANALYTICAL BIOCHEMISTRY, 2012, 430 (02) : 156 - 162
  • [10] An electrochemical biosensor for fructosyl valine for glycosylated hemoglobin detection based on core-shell magnetic bionanoparticles modified gold electrode
    Chawla, Sheetal
    Pundir, Chandra Shekhar
    [J]. BIOSENSORS & BIOELECTRONICS, 2011, 26 (08) : 3438 - 3443