Malfunction of the da Vinci robotic system in urology

被引:16
作者
Chen, Cheng-Che [1 ]
Ou, Yen-Chuan [1 ,2 ,3 ]
Yang, Cheng-Kuang [1 ]
Chiu, Kun-Yuan [1 ]
Wang, Shian-Shiang [1 ]
Su, Chung-Kuang [1 ]
Ho, Hao-Chung [1 ]
Cheng, Chen-Li [1 ]
Chen, Chuan-Shu [1 ]
Lee, Jian-Ri [1 ]
Chen, Wen-Min [1 ]
机构
[1] Taichung Vet Gen Hosp, Div Urol, Dept Surg, Taichung 40705, Taiwan
[2] Chung Sang Med Univ, Inst Med, Taichung, Taiwan
[3] Natl Chi Nan Univ, Grad Inst Biomed & Biomed Technol, Dept Appl Chem, Nantou, Taiwan
关键词
da Vinci; malfunction; robot; urology; RADICAL PROSTATECTOMIES;
D O I
10.1111/j.1442-2042.2012.03010.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To analyze the incidence of malfunction of the da Vinci robotic system in a single center and to provide potential solutions. Methods: A total of 400 patients underwent da Vinci robotic urological surgery at Taichung Veterans General Hospital in Taichung, Taiwan, from December 2005 to April 2011. Episodes of malfunction of the robotic system were analyzed by period of operation, type of procedure, type of malfunction and management of the event. Results: Overall, 14 cases of malfunction occurred (3.5% of the entire series). Among them, five (1.25%) occurred before the surgery and nine (2.25%) intraoperatively. Operative procedures included radical prostatectomy, bilateral pelvic lymph node dissection, dismembered pyeloplasty, partial nephrectomy, nephroureterectomy, and radical and partial cystectomies. Areas of malfunctions included the robotic arm system and joint (11/14), optical system (1/14), power system and connector (1/14), endoscopic instrument (1/14), and software (1/14). In 10 cases, the failure was recoverable, whereas in four cases there was a critical failure, requiring a conversion to standard laparoscopy in three of them, and the rescheduling of the surgery in one case. Conclusions: The da Vinci robotic system is extremely reliable for use in urology. Malfunction is rare and the risk of critical failure is very low. Managing mechanical failure before or during the surgery is the key to maintaining the safety of patients undergoing robotic surgical procedures.
引用
收藏
页码:736 / 740
页数:5
相关论文
共 12 条
[1]  
Borden Lester S Jr, 2007, Can J Urol, V14, P3499
[2]  
Eichel L., 2005, AUA UPDATE SERIES, V24, P106
[3]   Malfunction of the da Vinci Robotic System During Robot-Assisted Laparoscopic Prostatectomy: An International Survey [J].
Kaushik, Dharam ;
High, Robin ;
Clark, Curtis J. ;
LaGrange, Chad A. .
JOURNAL OF ENDOUROLOGY, 2010, 24 (04) :571-575
[4]   Failure and Malfunction of da Vinci Surgical Systems During Various Robotic Surgeries: Experience From Six Departments at a Single Institute [J].
Kim, Won Tae ;
Ham, Won Sik ;
Jeong, Wooju ;
Song, Hyun Jung ;
Rha, Koon Ho ;
Choi, Young Deuk .
UROLOGY, 2009, 74 (06) :1234-1237
[5]   Mechanical failure rate of davinci robotic system: Implications for pre-op patient counseling [J].
Kozlowski, PM ;
Porter, CR ;
Corman, JM .
JOURNAL OF UROLOGY, 2006, 175 (04) :372-373
[6]   Robotic Equipment Malfunction During Robotic Prostatectomy: A Multi-institutional Study [J].
Lavery, Hugh J. ;
Thaly, Rahul ;
Albala, David ;
Ahlering, Thomas ;
Shalhav, Arieh ;
Lee, David ;
Fagin, Randy ;
Wiklund, Peter ;
Dasgupta, Prokar ;
Costello, Anthony J. ;
Tewari, Ashutosh ;
Coughlin, Geoff ;
Patel, Vipul R. .
JOURNAL OF ENDOUROLOGY, 2008, 22 (09) :2165-2168
[7]   Robotic-assisted laparoscopic radical prostatectomy: Learning curve of first 100 cases [J].
Ou, Yen Chuan ;
Yang, Chi Rei ;
Wang, John ;
Cheng, Chen Li ;
Patel, Vipul R. .
INTERNATIONAL JOURNAL OF UROLOGY, 2010, 17 (07) :635-640
[8]   The learning curve for reducing complications of robotic-assisted laparoscopic radical prostatectomy by a single surgeon [J].
Ou, Yen-Chuan ;
Yang, Chi-Rei ;
Wang, John ;
Yang, Chun-Kuang ;
Cheng, Chen-Li ;
Patel, Vipul R. ;
Tewari, Ashutosh K. .
BJU INTERNATIONAL, 2011, 108 (03) :420-425
[9]   A Unique Instrumental Malfunction during Robotic Prostatectomy [J].
Park, Sung Yul ;
Ahn, Jenny Jin-Kyung ;
Jeong, Wooju ;
Ham, Won Sik ;
Rha, Koon Ho .
YONSEI MEDICAL JOURNAL, 2010, 51 (01) :148-150
[10]   Complications and incidences in our first 250 robotic radical prostatectomies [J].
Pereira Arias, J. G. ;
Gamarra Quintanilla, M. ;
Leibar Tamayo, A. ;
Astobieta Odriozola, A. ;
Ibarluzea Gonzalez, G. .
ACTAS UROLOGICAS ESPANOLAS, 2010, 34 (05) :428-439