Low-Dose Computed Tomography of the Chest Using Iterative Reconstruction Versus Filtered Back Projection: Comparison of Image Quality

被引:26
作者
Lee, Youkyung [1 ]
Jin, Kwang Nam [1 ]
Lee, Nyoung Keun [2 ]
机构
[1] Seoul Metropolitan Govt Seoul Natl Univ, Boramae Med Ctr, Dept Radiol, Seoul 156010, South Korea
[2] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
关键词
lung; computed tomography; iterative reconstruction; radiation dose; image quality; CT; REDUCTION;
D O I
10.1097/RCT.0b013e318260ce39
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this study was to compare image quality of iterative reconstruction (IR) to filtered back projection (FBP) in low-dose computed tomography of the chest. Methods: Forty-three consecutive patients were retrospectively enrolled. Eight series of images were reconstructed using FBP and 7 levels of IR in each subject. Image noise, signal-to-noise ratio (SNR), and SNR improvement were measured. Two radiologists evaluated subjective artifact, image artificiality, and subjective overall image quality with 4- or 5-point scales. Results: Iterative reconstruction showed significantly lower image noise (135.5 +/- 36.6 vs 219.9 +/- 40.9) and higher SNR (0.36 +/- 0.12 vs 0.21 +/- 0.05) than FBP (P < 0.001). Signal-to-noise ratio improvement was 72.4% +/- 44.9%. Subjective artifact of FBP was significantly higher than IR images (P < 0.001). Image artificiality of IR was significantly higher than that of FBP (P < 0.001). Overall, subjective image quality was poor in FBP and acceptable or good in IR. Conclusions: With the use of IR, low-dose computed tomography of the chest would achieve less image noise and better image quality compared to the FBP.
引用
收藏
页码:512 / 517
页数:6
相关论文
共 16 条
[1]   Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic Screening [J].
Aberle, Denise R. ;
Adams, Amanda M. ;
Berg, Christine D. ;
Black, William C. ;
Clapp, Jonathan D. ;
Fagerstrom, Richard M. ;
Gareen, Ilana F. ;
Gatsonis, Constantine ;
Marcus, Pamela M. ;
Sicks, JoRean D. .
NEW ENGLAND JOURNAL OF MEDICINE, 2011, 365 (05) :395-409
[2]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]   Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study [J].
Flicek, Kristina T. ;
Hara, Amy K. ;
Silva, Alvin C. ;
Wu, Qing ;
Peter, Mary B. ;
Johnson, C. Daniel .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) :126-131
[4]   The National Lung Screening Trial: Overview and Study Design [J].
Gatsonis, Constantine A. .
RADIOLOGY, 2011, 258 (01) :243-253
[5]   Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study [J].
Hara, Amy K. ;
Paden, Robert G. ;
Silva, Alvin C. ;
Kujak, Jennifer L. ;
Lawder, Holly J. ;
Pavlicek, William .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (03) :764-771
[6]   Radiation dose reduction in chest CT: A review [J].
Kubo, Takeshi ;
Lin, Pei-Jan Paul ;
Stiller, Wolfram ;
Takahashi, Masaya ;
Kauczor, Hans-Ulrich ;
Ohno, Yoshiharu ;
Hatabu, Hiroto .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) :335-343
[7]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[8]   A Prospective Evaluation of Dose Reduction and Image Quality in Chest CT Using Adaptive Statistical Iterative Reconstruction [J].
Leipsic, Jonathon ;
Nguyen, Giang ;
Brown, Jaqueline ;
Sin, Don ;
Mayo, John R. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (05) :1095-1099
[9]   Radiologic and Nuclear Medicine Studies in the United States and Worldwide: Frequency, Radiation Dose, and Comparison with Other Radiation Sources-1950-2007 [J].
Mettler, Fred A., Jr. ;
Bhargavan, Mythreyi ;
Faulkner, Keith ;
Gilley, Debbie B. ;
Gray, Joel E. ;
Ibbott, Geoffrey S. ;
Lipoti, Jill A. ;
Mahesh, Mahadevappa ;
McCrohan, John L. ;
Stabin, Michael G. ;
Thomadsen, Bruce R. ;
Yoshizumi, Terry T. .
RADIOLOGY, 2009, 253 (02) :520-531
[10]   Initial Performance Characterization of a Clinical Noise-Suppressing Reconstruction Algorithm for MDCT [J].
Noel, Peter B. ;
Fingerle, Alexander A. ;
Renger, Bernhard ;
Muenzel, Daniela ;
Rummeny, Ernst J. ;
Dobritz, Martin .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (06) :1404-1409