Incidence and mechanisms of longitudinal stent deformation associated with Biomatrix, Resolute, Element, and Xience stents: Angiographic and case-by-case review of 1,800 PCIs

被引:23
作者
Arnous, Samer [1 ,2 ]
Shakhshir, Nizar [1 ,2 ]
Wiper, Andrew [1 ,2 ]
Ordoubadi, Farzin-Farth [1 ,2 ]
Williams, Paul [1 ,2 ]
Clarke, Bernard [1 ,2 ]
Mahadavan, Vaikom [1 ,2 ]
El-Omar, Magdi [1 ,2 ]
Mamas, Mamas [1 ,2 ]
Fraser, Douglas [1 ,2 ]
机构
[1] Manchester Royal Infirm, Manchester Heart Ctr, Manchester M13 9WL, Lancs, England
[2] Univ Manchester, Cardiovasc Inst, Manchester M13 9PT, Lancs, England
关键词
longitudinal stent deformation; drug-eluting stents; percutaneous coronary intervention; EVEROLIMUS-ELUTING STENTS; CORONARY STENTS; PROMUS ELEMENT; DUTCH PEERS; MULTICENTER; EXPERIENCE; INSIGHTS; MANUFACTURER; COMPRESSION; OUTCOMES;
D O I
10.1002/ccd.25790
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThere is conflicting evidence regarding the incidence of longitudinal stent deformation (LSD) in contemporary practice. Methods and resultsTo assess the incidence and mechanism of LSD across commonly used DES platforms, we performed a case-by-case review of 1,800 PCI cases involving 450 consecutive procedures using Biomatrix Flex, Resolute Integrity, Promus Element, and Xience V stents, respectively, between January 2009 and December 2011. LSD was detected in a higher proportion with Promus Element [15 (3.1%)] compared with other platforms (Xience V [4 (0.9%)], Biomatrix [3 (0.7%)], Resolute [3 (0.7%)]; P=0.002). LSD was characterized as guide catheter/guide extension induced, or as impact from secondary devices such as postdilatation balloons or IVUS catheters. The incidence of guide catheter/guide extension LSD was similar across platforms; (Promus Element [5 (1.1%)], Xience V [4 (0.9%)], Biomatrix [3 (0.7%)], Resolute [3 (0.7%)]; P=0.85). Secondary device LSD occurred exclusively with Promus Element (9/450 cases [2%] (P<0.0001). Re-entering the deformed stent was more difficult in cases of secondary device LSD (6/9 compared with 0/12 treated cases; P<0.001). Univariate predictors of LSD were previous CABG, culprit vessel, ostial involvement, and lesion tortuosity. Multivariate predictors of LSD were the Promus Element stent (OR 5.53 CI[1.54-19.85]), Guideliner use (OR 22.09 CI[4.73-103]), postdilation balloons (OR 5.47 CI[1.31-22.81]) and number of stents deployed (OR 2.06 CI[1.45-2.9]. ConclusionLSD is more common than previously reported. LSD by a guide catheter/guide extension occurred equally with all platforms, however, LSD associated with secondary devices only occurred with the Element stent. These findings have important implications regarding current and future stent designs. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1002 / 1011
页数:10
相关论文
共 16 条
[11]  
Rigattieri S, 2013, J INVASIVE CARDIOL, V25, pE101
[12]   Angiographic and clinical analysis of 164 cases of longitudinal stent deformation: comparison of cases from a multicentre case series with cases identified from the MAUDE database [J].
Shakhshir, Nizar ;
Abdel-Wahab, Mohamed ;
Aminian, Adel ;
Arnous, Samer ;
Bartorelli, Antonio L. ;
Colombo, Antonio ;
Cristina, Ferrari ;
Foin, Nicolas ;
Latib, Azeem ;
Malik, Iqbal S. ;
Mamas, Mamas ;
Richardt, Gert ;
Williams, Paul D. ;
Fraser, Douglas G. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) :B145-B146
[13]   Procedural trauma risks longitudinal shortening of the promus element stent platform [J].
Shannon, Joanne ;
Latib, Azeem ;
Takagi, Kensuke ;
Chieffo, Alaide ;
Figini, Filippo ;
Sacco, Francesco ;
Ferrarello, Santo ;
Montorfano, Matteo ;
Colombo, Antonio .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (05) :810-817
[14]   DUrable polymer-based sTent CHallenge of Promus ElemEnt versus ReSolute integrity (DUTCH PEERS): Rationale and study design of a randomized multicenter trial in a Dutch all-comers population [J].
Tandjung, Kenneth ;
Basalus, Mounir W. Z. ;
Sen, Hanim ;
Jessurun, Gillian A. J. ;
Danse, Peter W. ;
Stoel, Martin ;
Linssen, Gerard C. M. ;
Derks, Anita ;
van Loenhout, Ton T. ;
Nienhuis, Mark B. ;
Hautvast, Raymond W. M. ;
von Birgelen, Clemens .
AMERICAN HEART JOURNAL, 2012, 163 (04) :557-562
[15]   Third-generation zotarolimus-eluting and everolimus-eluting stents in all-comer patients requiring a percutaneous coronary intervention (DUTCH PEERS): a randomised, single-blind, multicentre, non-inferiority trial [J].
von Birgelen, Clemens ;
Sen, Hanim ;
Lam, Ming Kai ;
Danse, Peter W. ;
Jessurun, Gillian A. J. ;
Hautvast, Raymond W. M. ;
van Houwelingen, Gert K. ;
Schramm, Alexander R. ;
Gin, R. Melvyn Tjon Joe ;
Louwerenburg, Johannes W. ;
de Man, Frits H. A. F. ;
Stoel, Martin G. ;
Lowik, Marije M. ;
Linssen, Gerard C. M. ;
Said, Salah A. M. ;
Nienhuis, Mark B. ;
Verhorst, Patrick M. J. ;
Basalus, Mounir W. Z. ;
Doggen, Carine J. M. ;
Tandjung, Kenneth .
LANCET, 2014, 383 (9915) :413-423
[16]   Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms [J].
Williams, Paul D. ;
Mamas, Mamas A. ;
Morgan, Kenneth P. ;
El-Omar, Magdi ;
Clarke, Bernard ;
Bainbridge, Anthony ;
Fath-Ordoubadi, Farzin ;
Fraser, Douglas G. .
EUROINTERVENTION, 2012, 8 (02) :267-274