Incidence and mechanisms of longitudinal stent deformation associated with Biomatrix, Resolute, Element, and Xience stents: Angiographic and case-by-case review of 1,800 PCIs

被引:23
作者
Arnous, Samer [1 ,2 ]
Shakhshir, Nizar [1 ,2 ]
Wiper, Andrew [1 ,2 ]
Ordoubadi, Farzin-Farth [1 ,2 ]
Williams, Paul [1 ,2 ]
Clarke, Bernard [1 ,2 ]
Mahadavan, Vaikom [1 ,2 ]
El-Omar, Magdi [1 ,2 ]
Mamas, Mamas [1 ,2 ]
Fraser, Douglas [1 ,2 ]
机构
[1] Manchester Royal Infirm, Manchester Heart Ctr, Manchester M13 9WL, Lancs, England
[2] Univ Manchester, Cardiovasc Inst, Manchester M13 9PT, Lancs, England
关键词
longitudinal stent deformation; drug-eluting stents; percutaneous coronary intervention; EVEROLIMUS-ELUTING STENTS; CORONARY STENTS; PROMUS ELEMENT; DUTCH PEERS; MULTICENTER; EXPERIENCE; INSIGHTS; MANUFACTURER; COMPRESSION; OUTCOMES;
D O I
10.1002/ccd.25790
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThere is conflicting evidence regarding the incidence of longitudinal stent deformation (LSD) in contemporary practice. Methods and resultsTo assess the incidence and mechanism of LSD across commonly used DES platforms, we performed a case-by-case review of 1,800 PCI cases involving 450 consecutive procedures using Biomatrix Flex, Resolute Integrity, Promus Element, and Xience V stents, respectively, between January 2009 and December 2011. LSD was detected in a higher proportion with Promus Element [15 (3.1%)] compared with other platforms (Xience V [4 (0.9%)], Biomatrix [3 (0.7%)], Resolute [3 (0.7%)]; P=0.002). LSD was characterized as guide catheter/guide extension induced, or as impact from secondary devices such as postdilatation balloons or IVUS catheters. The incidence of guide catheter/guide extension LSD was similar across platforms; (Promus Element [5 (1.1%)], Xience V [4 (0.9%)], Biomatrix [3 (0.7%)], Resolute [3 (0.7%)]; P=0.85). Secondary device LSD occurred exclusively with Promus Element (9/450 cases [2%] (P<0.0001). Re-entering the deformed stent was more difficult in cases of secondary device LSD (6/9 compared with 0/12 treated cases; P<0.001). Univariate predictors of LSD were previous CABG, culprit vessel, ostial involvement, and lesion tortuosity. Multivariate predictors of LSD were the Promus Element stent (OR 5.53 CI[1.54-19.85]), Guideliner use (OR 22.09 CI[4.73-103]), postdilation balloons (OR 5.47 CI[1.31-22.81]) and number of stents deployed (OR 2.06 CI[1.45-2.9]. ConclusionLSD is more common than previously reported. LSD by a guide catheter/guide extension occurred equally with all platforms, however, LSD associated with secondary devices only occurred with the Element stent. These findings have important implications regarding current and future stent designs. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1002 / 1011
页数:10
相关论文
共 16 条
[1]   Longitudinal Deformation of Contemporary Coronary Stents: An Integrated Analysis of Clinical Experience and Observations from the Bench [J].
Abdel-Wahab, Mohamed ;
Sulimov, Dmitriy S. ;
Kassner, Guido ;
Geist, Volker ;
Toelg, Ralph ;
Richardt, Gert .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2012, 25 (06) :576-585
[2]   Comments on "Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the FDA Manufacturer and User Facility Device Experience database" by Mamas et al [J].
Aminian, Adel ;
Lalmand, Jacques .
EUROINTERVENTION, 2012, 8 (08) :990-991
[3]   Longitudinal compression: a "new" complication with modern coronary stent platforms - time to think beyond deliverability? [J].
Hanratty, Colm G. ;
Walsh, Simon J. .
EUROINTERVENTION, 2011, 7 (07) :872-877
[4]   Longitudinal stent deformation: quantitative coronary angiographic analysis from the PERSEUS and PLATINUM randomised controlled clinical trials [J].
Kereiakes, Dean J. ;
Popma, Jeffrey J. ;
Cannon, Louis A. ;
Kandzari, David E. ;
Kimmelstiel, Carey D. ;
Meredith, Ian T. ;
Teirstein, Paul S. ;
Verheye, Stefan ;
Allocco, Dominic J. ;
Dawkins, Keith D. ;
Stone, Gregg W. .
EUROINTERVENTION, 2012, 8 (02) :187-195
[5]   Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: Predisposing mechanical properties, incidence, and predictors in a large patient cohort [J].
Leibundgut, Gregor ;
Gick, Michael ;
Toma, Aurel ;
Valina, Christian ;
Loeffelhardt, Nikolaus ;
Buettner, Heinz Joachim ;
Neumann, Franz-Josef .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (05) :E206-E214
[6]   Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration Manufacturer and User Facility Device Experience database [J].
Mamas, Mamas A. ;
Williams, Paul D. .
EUROINTERVENTION, 2012, 8 (02) :196-204
[7]   Stent Longitudinal Integrity Bench Insights Into a Clinical Problem [J].
Ormiston, John A. ;
Webber, Bruce ;
Webster, Mark W. I. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (12) :1310-1317
[8]   A Randomized Comparison of Platinum Chromium- Based Everolimus-Eluting Stents Versus Cobalt Chromium-Based ZotarolimusEluting Stents in All- Comers Receiving Percutaneous Coronary Intervention [J].
Park, Kyung Woo ;
Kang, Si-Hyuck ;
Kang, Hyun-Jae ;
Koo, Bon-Kwon ;
Park, Byoung-Eun ;
Cha, Kwang Soo ;
Rhew, Jay Young ;
Jeon, Hui-Kyoung ;
Shin, Eun-Seok ;
Oh, Ju Hyeon ;
Jeong, Myung-Ho ;
Kim, Sanghyun ;
Hwang, Kyung-Kuk ;
Yoon, Jung-Han ;
Lee, Sung Yun ;
Park, Tae-Ho ;
Moon, Keon Woong ;
Kwon, Hyuck-Moon ;
Hur, Seung-Ho ;
Ryu, Jae-Kean ;
Lee, Bong-Ryul ;
Park, Yong Whi ;
Chae, In-Ho ;
Kim, Hyo-Soo .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 63 (25) :2805-2816
[9]   The 'Harmonizing Optimal Strategy for Treatment of coronary artery stenosis - sAfety & effectiveneSS of drug-elUting stents & antiplatelet REgimen' (HOST-ASSURE) trial: study protocol for a randomized controlled trial [J].
Park, Kyung Woo ;
Park, Byoung-Eun ;
Kang, Si-Hyuck ;
Park, Jin-Joo ;
Lee, Seung-Pyo ;
Cha, Kwang Soo ;
Rhew, Jay Young ;
Jeon, Hui-Kyoung ;
Shin, Eun Seok ;
Oh, Ju Hyeon ;
Jeong, Myung-Ho ;
Kim, Sanghyun ;
Hwang, Kyung-Kuk ;
Yoon, Jung-Han ;
Lee, Sung Yun ;
Park, Tae-Ho ;
Moon, Keon Woong ;
Kwon, Hyuck-Moon ;
Chae, In-Ho ;
Kim, Hyo-Soo .
TRIALS, 2012, 13
[10]   Engineering assessment of the longitudinal compression behaviour of contemporary coronary stents [J].
Prabhu, Santosh ;
Schikorr, Tanya ;
Mahmoud, Tamer ;
Jacobs, James ;
Potgieter, Adriaan ;
Simonton, Charles .
EUROINTERVENTION, 2012, 8 (02) :275-281