Risk factors for aseptic loosening of Muller-type straight stems A registry-based analysis of 828 consecutive cases with a minimum follow-up of 16 years

被引:15
作者
Clauss, Martin [1 ]
Gersbach, Silke [1 ]
Butscher, Andre [2 ]
Ilchmann, Thomas [1 ]
机构
[1] Kantonsspital Baselland Liestal, Clin Orthoped & Trauma Surg, Bettlach, Switzerland
[2] RMS Fdn, Bettlach, Switzerland
关键词
TOTAL HIP-ARTHROPLASTY; ABDUCTOR MUSCLE STRENGTH; FEMORAL COMPONENTS; CEMENTED STEMS; FRENCH PARADOX; SURFACE FINISH; REPLACEMENT; PROSTHESES; SURVIVAL; MIGRATION;
D O I
10.3109/17453674.2013.810517
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose Even small differences in design variables for the femoral stem may influence the outcome of a hip arthroplasty. We performed a risk factor analysis for aseptic loosening of 4 different versions of cemented Muller-type straight stems with special emphasis on design modifications (2 shapes, MSS or SL, and 2 materials, CoNiCrMo (Co) or Ti-6Al-7Nb (Ti)). Methods We investigated 828 total hip replacements, which were followed prospectively in our in-house register. All stems were operated in the same setup, using Sulfix-6 bone cement and a second-generation cementing technique. Demographic and design-specific risk factors were analyzed using an adjusted Cox regression model. Results The 4 versions showed marked differences in 15-year stem survival with aseptic loosening as the endpoint: 94% (95% CI: 89-99) for MSS Co, 83% (CI: 75-91) for SL Co, 81% (CI: 76-87) for MSS Ti and 63% (CI: 56-71) for SL Ti. Cox regression analysis showed a relative risk (RR) for aseptic loosening of 3 (CI: 2-5) for stems made of Ti and of 2 (CI: 1-2) for the SL design. The RR for aseptic stem loosening increased to 8 (CI: 4-15) when comparing the most and the least successful designs (MSS Co and SL Ti). Interpretation Cemented Muller-type straight stems should be MSS-shaped and made of a material with high flexural strength (e. g. cobalt-chrome). The surface finish should be polished (Ra < 0.4 mu m). These technical aspects combined with modern cementing techniques would improve the survival of Miller-type straight stems. This may be true for all types of cemented stems.
引用
收藏
页码:353 / 359
页数:7
相关论文
共 34 条
[11]  
FOWLER JL, 1988, ORTHOP CLIN N AM, V19, P477
[12]   Is there still a place for the cemented titanium femoral stem? 10,108 cases from the Norwegian Arthroplasty Register [J].
Hallan, Geir ;
Espehaug, Birgitte ;
Furnes, Ove ;
Wangen, Helge ;
Hol, Paul J. ;
Ellison, Peter ;
Havelin, Leif I. .
ACTA ORTHOPAEDICA, 2012, 83 (01) :1-6
[13]   Minimum 10-year survival of Kerboull cemented stems according to surface finish [J].
Hamadouche, Moussa ;
Baque, Francois ;
Lefevre, Nicolas ;
Kerboull, Marcel .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (02) :332-339
[14]   The Norwegian Arthroplasty Register -: 11 years and 73,000 arthroplasties [J].
Havelin, LI ;
Engesæter, LB ;
Espehaug, B ;
Furnes, O ;
Lie, SA ;
Vollset, SE .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (04) :337-353
[15]   Long-term registration has improved the quality of hip replacement - A review of the Swedish THR Register comparing 160,000 cases [J].
Herberts, P ;
Malchau, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (02) :111-121
[16]   In vivo surface wear mechanisms of femoral components of cemented total hip arthroplasties -: The influence of wear mechanism on clinical outcome [J].
Howell, JR ;
Blunt, LA ;
Doyle, C ;
Hooper, RM ;
Lee, AJC ;
Ling, RSM .
JOURNAL OF ARTHROPLASTY, 2004, 19 (01) :88-101
[17]   Loosening of matt and polished cemented femoral stems [J].
Howie, DW ;
Middleton, RG ;
Costi, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (04) :573-576
[18]   Acetabular polyethylene wear volume after hip replacement: Reliability of volume calculations from plain radiographs [J].
Ilchmann, Thomas ;
Lueem, Martin ;
Pannhorst, Silke ;
Clauss, Martin .
WEAR, 2012, 282 :69-75
[19]   Finite-element analysis of failure of the Capital Hip designs [J].
Janssen, D ;
Aquarius, R ;
Stolk, J ;
Verdonschot, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (11) :1561-1567
[20]   A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 years [J].
Kadar, Thomas ;
Hallan, Geir ;
Aamodt, Arild ;
Indrekvam, Kari ;
Badawy, Mona ;
Havelin, Leif Ivar ;
Stokke, Terje ;
Haugan, Kristin ;
Espehaug, Birgitte ;
Furnes, Ove .
ACTA ORTHOPAEDICA, 2011, 82 (05) :538-544