Needle-tissue interaction forces - A survey of experimental data

被引:194
作者
van Gerwen, Dennis J. [1 ]
Dankelman, Jenny [1 ]
van den Dobbelsteen, John J. [1 ]
机构
[1] Delft Univ Technol, Dept Biomech Engn, NL-2628 CD Delft, Netherlands
关键词
Puncture; Friction; Cutting; Resistance; Insertion; SOFT-TISSUE; ELASTOMER MEMBRANES; INSERTION FORCE; PUNCTURE FORCE; SIMULATION; MODEL; FRICTION; SYSTEM; MOTION; IDENTIFICATION;
D O I
10.1016/j.medengphy.2012.04.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The development of needles, needle-insertion simulators, and needle-wielding robots for use in a clinical environment depends on a thorough understanding of the mechanics of needle-tissue interaction. It stands to reason that the forces arising from this interaction are influenced by numerous factors, such as needle type, insertion speed, and tissue characteristics. However, exactly how these factors influence the force is not clear. For this reason, the influence of various factors on needle insertion-force was investigated by searching literature for experimental data. This resulted in a comprehensive overview of experimental insertion-force data available in the literature, grouped by factor for quick reference. In total, 99 papers presenting such force data were found, with typical peak forces in the order of 1-10N. The data suggest, for example, that higher velocity tends to decrease puncture force and increase friction. Furthermore, increased needle diameter was found to increase peak forces, and conical needles were found to create higher peak forces than beveled needles. However, many questions remain open for investigation, especially those concerning the influence of tissue characteristics. (C) 2012 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:665 / 680
页数:16
相关论文
共 107 条
[91]   Deflections of a rubber membrane [J].
Selvadurai, APS .
JOURNAL OF THE MECHANICS AND PHYSICS OF SOLIDS, 2006, 54 (06) :1093-1119
[92]   Experimental investigation into the deep penetration of soft solids by sharp and blunt punches, with application to the piercing of skin [J].
Shergold, OA ;
Fleck, NA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (05) :838-848
[93]  
Simone C, 2002, 2002 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS I-IV, PROCEEDINGS, P2085, DOI 10.1109/ROBOT.2002.1014848
[94]  
SIMONE C, 2002, THESIS J HOPKINS U
[95]  
Stellman J.T., 2009, THESIS GEORGIA I TEC
[96]  
Suzuki Toshiyasu, 2004, Tokai J Exp Clin Med, V29, P175
[97]   Instrumentation of the Loss-of-Resistance Technique for Epidural Needle Insertion [J].
Tran, Denis ;
Kamani, Allaudin A. ;
Lessoway, Victoria A. ;
Rohling, Robert N. ;
Hor, King-Wei .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (03) :820-827
[98]   Force sensing and its application in minimally invasive surgery and therapy: a survey [J].
Trejos, A. L. ;
Patel, R. V. ;
Naish, M. D. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2010, 224 (C7) :1435-1454
[99]   A haptic unit designed for magnetic-resonance-guided biopsy [J].
Tse, Z. T. H. ;
Elhawary, H. ;
Rea, M. ;
Young, I. ;
Davis, B. L. ;
Lamperth, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H2) :159-172
[100]   Simulation of ultrasound guided needle puncture using patient specific data with 3D textures and volume haptics [J].
Vidal, F. P. ;
John, N. W. ;
Healey, A. E. ;
Gould, D. A. .
COMPUTER ANIMATION AND VIRTUAL WORLDS, 2008, 19 (02) :111-127