An automatic evaluation system for contrast-detail phantom images in digital radiography

被引:0
作者
Lin, Chuen-Horng [1 ]
Lin, Chih-Yang [2 ]
Cheng, Yu-Jung [1 ]
Chan, Yung-Kuan [3 ]
机构
[1] Natl Taichung Univ Sci & Technol, Dept Comp Sci & Informat Engn, Taichung, Taiwan
[2] Asia Univ, Dept Comp Sci & Informat Engn, Taichung, Taiwan
[3] Natl Chung Hsing Univ, Dept Management Informat Syst, Taichung 40227, Taiwan
关键词
contrast-detail phantom; bilinear pixel interpolation; region shifting; QUALITY;
D O I
10.1002/ima.22025
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Currently, contrast-detail phantom (CD phantom) images are widely used to assist medical experts for diagnosis, because these images can be used to study imaging quality and assess the radiation dose amount for digital X-ray imaging systems. However, the quality of the CD phantom image varies from person to person, increasing the possibility of false diagnosis. In this article, we propose an automatic evaluation system to measure the quality of a phantom image an alternative to assessment by radiologists. The proposed method is based on bilinear pixel interpolation, geometric transformation, and region shifting to efficiently and objectively evaluate the quality of a phantom image. Moreover, misjudgment modification is also proposed to further improve the recognition accuracy. The experimental results show that the accuracy of the proposed method is very close to that of the radiologists' observations. The results have also proven that the proposed method is a feasible way to evaluate the quality of a phantom image in an objective and automatic manner. (c) 2012 Wiley Periodicals, Inc. Int J Imaging Syst Technol, 22, 214225, 2012 ; Published online in Wiley Online Library (wileyonlinelibrary.com). DOI 10.1002/ima.22025
引用
收藏
页码:214 / 225
页数:12
相关论文
共 15 条
  • [1] Bijkerk K. R., 2000, CDMAM PHANTOM TYPE 3, V3, P4
  • [2] Optimization of tube potential-filter combinations for film-screen mammography - A contrast detail phantom study
    Chida, K
    Zuguchi, M
    Sai, M
    Saito, H
    Yamada, T
    Ishibashi, T
    Ito, D
    Kimoto, N
    Kohzuki, M
    Takahashi, S
    [J]. CLINICAL IMAGING, 2005, 29 (04) : 246 - 250
  • [3] Quality control phantom for digital chest radiography
    Chotas, HG
    Floyd, CE
    Johnson, GA
    Ravin, CE
    [J]. RADIOLOGY, 1997, 202 (01) : 111 - 116
  • [4] ASSESSMENT OF NOISE IN A DIGITAL IMAGE USING THE JOIN-COUNT STATISTIC AND THE MORAN TEST
    CHUANG, KS
    HUANG, HK
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (02) : 357 - 369
  • [5] GELEIJNS J, 1993, RADIAT PROT DOSIM, V49, P83
  • [6] Gonzalez R. C., 2004, Digital image processing using MATLAB, VSecond
  • [7] Gonzalez R. C., 1992, DIGITAL IMAGE PROCES, V2nd
  • [8] Lu Z F, 2003, J Appl Clin Med Phys, V4, P91, DOI 10.1120/1.1524950
  • [9] Michette A.G., 1993, X-ray Science and Technology
  • [10] A clinical evaluation of the Image Quality computer program, CoCIQ
    Norrman, E
    Gårdestig, M
    Persliden, J
    Geijer, H
    [J]. JOURNAL OF DIGITAL IMAGING, 2005, 18 (02) : 138 - 144