The influence of wear paths produced by hip replacement patients during normal walking on wear rates

被引:22
作者
Bennett, Damien [1 ]
Humphreys, Lee [2 ]
O'Brien, Seamus [1 ]
Kelly, Connor [2 ]
Orr, John [3 ]
Beverland, David E. [1 ]
机构
[1] Musgrave Pk Hosp, Outcome Assessment Unit, Belfast, Antrim, North Ireland
[2] Musgrave Pk Hosp, Gait Anal Lab, Belfast, Antrim, North Ireland
[3] Queens Univ Belfast, Sch Mech & Aerosp Engn, Belfast, Antrim, North Ireland
关键词
wear; gait kinematics; hip replacement; multidirectional motion;
D O I
10.1002/jor.20583
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Variation in wear paths is known to greatly affect wear rates in vitro, with multidirectional paths producing much greater wear than unidirectional paths. This study investigated the relationship between multidirectional motion at the hip joint, as measured by aspect ratio, sliding distance, and wear rate for 164 hip replacements. Kinematic input from three-dimensional gait analysis was used to determine the wear paths. Activity cycles were determined for a subgroup of 100 patients using a pedometer study, and the relationship between annual sliding distance and wear rate was analyzed. Poor correlations were found between both aspect ratio and sliding distance and wear rate for the larger group and between annual sliding distance and wear rate for the subgroup. However, patients who experienced a wear rate <0.08 mm/year showed a strong positive correlation between the combination of sliding distance, activity levels, and aspect ratio and wear rate (adjusted r(2) = 55.4%). This group may represent those patients who experience conditions that most closely match those that prevail in simulator and laboratory tests. Although the shape of wear paths, their sliding distance, and the number of articulation cycles at the hip joint affect wear rates in simulator studies, this relationship was not seen in this clinical study. Other factors such as lubrication, loading conditions and roughness of the femoral head may influence the wear rate. (C) 2008 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.
引用
收藏
页码:1210 / 1217
页数:8
相关论文
共 34 条
[1]   Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions [J].
Affatato, S ;
Bersaglia, G ;
Rocchi, M ;
Taddei, P ;
Fagnano, C ;
Toni, A .
BIOMATERIALS, 2005, 26 (16) :3259-3267
[2]   A new approach to determine the hip rotation profile from clinical gait analysis data [J].
Baker, R ;
Finney, L ;
Orr, J .
HUMAN MOVEMENT SCIENCE, 1999, 18 (05) :655-667
[3]   Accuracy of five electronic pedometers for measuring distance walked [J].
Bassett, DR ;
Ainsworth, BE ;
Leggett, SR ;
Mathien, CA ;
Main, JA ;
Hunter, DC ;
Duncan, GE .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1996, 28 (08) :1071-1077
[4]   The influence of shape and sliding distance of femoral head movement loci on the wear of acetabular cups in total hip arthroplasty [J].
Bennett, D ;
Orr, JF ;
Beverland, DE ;
Baker, R .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H6) :393-402
[5]   Movement loci of selected points on the femoral head for individual total hip arthroplasty patients using three-dimensional computer simulation [J].
Bennett, DB ;
Orr, JF ;
Baker, R .
JOURNAL OF ARTHROPLASTY, 2000, 15 (07) :909-915
[6]   Metal-on-metal hip simulator study of increased wear particle surface area due to 'severe' patient activity [J].
Bowsher, JG ;
Hussain, A ;
Williams, PA ;
Shelton, JC .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :279-287
[7]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[8]  
BYGOTT J, 1967, INTRO MAPWORK PRACTI
[9]   Force track analysis of contemporary hip simulators [J].
Calonius, O ;
Saikko, V .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1719-1726
[10]  
DEVANE PA, 1995, CLIN ORTHOP RELAT R, P303