Experimental Validation of Finite Element Models of Intact and Implanted Composite Hemipelvises Using Digital Image Correlation

被引:48
作者
Ghosh, Rajesh [1 ]
Gupta, Sanjay [1 ]
Dickinson, Alexander [2 ]
Browne, Martin [2 ]
机构
[1] Indian Inst Technol Kharagpur, Dept Mech Engn, Kharagpur 721302, W Bengal, India
[2] Univ Southampton, Sch Engn Sci, Bioengn Sci Res Grp, Southampton S017 1BJ, Hants, England
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2012年 / 134卷 / 08期
关键词
pelvis biomechanics; in vitro testing; digital image correlation; strain gauging; finite element analysis; STRAIN DISTRIBUTION; TRABECULAR BONE; ACETABULAR CUP; PROXIMAL FEMUR; LOAD-TRANSFER; PELVIS; REPLACEMENT; DENSITY;
D O I
10.1115/1.4007173
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A detailed understanding of the changes in load transfer due to implantation is necessary to identify potential failure mechanisms of orthopedic implants. Computational finite element (FE) models provide full field data on intact and implanted bone structures, but their validity must be assessed for clinical relevance. The aim of this study was to test the validity of FE predicted strain distributions for the intact and implanted pelvis using the digital image correlation (DIC) strain measurement technique. FE models of an in vitro hemipelvis test setup were produced, both intact and implanted with an acetabular cup. Strain predictions were compared to DIC and strain rosette measurements. Regression analysis indicated a strong linear relationship between the measured and predicted strains, with a high correlation coefficient (R = 0.956 intact, 0.938 implanted) and a low standard error of the estimate (SE = 69.53 mu epsilon, 75.09 mu epsilon). Moreover, close agreement between the strain rosette and DIC measurements improved confidence in the validity of the DIC technique. The FE model therefore was supported as a valid predictor of the measured strain distribution in the intact and implanted composite pelvis models, confirming its suitability for further computational investigations. [DOI: 10.1115/1.4007173]
引用
收藏
页数:9
相关论文
共 53 条
[1]   Subject-specific finite element model of the pelvis: Development, validation and sensitivity studies [J].
Anderson, AE ;
Peters, CL ;
Tuttle, BD ;
Weiss, JA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (03) :364-373
[2]  
[Anonymous], 2009, VIC 3D TEST GUID 200
[3]   TEXTURE CORRELATION - A METHOD FOR THE MEASUREMENT OF DETAILED STRAIN DISTRIBUTIONS WITHIN TRABECULAR BONE [J].
BAY, BK .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1995, 13 (02) :258-267
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]   Vibration-assisted bone-graft compaction in impaction bone grafting of the femur [J].
Bolland, B. J. R. F. ;
New, A. M. R. ;
Madabhushi, S. P. G. ;
Oreffo, R. O. C. ;
Dunlop, D. G. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (05) :686-692
[6]   The biomechanics of a validated finite element model of stress shielding in a novel hybrid total knee replacement [J].
Bougherara, H. ;
Zdero, R. ;
Mahboob, Z. ;
Dubov, A. ;
Shah, S. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H10) :1209-1219
[7]  
Bremand F., 2008, P 10 INT C EXP ORL F
[8]   Trabecular bone strain changes associated with subchondral bone defects of the tibial plateau [J].
Brown, AN ;
McKinley, TO ;
Bay, BK .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (09) :638-643
[9]  
Christopher L. G., 2003, SUMM BIOENG C KEY BI
[10]   Effect of undersizing on the long-term stability of the Exeter hip stem: A comparative in vitro study [J].
Cristofolini, Luca ;
Erani, Paolo ;
Bialoblocka-Juszczyk, Ewa ;
Ohashi, Hirotsugu ;
Iida, Satoshi ;
Minato, Izumi ;
Viceconti, Marco .
CLINICAL BIOMECHANICS, 2010, 25 (09) :899-908