Potential and feasibility of parallel MRI at high field

被引:93
作者
Wiesinger, Florian
Van de Moortele, Pierre-Francois
Adriany, Gregor
De Zanche, Nicola
Ugurbil, Kamil
Pruessmann, Klaas P.
机构
[1] Univ Zurich, Inst Biomed Engn, CH-8092 Zurich, Switzerland
[2] Swiss Fed Inst Technol, CH-8092 Zurich, Switzerland
[3] Univ Minnesota, Ctr MR Res, Minneapolis, MN USA
关键词
parallel imaging; high field; 7T; SNR; SMASH; SENSE; GRAPPA;
D O I
10.1002/nbm.1050
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This survey focuses on the fusion of two major lines of recent progress in MRI methodology: parallel imaging with receiver coil arrays and the transition to high and ultra-high field strength for human applications. As discussed in this paper, combining the two developments has vast potential due to multiple specific synergies. First, parallel acquisition and high field are highly complementary in terms of their individual advantages and downsides. As a consequence, the joint approach generally offers enhanced flexibility in the design of scanning strategies. Second, increasing resonance frequency changes the electrodynamics of the MR signal in such a way that parallel imaging becomes more effective in large objects. The underlying conceptual and theoretical considerations are reviewed in detail. In further sections, technical challenges and practical aspects are discussed. The feasibility of parallel MRI at ultra-high field is illustrated by current results of parallel human MRI at 7 T. Copyright (C) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:368 / 378
页数:11
相关论文
共 107 条
[1]   Transmit and receive transmission line arrays for 7 tesla parallel imaging [J].
Adriany, G ;
Van de Moortele, PF ;
Wiesinger, F ;
Moeller, S ;
Strupp, JP ;
Andersen, P ;
Snyder, C ;
Zhang, XL ;
Chen, W ;
Pruessmann, KP ;
Boesiger, P ;
Vaughan, T ;
Ugurbil, K .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (02) :434-445
[2]  
AHN BC, 1986, IEEE T MED IMAGING, V5, P2
[3]  
[Anonymous], [No title captured], Patent No. 6633161
[4]   MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES [J].
MANSFIELD, P .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03) :L55-L58
[5]  
[Anonymous], P 8 ISMRM ANN M DENV
[6]   Coil-based artifact reduction [J].
Atkinson, D ;
Larkman, DJ ;
Batchelor, PG ;
Hill, DLG ;
Hajnal, JV .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (04) :825-830
[7]   Diffusion tensor imaging using single-shot SENSE-EPI [J].
Bammer, R ;
Auer, M ;
Keeling, SL ;
Augustin, M ;
Stables, LA ;
Prokesch, RW ;
Stollberger, R ;
Moseley, ME ;
Fazekas, F .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) :128-136
[8]   Improved diffusion-weighted single-shot echo-planar imaging (EPI) in stroke using Sensitivity Encoding (SENSE) [J].
Bammer, R ;
Keeling, SL ;
Augustin, M ;
Pruessmann, KP ;
Wolf, R ;
Stollberger, R ;
Hartung, HP ;
Fazekas, F .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) :548-554
[9]  
BAMMER R, 2004, P 12 ANN M ISMRM KYO, P2414
[10]   Simulation-based investigation of partially parallel imaging with a linear array at high accelerations [J].
Bankson, JA ;
Wright, SM .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (04) :777-786