HYBRID BEAMFORMATION FOR VOLUMETRIC ULTRASOUND IMAGING SCANNERS USING 2-D ARRAY TRANSDUCERS

被引:0
作者
Kim, Bae-Hyung [1 ]
Lee, Seungheun [1 ]
Kim, Youngil [1 ]
Cho, Kyungil [1 ]
Kim, Dongwook [1 ]
Song, Jongkeun [1 ]
机构
[1] Samsung Elect Co, Samsung Adv Inst Technol, Seoul, South Korea
来源
2012 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP 2012) | 2012年
关键词
Ultrasonic imaging; Hybrid beamforming; 2-D array transducer; Ultrasound scanner; Phased array system; COHERENT; DESIGN; SYSTEM;
D O I
暂无
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
In real-time 3-D ultrasound imaging using 2-D array transducers, manipulating massive ultrasound data acquired from a large number of system channels is a challenge as is interconnecting hundreds or thousands of elements of 2-D array with the imaging system's front-end electronics. Minimizing the number of transmitting and receiving elements and the firing events without degrading the spatial resolution is one of the solutions to reduce the overall system complexity and improve the frame rate. This paper introduces a new hybrid beamformation method for a medical image system. We also present an efficient 2-D array scanning method for miniaturized volumetric ultrasound imaging scanners. To evaluate our works, computer simulations are performed. From the simulation results, we have shown that the proposed hybrid beamforming method can provide comparable image quality to fully-sampled array beamformation, and the proposed method may be feasible for miniaturized real-time volumetric ultrasound scanners.
引用
收藏
页码:2297 / 2300
页数:4
相关论文
共 18 条
  • [1] Chang S.-H., U.S. Patent, Patent No. [5,901,708, 5901708]
  • [2] Coded excitation for diagnostic ultrasound: A system developer's perspective
    Chiao, RY
    Hao, XH
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (02) : 160 - 170
  • [3] Daft C. M. W., U.S. Patent Application, Patent No. 20070242567
  • [4] Three-dimensional ultrasound imaging
    Fenster, A
    Downey, DB
    Cardinal, HN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) : R67 - R99
  • [5] Packaging and design of reconfigurable arrays for volumetric imaging
    Fisher, R.
    Wodnicki, R.
    Cogan, S.
    Thomas, R.
    Mills, D.
    Woychik, C.
    Lewandowski, R.
    Thomenius, K.
    [J]. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 407 - 410
  • [6] Multielement synthetic transmit aperture imaging using temporal encoding
    Gammelmark, KL
    Jensen, JA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (04) : 552 - 563
  • [7] THE UNIFYING ROLE OF THE COARRAY IN APERTURE SYNTHESIS FOR COHERENT AND INCOHERENT IMAGING
    HOCTOR, RT
    KASSAM, SA
    [J]. PROCEEDINGS OF THE IEEE, 1990, 78 (04) : 735 - 752
  • [8] Jensen J. A., 1996, Medical & Biological Engineering & Computing, V34, P351
  • [9] Coherent array imaging using phased subarmys.: Part II:: Simulations and experimental results
    Johnson, JA
    Oralkan, Ö
    Ergun, S
    Demirci, U
    Karaman, M
    Khuri-Yakub, BT
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (01) : 51 - 64
  • [10] Capacitive micromachined ultrasonic transducers for medical imaging and therapy
    Khuri-Yakub, Butrus T.
    Oralkan, Oemer
    [J]. JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2011, 21 (05)