ECG Artifact Removal from EMG Recordings using Independent Component Analysis and Adapted Filter

被引:0
|
作者
Li, Yun [1 ]
Chen, Xiang [1 ]
Zhang, Xu [1 ]
Zhou, Ping
机构
[1] Univ Sci & Technol China, Dept Elect Sci & Technol, Hefei 230027, Peoples R China
关键词
SURFACE EMG; NOISE;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Surface electromyography (sEMG) recordings from trunk or limb muscles are often easily corrupted by electrocardiography (ECG) signals. In order to remove or reduce ECG in sEMG so as to improve the practicability, a novel signal filtering method with joint independent component analysis (ICA) and adaptive filtering (AF) is proposed in this paper. The method is validated with synthetic noisy EMG signals derived from 8-channel real sEMG added with 8-channel ECG recordings. Two groups of sEMG signals and two groups of ECG signals were used to examine the performance of the proposed method in our validation study. Experimental results demonstrate that the ICA+AF signal filtering method achieves better performance on reduction ECG artifact than the conventional Butterworth High-pass filter with 30 Hz cutoff frequency. The proposed method also performed well with 8-channel real ECG contaminated sEMG signals.
引用
收藏
页码:347 / 350
页数:4
相关论文
共 50 条
  • [1] ECG artifact removal from surface EMG using adaptive filter algorithm
    Yeom, H. (hyeom@eulji.ac.kr), 1600, Science and Engineering Research Support Society, Room 402, Man-Je Bld., 449-8, Ojung-Dong, Daedoek-Gu, Korea, Republic of (07):
  • [2] ECG ARTIFACT REMOVAL FROM SURFACE EMG SIGNALS BY COMBINING EMPIRICAL MODE DECOMPOSITION AND INDEPENDENT COMPONENT ANALYSIS
    Taelman, Joachim
    Mijovic, Bogdan
    Van Huffel, Sabine
    Devuyst, Stephanie
    Dutoit, Thierry
    BIOSIGNALS 2011, 2011, : 421 - +
  • [3] Removal of ECG interference from the EEG recordings in small animals using independent component analysis
    Tong, S
    Bezerianos, A
    Paul, J
    Zhu, Y
    Thakor, N
    JOURNAL OF NEUROSCIENCE METHODS, 2001, 108 (01) : 11 - 17
  • [4] The Removal of Blink and Saccade Artifact in EEG recordings by Independent Component Analysis
    Dong, Jie
    Wang, Tao
    Zhang, Ai-tao
    Dai, Hong-ya
    2010 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2010), VOLS 1-7, 2010, : 1071 - 1075
  • [5] An automated ECG-artifact removal method for trunk muscle surface EMG recordings
    Mak, Joseph N. F.
    Hu, Yong
    Luk, Keith D. K.
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (08) : 840 - 848
  • [6] CPR Artifact Reduction in the Human ECG Using Independent Component Analysis
    Granegger, M.
    Werther, T.
    Roehrich, M.
    Gilly, H.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 980 - 983
  • [7] CPR Artifact Reduction in the Human ECG by Using Constrained Independent Component Analysis
    Traxler, L.
    Granegger, M.
    Gilly, H.
    5TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2012, 37 : 207 - +
  • [8] Muscle Artifact Removal from Human Sleep EEG by Using Independent Component Analysis
    Maite Crespo-Garcia
    Mercedes Atienza
    Jose L. Cantero
    Annals of Biomedical Engineering, 2008, 36 : 467 - 475
  • [9] Muscle artifact removal from human sleep EEG by using independent component analysis
    Crespo-Garcia, Maite
    Atienza, Mercedes
    Cantero, Jose L.
    ANNALS OF BIOMEDICAL ENGINEERING, 2008, 36 (03) : 467 - 475
  • [10] Artifact Removal from ECG Signal using Notch Filter for Developing Countries
    Gupta, Varun
    Singh, Gavendra
    Sharma, Rohit
    Pahuja, Sharvan Kumar
    MEMS, NANO AND SMART SYSTEMS, PTS 1-6, 2012, 403-408 : 120 - +