Feasibility and Safety of Remote-Controlled Magnetic Navigation for Ablation of Atrial Fibrillation

被引:48
作者
Katsiyiannis, William T. [1 ]
Melby, Daniel P. [1 ]
Matelski, Jayme L. [1 ]
Ervin, Vanessa L. [1 ]
Laverence, Kerri L. [1 ]
Gornick, Charles C. [1 ]
机构
[1] Minneapolis Heart Inst Fdn, Minneapolis, MN USA
关键词
D O I
10.1016/j.amjcard.2008.08.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Radiofrequency ablation for atrial fibrillation (AF) involves complex catheter manipulation resulting in prolonged procedure time and fluoroscopy exposure. Remote magnetic navigation (RMN) represents a novel approach toward improving the ability to perform complex ablation. Forty patients underwent ablation for AF, 20 using RMN (NIOBE II, Stereotaxis, Inc) with a 4-mm-tip magnetic catheter (Celsius, Biosense Webster) and 20 using a conventional 8-mm-tip bidirectional ablation catheter (Blazer, Boston Scientific). All patients underwent a combined wide area circumferential ablation and segmental pulmonary vein (PV) isolation using a circular mapping catheter and cavotricuspid isthmus ablation for right atrial flutter. The procedural end point was PV entrance block. There was no difference in atrial size, left ventricular systolic function, or type of AF between groups. PV entrance block was achieved in all patients. Mean procedure time was 279 +/- 60 minutes in the conventional group versus 209 +/- 56 minutes in the RMN group (p <0.001). Mean fluoroscopy time in the conventional group was 58.6 +/- 21 minutes versus 19.5 +/- 9.8 in the RMN group (p <0.001.). At I year there were 15 patients in the conventional group and 16 in the RMN group free from clinical AF and off antiarrhythmic drugs (p = NS). There were 2 additional ablations performed for atypical atrial flutter in the conventional group and 3 in the RMN group (p = ns). Ablation catheter char formation was not observed. There were no procedural complications. In conclusion, radiofrequency ablation of AF performed with RMN is safe and feasible. Compared with conventional hand-navigated ablation, RMN ablation results in similar clinical outcomes with decreased fluoroscopy and procedure times. (C) 2008 Elsevier Inc. (Am J Cardiol 2008;102: 1674-1676)
引用
收藏
页码:1674 / 1676
页数:3
相关论文
共 12 条
[1]  
Chen X, 2007, EUR HEART J, V28, P164
[2]   Remote magnetic navigation - Human experience in pulmonary vein ablation [J].
Di Blase, Luigi ;
Fahmy, Tamer S. ;
Patel, Dimpi ;
Bai, Rong ;
Civello, Kenneth ;
Wazni, Oussama M. ;
Kanj, Mohamed ;
Elayi, Claude S. ;
Ching, Chi Keong ;
Khan, Mohamed ;
Popova, Lucie ;
Schweikert, Robert A. ;
Cummings, Jennifer E. ;
Burkhardt, J. David ;
Martin, David O. ;
Bhargava, Mandeep ;
Dresing, Thomas ;
Saliba, Walid ;
Arruda, Mauricio ;
Natale, Andrea .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (09) :868-874
[3]   Initial experience with remote catheter ablation using a novel magnetic navigation system -: Magnetic remote catheter ablation [J].
Ernst, S ;
Ouyang, FF ;
Linder, C ;
Hertting, K ;
Stahl, F ;
Chun, J ;
Hachiya, H ;
Bänsch, D ;
Antz, M ;
Kuck, KH .
CIRCULATION, 2004, 109 (12) :1472-1475
[4]   Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation [J].
Faddis, MN ;
Blume, W ;
Finney, J ;
Hall, A ;
Rauch, J ;
Sell, J ;
Bae, KT ;
Talcott, M ;
Lindsay, B .
CIRCULATION, 2002, 106 (23) :2980-2985
[5]   Magnetic guidance system for cardiac electrophysiology - A prospective trial of safety and efficacy in humans [J].
Faddis, MN ;
Chen, J ;
Osborn, J ;
Talcott, M ;
Cain, ME ;
Lindsay, BD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (11) :1952-1958
[6]   Remote controlled magnetically guided pulmonary vein isolation in canines [J].
Greenberg, S ;
Blume, W ;
Faddis, M ;
Finney, J ;
Hall, A ;
Talcott, M ;
Lindsay, B .
HEART RHYTHM, 2006, 3 (01) :71-76
[7]   Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci [J].
Haïssaguerre, M ;
Jaïs, P ;
Shah, DC ;
Garrigue, S ;
Takahashi, A ;
Lavergne, T ;
Hocini, M ;
Peng, JT ;
Roudaut, R ;
Clementy, J .
CIRCULATION, 2000, 101 (12) :1409-1417
[8]   Circular mapping and ablation of the pulmonary vein for treatment of atrial fibrillation - Impact of different catheter technologies [J].
Marrouche, NF ;
Dresing, T ;
Cole, C ;
Bash, D ;
Saad, E ;
Balaban, K ;
Pavia, SV ;
Schweikert, R ;
Saliba, W ;
Abdul-Karim, A ;
Pisano, E ;
Fanelli, R ;
Tchou, P ;
Natale, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (03) :464-474
[9]   Catheter ablation for paroxysmal atrial fibrillation - Segmental pulmonary vein ostial ablation versus left atrial ablation [J].
Oral, H ;
Scharf, C ;
Chugh, A ;
Hall, B ;
Cheung, P ;
Good, E ;
Veerareddy, S ;
Pelosi, F ;
Morady, F .
CIRCULATION, 2003, 108 (19) :2355-2360
[10]   Segmental ostial ablation to isolate the pulmonary veins during atrial fibrillation -: Feasibility and mechanistic insights [J].
Oral, H ;
Knight, BP ;
Özaydin, M ;
Chugh, A ;
Lai, SWK ;
Scharf, C ;
Hassan, S ;
Greenstein, R ;
Han, JD ;
Pelosi, F ;
Strickberger, SA ;
Morady, F .
CIRCULATION, 2002, 106 (10) :1256-1262