The dosimetric impact of daily setup error on target volumes and surrounding normal tissue in the treatment of prostate cancer with intensity-modulated radiation therapy

被引:11
作者
Algan, Ozer [1 ]
Jamgade, Ambarish [1 ]
Ali, Imad [1 ]
Christie, Alana [1 ]
Thompson, J. Spencer [1 ]
Thompson, David [1 ]
Ahmad, Salahuddin [1 ]
Herman, Terence [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol Biostat & Epidemiol, Oklahoma City, OK 73104 USA
关键词
Prostate cancer; IMRT; Setup error; Fiducial marker; CONFORMAL RADIOTHERAPY; LOCALIZATION; MOTION; CT;
D O I
10.1016/j.meddos.2012.03.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The purpose of this study was to evaluate the impact of daily setup error and interfraction organ motion on the overall dosimetric radiation treatment plans. Twelve patients undergoing definitive intensity-modulated radiation therapy (IMRT) treatments for prostate cancer were evaluated in this institutional review board-approved study. Each patient had fiducial markers placed into the prostate gland before treatment planning computed tomography scan. IMRT plans were generated using the Eclipse treatment planning system. Each patient was treated to a dose of 8100 cGy given in 45 fractions. In this study, we retrospectively created a plan for each treatment day that had a shift available. To calculate the dose, the patient would have received under this plan, we mathematically "negated" the shift by moving the isocenter in the exact opposite direction of the shift. The individualized daily plans were combined to generate an overall plan sum. The dose distributions from these plans were compared with the treatment plans that were used to treat the patients. Three-hundred ninety daily shifts were negated and their corresponding plans evaluated. The mean isocenter shift based on the location of the fiducial markers was 3.3 +/- 6.5 mm to the right, 1.6 +/- 5.1 mm posteriorly, and 1.0 +/- 5.0 mm along the caudal direction. The mean D95 doses for the prostate gland when setup error was corrected and uncorrected were 8228 and 7844 cGy (p < 0.002), respectively, and for the planning target volume (PTV8100) was 8089 and 7303 cGy (p < 0.001), respectively. The mean V95 values when patient setup was corrected and uncorrected were 99.9% and 87.3%, respectively, for the PTV8100 volume (p < 0.0001). At an individual patient level, the difference in the D95 value for the prostate volume could be >1200 cGy and for the PIV8100 could approach almost 2000 cGy when comparing corrected against uncorrected plans. There was no statistically significant difference in the D35 parameter for the surrounding normal tissue except for the dose received by the penile bulb and the right hip. Our dosimetric evaluation suggests significant underdosing with inaccurate target localization and emphasizes the importance of accurate patient setup and target localization. Further studies are needed to evaluate the impact of intrafraction organ motion, rotation, and deformation on doses delivered to target volumes. (C) 2012 American Association of Medical Dosimetrists.
引用
收藏
页码:406 / 411
页数:6
相关论文
共 14 条
[1]  
[Anonymous], 2010, Cancer facts and figures
[2]   PROSTATE MOTION DURING STANDARD RADIOTHERAPY AS ASSESSED BY FIDUCIAL MARKERS [J].
CROOK, JM ;
RAYMOND, Y ;
SALHANI, D ;
YANG, H ;
ESCHE, B .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :35-42
[3]   What dose of external-beam radiation is high enough for prostate cancer? [J].
Eade, Thomas N. ;
Hanlon, Alexandra L. ;
Horwitz, Eric M. ;
Buyyounouski, Mark K. ;
Hanks, Gerald E. ;
Pollack, Alan .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (03) :682-689
[4]   Dosimetric effects of patient rotational setup errors on prostate IMRT treatments [J].
Fu, Weihua ;
Yang, Yong ;
Li, Xiang ;
Heron, Dwight E. ;
Huq, M. Saiful ;
Yue, Ning J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (20) :5321-5331
[5]  
Kuban D.A., 2010, INT J RADIAT ONCOL, V77, P1
[6]   PRACTICAL METHOD OF ADAPTIVE RADIOTHERAPY FOR PROSTATE CANCER USING REAL-TIME ELECTROMAGNETIC TRACKING [J].
Olsen, Jeffrey R. ;
Noel, Camille E. ;
Baker, Kenneth ;
Santanam, Lakshmi ;
Michalski, Jeff M. ;
Parikh, Parag J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05) :1903-1911
[7]   The detectability and localization accuracy of implanted fiducial markers determined on in-room computerized tomography (CT) and electronic portal images (EPI) [J].
Owen, Rebecca ;
Kron, Tomas ;
Foroudi, Farshad ;
Cox, Jennifer ;
Zhu, Li ;
Cramb, Jim ;
Sparks, Laura ;
Duchesne, Gillian .
MEDICAL DOSIMETRY, 2008, 33 (03) :226-233
[8]  
Roach M 3rd, 1997, Radiat Oncol Investig, V5, P187, DOI 10.1002/(SICI)1520-6823(1997)5:4<187::AID-ROI4>3.3.CO
[9]  
2-B
[10]   Combined error of patient positioning variability and prostate motion uncertainty in 3D conformal radiotherapy of localized prostate cancer [J].
Rudat, V ;
Schraube, P ;
Oetzel, D ;
Zierhut, D ;
Flentje, M ;
Wannenmacher, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (05) :1027-1034