Time of flight measurements based on FPGA and SiPMs for PET-MR

被引:7
作者
Aguilar, Albert [1 ]
Garcia-Olcina, Raimundo [1 ]
Martinez, Pedro A. [1 ]
Martos, Julio [1 ]
Soret, Jesus [1 ]
Torres, Jose [1 ]
Benlloch, Jose M. [2 ]
Gonzalez, Antonio J. [2 ]
Sanchez, Filomeno [2 ]
机构
[1] Univ Valencia, Commun & Digital Syst Design Grp DSDC, E-46003 Valencia, Spain
[2] Univ Politecn Valencia, CSIC, CIEMAT, I3M, Valencia, Spain
关键词
Time of flight; Silicon photomultipliers; Time-to-digital converter; Field programable gate array; Positron emission tomography and magnetic resonance;
D O I
10.1016/j.nima.2013.09.008
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Coincidence time measurements with SiPMs have shown to be suitable for PET/MR systems. The present study is based on 3 x 3 mm(2) SiPMs, LSO crystals and a conditioning signal electronic circuit. A Constant Fraction Discriminator (CFD) is used to digitalize the signals and a TDC FPGA-implemented is employed for fine time measurements. TDC capability allows processing the arrival of multiple events simultaneously, measuring times under 100 ps. The complete set-up for time measurements results on a resolution of 892 +/- 41 ps for a pair of detectors. The details of such implementation are exposed and the trade-offs of each configuration are discussed. (C) 2013 Elsevier By, All rights reserved,
引用
收藏
页码:127 / 131
页数:5
相关论文
共 15 条
[1]  
Auffray E., 2011, P IEEE NSS, VN4-61
[2]  
Barral J., 2004, STUDY SILICON PHOTOM
[3]   Performance evaluation of SiPM photodetectors for PET imaging in the presence of magnetic fields [J].
Espana, S. ;
Fraile, L. M. ;
Herraiz, J. L. ;
Udias, J. M. ;
Desco, M. ;
Vaquero, J. J. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 613 (02) :308-316
[4]  
Favi E., 2009, P ACMSIGDA INT S FIE, P113, DOI 10.1145/1508128.1508145
[5]   Analog Circuit for Timing Measurements With Large Area SiPMs Coupled to LYSO Crystals [J].
Gola, Alberto ;
Piemonte, Claudio ;
Tarolli, Alessandro .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) :1296-1302
[6]  
Haselman M.D., 2010, DIGITAL PULSE TIMING, DOI 10.1.1.152.3344
[7]   Free-running ADC- and FPGA-based signal processing method for brain PET using GAPD arrays [J].
Hu, Wei ;
Choi, Yong ;
Hong, Key Jo ;
Rang, Jihoon ;
Jung, Jin Ho ;
Huh, Youn Sul ;
Lim, Hyun Keong ;
Kim, Sang Su ;
Kim, Byung-Tae ;
Chung, Yonghyun .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 664 (01) :370-375
[8]   A fast preamplifier concept for SiPM-based time-of-flight PET detectors [J].
Huizenga, J. ;
Seifert, S. ;
Schreuder, F. ;
van Dam, H. T. ;
Dendooven, P. ;
Lohner, H. ;
Vinke, R. ;
Schaart, D. R. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 695 :379-384
[9]   ERROR ANALYSIS AND DESIGN OF THE NUTT TIME-INTERVAL DIGITIZER WITH PICOSECOND RESOLUTION [J].
KALISZ, J ;
PAWLOWSKI, M ;
PELKA, R .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1987, 20 (11) :1330-1341
[10]   Multi-Pixel Photon Counters for TOF PET Detector and Its Challenges [J].
Kim, Chang Lyong ;
Wang, Gin-Chung ;
Dolinsky, Sergei .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) :2580-2585