Energetics of blood flow in Fontan circulation under VAD support

被引:9
作者
Telyshev, Dmitry [1 ,2 ]
Denisov, Maxim [1 ]
Markov, Aleksandr [2 ]
Fresiello, Libera [3 ]
Verbelen, Tom [3 ]
Selishchev, Sergey [1 ]
机构
[1] Natl Res Univ Elect Technol, Inst Biomed Syst, Zelenograd 124498, Russia
[2] IM Sechenov First Moscow State Med Univ, Inst Bion Technol & Engn, Moscow, Russia
[3] Katholieke Univ Leuven, Dept Cardiac Surg, Leuven, Belgium
基金
俄罗斯科学基金会;
关键词
flow hemodynamics; kinetic energy distribution; scalar shear stress; total cavopulmonary connection; ventricular assist device; IN-VITRO; CAVOPULMONARY CONNECTION; SURGICAL REPAIR; VENTRICLE; PUMP;
D O I
10.1111/aor.13564
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The need to simulate the normal operating conditions of the human body is the key factor in every study and engineering process of bioelectronic devices designed for implantation. The Fontan procedure is an example of such a process aimed to support the human body function. It is a standard treatment method for patients with a functionally univentricular heart. However, it has significant drawbacks such as overload of the only functional heart ventricle that often leads to the necessity of the heart transplantation. In this study, we analyze the total cavopulmonary connection (TCPC) influence on the blood with and without connected auxiliary blood circulation pump. We investigate four different types of TCPC configurations, analyze blood pressure and different flow rate, study the turbulent kinetic energy distribution, and evaluate hydraulic and power losses for various cases. Finally, we calculate volumetric scalar shear stresses distribution and demonstrate the high potential of TCPC configuration with connected rotary pump as a tool for the load redistribution in the functional heart ventricle. This work is particularly relevant for improving existing TCPCs' quality that can extend the life of Fontan patients. Moreover, it also applies to the reduction of morbidity and mortality of the patients waiting for the heart transplantation.
引用
收藏
页码:50 / 57
页数:8
相关论文
共 34 条
[31]   The Effect of Rotor Geometry on the H−Q Curves of the Sputnik Implantable Pediatric Rotary Blood Pump [J].
Telyshev D.V. ;
Denisov M.V. ;
Selishchev S.V. .
Biomedical Engineering, 2017, 50 (06) :420-424
[32]   The Progress in the Novel Pediatric Rotary Blood Pump Sputnik Development [J].
Telyshev, Dmitry ;
Denisov, Maxim ;
Pugovkin, Alexander ;
Selishchev, Sergey ;
Nesterenko, Igor .
ARTIFICIAL ORGANS, 2018, 42 (04) :432-443
[33]   Distribution of hepatic venous blood in the total cavo pulmonary connection: An in vitro study into the effects of connection geometry [J].
Walker, PG ;
Oweis, GF ;
Watterson, KG .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2001, 123 (06) :558-564
[34]   Cavopulmonary Support with a Microaxial Pump for the Failing Fontan Physiology [J].
Zhu, Jiaquan ;
Kato, Hideyuki ;
Fu, Yaqin Y. ;
Zhao, Lisa ;
Foreman, Celeste ;
Davey, Lisa ;
Weisel, Richard D. ;
Van Arsdell, Glen S. ;
Honjo, Osami .
ASAIO JOURNAL, 2015, 61 (01) :49-54