Attenuation correction in SPECT using consistency conditions for the exponential ray transform

被引:27
作者
Mennessier, C
Noo, F
Clackdoyle, R
Bal, G
Desbat, L
机构
[1] Univ Utah, Dept Radiol, Salt Lake City, UT 84132 USA
[2] Univ Liege, Inst Elect Montefiore, B-4000 Liege, Belgium
[3] Univ Grenoble 1, Grenoble, France
关键词
D O I
10.1088/0031-9155/44/10/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Using data consistency conditions for the exponential ray transform, a method is derived to correct SPECT data for attenuation effects. No transmission measurements are required, and no operator-defined contours are needed. Furthermore, any 3D parallel-ray geometry can be considered for SPECT data acquisition, even unconventional geometries which do not lead to a set of 2D parallel-beam sinograms. The method is presented for both the 2D parallel-beam geometry and a particular 3D case, called the rotating slant hole geometry. Full details of the algorithms are given. Implementation has been carried out and results are presented in 2D and in 3D using simulated data.
引用
收藏
页码:2483 / 2510
页数:28
相关论文
共 33 条
[1]   Range conditions for the exponential radon transform [J].
Aguilar, V ;
Ehrenpreis, L ;
Kuchment, P .
JOURNAL D ANALYSE MATHEMATIQUE, 1996, 68 :1-13
[2]   RANGE CONDITIONS FOR THE MULTIDIMENSIONAL EXPONENTIAL X-RAY TRANSFORM [J].
AGUILAR, V ;
KUCHMENT, P .
INVERSE PROBLEMS, 1995, 11 (05) :977-982
[3]  
BAILEY DL, 1987, J NUCL MED, V28, P844
[4]   APPROXIMATE RECONSTRUCTION OF ATTENUATION MAP IN SPECT IMAGING [J].
BRONNIKOV, AV .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (05) :1483-1488
[5]   NEW APPROACH TO THE EMISSION COMPUTERIZED TOMOGRAPHY PROBLEM - SIMULTANEOUS CALCULATION OF ATTENUATION AND ACTIVITY-COEFFICIENTS [J].
CENSOR, Y ;
GUSTAFSON, DE ;
LENT, A ;
TUY, H .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) :2775-2779
[6]   ASYMMETRIC FAN TRANSMISSION CT ON SPECT SYSTEMS [J].
CHANG, W ;
LONCARIC, S ;
HUANG, G ;
SANPITAK, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :913-928
[7]  
CLACK R, 1996, 1995 IEEE MED IM C, P1948
[8]  
DESBAT L, 1995, COMPUTERIZED TOMOGRAPHY, P124
[9]  
DICKEN V, 1998, THESIS U POTSDAM GER
[10]  
Gilland DR, 1998, J NUCL MED, V39, P1105