Endoleaks after Endovascular Abdominal Aortic Aneurysm Repair: Detection with Dual-Energy Dual-Source CT

被引:170
作者
Stolzmann, Paul [1 ]
Frauenfelder, Thomas [1 ]
Pfammatter, Thomas [1 ]
Peter, Nicole [1 ]
Scheffel, Hans [1 ]
Lachat, Mario [2 ]
Schmidt, Bernhard [3 ]
Marincek, Borut [1 ]
Alkadhi, Hatem [1 ]
Schertler, Thomas [1 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Cardiovasc Surg Clin, CH-8091 Zurich, Switzerland
[3] Siemens Med Solut, Forchheim, Germany
基金
瑞士国家科学基金会;
关键词
D O I
10.1148/radiol.2483080193
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the diagnostic performance of dual-energy dual-source computed tomography (CT) in the detection of endoleaks after endovascular abdominal aortic aneurysm (AAA) repair. Materials and Methods: This study was local ethics board approved, and written informed consent was obtained from all patients. One hundred eighteen patients ( 21 women, 97 men; mean age, 74 years +/- 8 [ standard deviation]) underwent follow-up dual-energy dual-source CT during the nonenhanced, arterial, and delayed phases after AAA repair. Delayed phase CT images were acquired in the dual-energy mode for reconstruction of virtual nonenhanced images. Two blinded and independent readers evaluated the data for the presence or absence of endoleaks during three reading sessions: Standard nonenhanced, arterial phase, and delayed phase images were read during session A; virtual nonenhanced, arterial phase, and delayed phase images, during session B; and virtual nonenhanced and delayed phase images, during session C. Sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV) were calculated, with the session A image data set as the reference standard. Radiation dose parameters were estimated. Results: Reading session A revealed that 52 (44%) of 118 patients had endoleaks. Overall sensitivity, specificity, NPV, and PPV for CT endoleak detection during sessions B and C were identical: 100%, 97%, 100%, and 96%, respectively. The accuracy of the session B and session C readings was not significantly different from that of the session A reading (P = .50). The effective radiation dose in the image acquisition protocol involving one dual-energy scan was significantly (P < .001) lower than the effective doses in the protocols involving standard triple-phase scanning (mean difference, 61%) and standard nonenhanced and delayed phase scanning (mean difference, 41%). Conclusion: Compared with standard protocols, one dual-energy dual-source CT scan performed during the delayed phase with reconstruction of virtual nonenhanced images enables detection of endoleaks after endovascular AAA repair with high accuracy and a considerably lower radiation dose. (c) RSNA, 2008
引用
收藏
页码:682 / 691
页数:10
相关论文
共 31 条
[1]   Emergency endovascular repair for ruptured abdominal aortic aneurysms: Feasibility and comparison of early results with conventional open repair [J].
Alsac, JM ;
Desgranges, P ;
Kobeiter, H ;
Becquemin, JP .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2005, 30 (06) :632-639
[2]   CLINICAL APPLICATION OF COMPTON AND PHOTOELECTRIC RECONSTRUCTION IN COMPUTED TOMOGRAPHY - PRELIMINARY-RESULTS [J].
AVRIN, DE ;
MACOVSKI, A ;
ZATZ, LM .
INVESTIGATIVE RADIOLOGY, 1978, 13 (03) :217-222
[3]   Endoluminal stent-grafts for infrarenal abdominal aortic aneurysms [J].
Blum, U ;
Voshage, G ;
Lammer, J ;
Beyersdorf, F ;
Tollner, D ;
Kretschmer, G ;
Spillner, G ;
Polterauer, P ;
Nagel, G ;
Holzenbein, T ;
Thurnher, S ;
Langer, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 336 (01) :13-20
[4]   Endovascular versus open surgical elective repair of infrarenal abdominal aortic aneurysm: Predictors of patient discharge destination [J].
Bosch, JL ;
Beinfeld, MT ;
Halpern, EF ;
Lester, JS ;
Gazelle, GS .
RADIOLOGY, 2001, 220 (03) :576-580
[5]   Estimated radiation risks potentially associated with full-body CT screening [J].
Brenner, DJ ;
Elliston, CD .
RADIOLOGY, 2004, 232 (03) :735-738
[6]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[7]   Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography [J].
Gerber, TC ;
Kuzo, RS ;
Morin, RL .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2005, 21 (01) :165-176
[8]   Helical CT of aorta after endoluminal stent-graft therapy: Value of biphasic acquisition [J].
Golzarian, J ;
Dussaussois, L ;
Abada, HT ;
Gevenois, PA ;
Van Gansbeke, D ;
Ferreira, J ;
Struyven, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (02) :329-331
[9]   Endovascular aortic stent-grafts: Transcatheter embolization of persistent perigraft leaks [J].
Golzarian, J ;
Struyven, J ;
Abada, HT ;
Wery, D ;
Dussaussois, L ;
Madani, A ;
Ferreira, J ;
Dereume, JP .
RADIOLOGY, 1997, 202 (03) :731-734
[10]  
Görich J, 1999, RADIOLOGY, V213, P767