Ambulatory Holter monitoring in asymptomatic patients with DDD pacemakers - do we need ACC/AHA Guidelines revision?

被引:6
作者
Chudzik, Michal [1 ]
Klimczak, Artur [1 ]
Wranicz, Jerzy Krzysztof [1 ]
机构
[1] Med Univ Lodz, Dept Electrocardiol, PL-91425 Lodz, Poland
关键词
DDD pacemaker; ambulatory Holter monitoring; pacemaker dysfunction; DUAL-CHAMBER PACEMAKERS; DEMAND PACEMAKERS; FIELD; INHIBITION; DIAGNOSIS; TELEMETRY; SIGNALS; SYSTEM; IMPACT;
D O I
10.5114/aoms.2013.38675
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: We sought to determine the usefulness of ambulatory 24-hour Hotter monitoring in detecting asymptomatic pacemaker (PM) malfunction episodes in patients with dual-chamber pacemakers whose pacing and sensing parameters were proper, as seen in routine post-implantation follow-ups. Material and methods: Ambulatory 24-hour Hotter recordings (HM) were performed in 100 patients with DDD pacemakers 1 day after the implantation. Only asymptomatic patients with proper pacing and sensing parameters (assessed on PM telemetry on the first day post-implantation) were enrolled in the study. The following parameters were assessed: failure to pace, failure to sense (both oversensing and undersensing episodes) as well as the percentage of all PM disturbances. Results: Despite proper sensing and pacing parameters, HM revealed PM disturbances in 23 patients out of 100 (23%). Atrial undersensing episodes were found in 12 patients (p < 0.005) with totally 963 episodes and failure to capture in 1 patient (1%). T wave oversensing was the most common ventricular channel disorder (1316 episodes in 9 patients, p < 0.0005). Malfunction episodes occurred sporadically, leading to pauses of up to 1.6 s or temporary bradycardia, which were, nevertheless, not accompanied by clinical symptoms. No ventricular pacing disturbances were found. Conclusions: Asymptomatic pacemaker dysfunction may be observed in nearly 25% of patients with proper DDD parameters after implantation. Thus, ambulatory HM during the early post-implantation period may be a useful tool to detect the need to reprogram PM parameters.
引用
收藏
页码:815 / 820
页数:6
相关论文
共 29 条
  • [1] BLEIFER S B, 1974, Progress in Cardiovascular Diseases, V16, P569, DOI 10.1016/0033-0620(74)90019-X
  • [2] BONNET IL, 1996, PACING CLIN ELECTROP, V19, P1724
  • [3] Impact and prevention of far-field sensing in fallback mode switches
    Bordacher, P
    Cazeau, S
    Graindorg, L
    Ritter, P
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 206 - 209
  • [4] FAR-FIELD QRS COMPLEX SENSING VIA THE ATRIAL PACEMAKER LEAD .1. MECHANISM, CONSEQUENCES, DIFFERENTIAL-DIAGNOSIS AND COUNTERMEASURES IN AAI AND VDD/DDD PACING
    BRANDT, J
    FAHRAEUS, T
    SCHULLER, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (10): : 1432 - 1438
  • [5] Far-field QRS complex sensing: Prevalence and timing with bipolar atrial leads
    Brandt, J
    Worzewski, T
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (03): : 315 - 320
  • [6] Analysis of atrial sensed far-field ventricular signals: A reassessment
    Brouwer, J
    Nagelkerke, D
    denHeijer, P
    Ruiter, JH
    Mulder, H
    Begemann, MJS
    Lie, KI
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (04): : 916 - 922
  • [7] THE CLINICAL INCIDENCE AND SIGNIFICANCE OF MYOPOTENTIAL SENSING WITH UNIPOLAR PACEMAKERS
    FETTER, J
    BOBELDYK, GL
    ENGMAN, FJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (05): : 871 - 881
  • [8] High rate atrial tachyarrhythmia detections in implantable pulse generators: Low incidence of false-positive detections
    Fitts, SM
    Hill, MRS
    Mehra, R
    Gillis, AM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (07): : 1080 - 1086
  • [9] Bipolar ventricular far-field signals in the atrium
    Fröhlig, G
    Helwani, Z
    Kusch, O
    Berg, M
    Schieffer, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (11): : 1604 - 1613
  • [10] FURMAN S, 1993, TELEMETRY PRACTICE C, P633