Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method

被引:72
作者
Kellman, Peter [1 ]
Herzka, Daniel A. [2 ]
Arai, Andrew E. [1 ]
Hansen, Michael Schacht [1 ]
机构
[1] NHLBI, NIH, DHHS, Bethesda, MD 20892 USA
[2] Johns Hopkins Univ, Sch Med, Baltimore, MD USA
基金
美国国家卫生研究院;
关键词
T1; map; Error; Off-resonance; MOLLI; Shim; SSFP; CARDIOVASCULAR MAGNETIC-RESONANCE; INVERSION-RECOVERY MOLLI; T1; VALIDATION;
D O I
10.1186/1532-429X-15-63
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts. Methods: The sensitivity of T1-estimates based on the SSFP based MOLLI sequence to errors in center frequency are calculated by means of a Bloch simulation and validated by phantom measurements. Typical off-resonance errors following local cardiac shimming are determined by field mapping at both 1.5 and 3.0T. In vivo examples demonstrate the artifactual appearance of T1-maps in the presence of off-resonance variation. Results: Off-resonance varied 61.8 +/- 15.5 Hz (mean +/- SD, n = 18) across the heart at 1.5T and 125.0 +/- 40.6 Hz (mean +/- SD, n = 18) at 3.0T. For T1 = 1000 ms, the variation in T1 due to off-resonance variation was approximately 20 ms at 62 Hz, and > 50 ms at 125 Hz. Conclusions: Regional variations due to the inability to completely shim the B0-field variation around the heart appear as regional variation in T1, which is artifactual.
引用
收藏
页数:8
相关论文
共 21 条
[1]  
Banypersad SM, 2013, JACC-CARDIOVASC IMAG, P00138, DOI [10.1016/j.jcmg.2012.11.013, DOI 10.1016/J.JCMG.2012.11.013.PII:S1936-878X]
[2]   Fundamentals of balanced steady state free precession MRI [J].
Bieri, Oliver ;
Scheffler, Klaus .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (01) :2-11
[3]  
Bull S, 2013, HEART
[4]   QUANTIFICATION OF T1 VALUES BY SNAPSHOT-FLASH NMR IMAGING [J].
DEICHMANN, R ;
HAASE, A .
JOURNAL OF MAGNETIC RESONANCE, 1992, 96 (03) :608-612
[5]   Reduction of transient signal oscillations in True-FISP using a linear flip angle series magnetization preparation [J].
Deshpande, VS ;
Chung, YC ;
Zhang, Q ;
Shea, SM ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :151-157
[6]   Equilibrium Contrast Cardiovascular Magnetic Resonance for the Measurement of Diffuse Myocardial Fibrosis Preliminary Validation in Humans [J].
Flett, Andrew S. ;
Hayward, Martin P. ;
Ashworth, Michael T. ;
Hansen, Michael S. ;
Taylor, Andrew M. ;
Elliott, Perry M. ;
McGregor, Christopher ;
Moon, James C. .
CIRCULATION, 2010, 122 (02) :138-U72
[7]   Modified Look-Locker T1 evaluation using Bloch simulations: Human and phantom validation [J].
Gai, Neville D. ;
Stehning, Christian ;
Nacif, Marcelo ;
Bluemke, David A. .
MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (02) :329-336
[8]   Robust Water/Fat Separation in the Presence of Large Field Inhomogeneities Using a Graph Cut Algorithm [J].
Hernando, Diego ;
Kellman, P. ;
Haldar, J. P. ;
Liang, Z-P. .
MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (01) :79-90
[9]   Adiabatic Inversion Pulses for Myocardial T1 Mapping [J].
Kellman, Peter ;
Herzka, Daniel A. ;
Hansen, Michael Schacht .
MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (04) :1428-1434
[10]   T1 and extracellular volume mapping in the heart: estimation of error maps and the influence of noise on precision [J].
Kellman, Peter ;
Arai, Andrew E. ;
Xue, Hui .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2013, 15