Safety interlock and vent system to alleviate potentially dangerous ice blockage of top-loading cryostat sample sticks

被引:2
作者
Pangelis, S. [1 ]
Olsen, S. R. [1 ]
Scherschligt, J. [2 ]
Leao, J. B. [2 ]
Pullen, S. A. [1 ]
Dender, D. [2 ]
Hester, J. R. [1 ]
Imperia, P. [1 ]
机构
[1] Australian Nucl Sci & Technol Org, Lucas Heights, NSW 2234, Australia
[2] NIST, Ctr Neutron Res, Gaithersburg, MD 20899 USA
关键词
closed cycle refrigerators; cryo-coolers; cryo-furnaces; cryogen free; cryostats; ice blockage; icing up; neutron scattering; safety interlocks; sample environment; sample sticks; top loading;
D O I
10.1107/S0021889813010868
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A combined solution is presented for minimizing the safety hazards associated with closed cycle cryostats described by Swainson & Cranswick [J. Appl. Cryst. (2010), 43, 206-210]. The initial solution is to install a vent tube with one open end deep inside the sample space and a pressure relief valve at the top. This solution works for either a cryogen or a cryogen-free (closed cycle) system. The second approach, which can be combined with the first and is applicable to cryogen-free cryostats, involves electrically interlocking the closed cycle refrigerator compressor to the sample space, so that the system cannot be cooled in the presence of a leak path to air.
引用
收藏
页码:1236 / 1239
页数:4
相关论文
共 5 条
[1]  
Norris M. A., 1997, CRYOG ENG C INT CRYO
[2]  
PARAMESWARAN VR, 1975, J GLACIOL, V14, P305
[3]  
Sumitomo Heavy Industries, 2003, OP MAN SRDK SER CRYO, P38
[4]   Safety considerations with top-loading cryostats, involving 'icing-up' of the sample space with solid air [J].
Swainson, Ian Peter ;
Cranswick, Lachlan Michael David .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2010, 43 :206-210
[5]   MEASUREMENTS CONCERNING THE VAPOUR-LIQUID EQUILIBRUM OF SOLUTIONS OF HE-3 IN HE-4 BELOW 2.19-DEGREES-K [J].
TACONIS, KW ;
BEENAKKER, JJM ;
NIER, AOC ;
ALDRICH, LT .
PHYSICA, 1949, 15 (8-9) :733-739