Dynamic shim updating on the human brain

被引:68
作者
Koch, Kevin M. [1 ]
McIntyre, Scott [1 ]
Nixon, Terence W. [1 ]
Rothman, Douglas L. [1 ]
de Graaf, Robin A. [1 ]
机构
[1] Yale Univ, Magnet Resonance Res Ctr, New Haven, CT 06520 USA
关键词
homogeneity; shimming; dynamic shimming; shim degeneracy;
D O I
10.1016/j.jmr.2006.03.007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Dynamic alteration of shim settings during a multi-slice imaging experiment can improve static magnetic-field homogeneity over extended volumes. In this report, a pre-emphasized dynamic shim updating (DSU) system capable of rapidly updating all non-degenerate zeroth through second-order shims is presented and applied to high-field multi-slice imaging studies on the human brain. DSU is utilized in both non-oblique and oblique slicing geometries while updating in-plane and through-slice shims. Image-based magnetic-field maps are used to quantify homogeneity improvements and comparisons are made on a slice-specific basis between static global shimming and increasing orders of shim inclusion utilized DSU. The influence of oblique slicing geometry on DSU-utilized global homogeneity is also quantified computationally. Finally, the effect of DSU on susceptibility artifact reduction in single-shot axial-sliced EPI is analyzed using experimental acquisitions. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:286 / 296
页数:11
相关论文
共 27 条
  • [1] Andersson J.L., 2004, P INT SOC MAGN RESON, V11, P87
  • [2] [Anonymous], 1999, CLASSICAL ELECTROMAG
  • [3] Dynamic shim updating: A new approach towards optimized whole brain shimming
    Blamire, AM
    Rothman, DL
    Nixon, T
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (01) : 159 - 165
  • [4] Chen NK, 1999, MAGNET RESON MED, V42, P807, DOI 10.1002/(SICI)1522-2594(199910)42:4<807::AID-MRM25>3.0.CO
  • [5] 2-8
  • [6] Dynamic shim updating (DSU) for multislice signal acquisition
    de Graaf, RA
    Brown, PB
    McIntyre, S
    Rothman, DL
    Nixon, TW
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (03) : 409 - 416
  • [7] A novel target-field method for magnetic resonance shim coils: III. Shielded zonal and tesseral coils
    Forbes, LK
    Crozier, S
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2003, 36 (02) : 68 - 80
  • [8] Goldstein H, 1980, CLASSICAL MECH
  • [9] AUTOMATIC, LOCALIZED INVIVO ADJUSTMENT OF ALL 1ST-ORDER AND 2ND-ORDER SHIM COILS
    GRUETTER, R
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (06) : 804 - 811
  • [10] HETHERINGTON HP, 2005, P 13 ISMRM