Time-over-threshold for pulse shape discrimination in a time-of-flight phoswich PET detector

被引:35
作者
Chang, Chen-Ming [1 ,2 ]
Cates, Joshua W. [2 ]
Levin, Craig S. [2 ,3 ,4 ,5 ]
机构
[1] Stanford Univ, Dept Appl Phys, Stanford, CA 94305 USA
[2] Stanford Univ, MIPS, Dept Radiol, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
[4] Stanford Univ, Dept Phys, Stanford, CA 94305 USA
[5] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
关键词
PET; phoswich; pulse shape discrimination; SiPM; time-of-flight (TOF); depth-of-interaction (DOI); time-over-threshold (TOT); POSITRON-EMISSION-TOMOGRAPHY; WHOLE-BODY PET; SCINTILLATION DETECTORS; TIMING PERFORMANCE; RESOLUTION; DEPTH; SCANNER; DOI; SYSTEM; TOF;
D O I
10.1088/1361-6560/62/1/258
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is well known that a PET detector capable of measuring both photon time-of-flight (TOF) and depth-of-interaction (DOI) improves the image quality and accuracy. Phoswich designs have been realized in PET detectors to measure DOI for more than a decade. However, PET detectors based on phoswich designs put great demand on the readout circuits, which have to differentiate the pulse shape produced by different crystal layers. A simple pulse shape discrimination approach is required to realize the phoswich designs in a clinical PET scanner, which consists of thousands of scintillation crystal elements. In this work, we studied time-over-threshold (ToT) as a pulse shape parameter for DOI. The energy, timing and DOI performance were evaluated for a phoswich detector design comprising 3 mm x 3 mm x 10 mm LYSO:Ce crystal optically coupled to 3 mm x 3 mm x 10 mm calcium co-doped LSO:Ce,Ca(0.4%) crystal read out by a silicon photomultiplier (SiPM). A DOI accuracy of 97.2% has been achieved for photopeak events using the proposed time-over-threshold (ToT) processing. The energy resolution without correction for SiPM non-linearity was 9.7 +/- 0.2% and 11.3 +/- 0.2% FWHM at 511 keV for LYSO and LSO crystal layers, respectively. The coincidence time resolution for photopeak events ranges from 164.6 ps to 183.1 ps FWHM, depending on the layer combinations. The coincidence time resolution for inter-crystal scatter events ranges from 214.6 ps to 418.3 ps FWHM, depending on the energy windows applied. These results show great promises of using ToT for pulse shape discrimination in a TOF phoswich detector since a ToT measurement can be easily implemented in readout electronics.
引用
收藏
页码:258 / 271
页数:14
相关论文
共 35 条
[1]   Waveform-Sampling Electronics for a Whole-Body Time-of-Flight PET Scanner [J].
Ashmanskas, W. J. ;
LeGeyt, B. C. ;
Newcomer, F. M. ;
Panetta, J. V. ;
Ryan, W. A. ;
Van Berg, R. ;
Wiener, R. I. ;
Karp, J. S. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (03) :1174-1181
[2]   A combined time-of-flight and depth-of-interaction detector for total-body positron emission tomography [J].
Berg, Eric ;
Roncali, Emilie ;
Kapusta, Maciej ;
Du, Junwei ;
Cherry, Simon R. .
MEDICAL PHYSICS, 2016, 43 (02) :939-950
[3]   Achieving fast timing performance with multiplexed SiPMs [J].
Bieniosek, M. F. ;
Cates, J. W. ;
Levin, C. S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (07) :2879-2892
[4]   Compact pulse width modulation circuitry for silicon photomultiplier readout [J].
Bieniosek, M. F. ;
Olcott, P. D. ;
Levin, C. S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (15) :5049-5059
[5]   DESIGN OF A HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPH USING SOLID-STATE SCINTILLATION DETECTORS [J].
CARRIER, C ;
MARTEL, C ;
SCHMITT, D ;
LECOMTE, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :685-690
[6]   Investigation of LSO crystals for high spatial resolution positron emission tomography [J].
Casey, ME ;
Eriksson, L ;
Schmand, M ;
Andreaco, MS ;
Paulus, M ;
Dahlbom, M ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1109-1113
[7]   Advances in coincidence time resolution for PET [J].
Cates, Joshua W. ;
Levin, Craig S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (06) :2255-2264
[8]   Analytical calculation of the lower bound on timing resolution for PET scintillation detectors comprising high-aspect-ratio crystal elements [J].
Cates, Joshua W. ;
Vinke, Ruud ;
Levin, Craig S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (13) :5141-5161
[9]   Performance characterization of compressed sensing positron emission tomography detectors and data acquisition system [J].
Chang, Chen-Ming ;
Grant, Alexander M. ;
Lee, Brian J. ;
Kim, Ealgoo ;
Hong, KeyJo ;
Levin, Craig S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (16) :6407-6421
[10]   Petiroc and Citiroc: front-end ASICs for SiPM read-out and ToF applications [J].
Fleury, J. ;
Callier, S. ;
de La Taille, C. ;
Seguin, N. ;
Thienpont, D. ;
Dulucq, F. ;
Ahmad, S. ;
Martin, G. .
JOURNAL OF INSTRUMENTATION, 2014, 9