Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods

被引:355
作者
Jaffrin, Michel Y. [1 ]
Morel, Helene [1 ]
机构
[1] Univ Technol Compiegne, Dept Biol Engn, F-60205 Compiegne, France
关键词
Body fluid resistivity; Electrical modelling of tissues; Extracellular fluid; Total body water;
D O I
10.1016/j.medengphy.2008.06.009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reviews various bioimpedance methods permitting to measure non-invasively, extracellular, intracellular and total body water (TBW) and compares BIA methods based oil empirical equations of the wrist-ankle resistance or impedance at 50 kHz, height and weight with BIS methods which rely on an electrical model of tissues and resistances measured at zero and infinite frequencies. In order to compare these methods, impedance measurements were made with a multifrequency Xitron 4200 impedancemeter on 57 healthy subjects which had undergone simultaneously it Dual X-ray absorptiometry examination (DXA), in order to estimate their TBW from their fat-free-mass. Extracellular (ECW) and TBW volumes were calculated for these Subjects using the original BIS method and modifications of Matthie [Matthie JR. Second generation mixture theory equation for estimating intracellular Water using bioimpedance spectroscopy. J Appl Physiol 2005;99:780-1], Jaffrin et al. [Jaffrin MY, Fenech M, Moreno MV, Kieffer R. Total body water measurement by a modification of the bioimpedance spectroscopy method. Med Bio Eng Comput 2006;44:873-82], Moissl et al. [Moissl UM, Wabel P, Chamney PW, Bosaeus I, Levin NW, et al. Body fluid volume determination via body composition spectroscopy in health and disease. Physiol Meas 2006;27:921-33] and their TBW resistivities were compared and discussed. ECW volumes were calculated by BIA methods of Sergi et al. [Sergi G, Bussolotto M, Perini P, Calliari I, et al. Accuracy of bioelectrical bioimpedance analysis for the assessment of extracellular space in healthy subjects and in fluid retention states. Ann Nutr Metab 1994;38(3):158-65] and Hannan et al. [Hannan WJ, Cowen SJ, Fearon KC, Plester CE, Falconer JS, Richardson RA. Evaluation of multi-frequency bio-impedance analysis for the assessment of extracellular and total body water in surgical patients. Clin Sci 1994;86:479-85] and TBW volumes by BIA methods of Kushner and Schoeller [Kushner RF, Schoeller DA. Estimation of total body water by bioelectrical impedance analysis. Am J Clin Nutr 1986;44(3):417-24], Lukaski et al. [Lukaski HC, Bolonchuk WW. Estimation of body fluid Volumes using tetrapolar bioelectrical impedance measurements. Aviat Space Environ Med 1988;59:1163-9], Hannan et al. [Hannan WJ, Cowen SJ, Fearon KC, Plester CE, Falconer JS, Richardson RA. Evaluation Of multi-frequency bio-impedance analysis for the assessment of extracellular and total body water in surgical patients. Clinical Science 1994;86:479-85], Deurenberg et al. [Deurenberg F. van der Koy K, Leenen R, Westrate JA, Seidell JC. Sex and age specific prediction formulas for estimating body composition from bioelectric impedance: a cross validation Study. Int J Obesity 1991; 15:17-25] These volumes were compared against those given by BIS method and, in the case of TBW, with those by DXA. For ECW, a good agreement was found between various BIS methods and that of Sergi while Hannan's values were higher. Both Matthie's and Moissl's methods gave mean TBW resistivities and Volumes lower than those of Jaffrin's and DXA methods. Kushner et al. method gave values of TBW not significantly different from those of Jaffrin et al. and DXA, as Hannan's method in men, but Lukaski and Deurenberg methods led to an underestimation. (C) 2008 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1257 / 1269
页数:13
相关论文
共 34 条
[1]  
Buchholz Andrea C, 2004, Nutr Clin Pract, V19, P433, DOI 10.1177/0115426504019005433
[2]  
Cole KS., 1972, Membranes, Ions and Impulses: A Chapter of Classical Biophysics
[3]   Predicting body cell mass with bioimpedance by using theoretical methods: A technological review [J].
DeLorenzo, A ;
Andreoli, A ;
Matthie, J ;
Withers, P .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (05) :1542-1558
[4]  
DEURENBERG P, 1991, INT J OBESITY, V15, P17
[5]   A comparison of bioimpedance methods for detection of body cell mass change in HIV infection [J].
Earthman, CP ;
Matthie, JR ;
Reid, PM ;
Harper, IT ;
Ravussin, E ;
Howell, WH .
JOURNAL OF APPLIED PHYSIOLOGY, 2000, 88 (03) :944-956
[6]   Human hydrometry:: comparison of multifrequency bioelectrical impedance with 2H2O and bromine dilution [J].
Ellis, KJ ;
Wong, WW .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 85 (03) :1056-1062
[7]  
Fenech M., 2003, Archives of Physiology and Biochemistry, V111, P21
[8]   Fluid volumes determination by impedance spectroscopy and hematocrit monitoring: Application to pediatric hemodialysis [J].
Fenech, M ;
Maasrani, M ;
Jaffrin, MY .
ARTIFICIAL ORGANS, 2001, 25 (02) :89-98
[9]   Dual-energy x-ray absorptiometry and body composition: Differences between devices and comparison with reference methods [J].
Genton, L ;
Hans, D ;
Kyle, UG ;
Pichard, C .
NUTRITION, 2002, 18 (01) :66-70
[10]  
Hanai T., 1968, EMULSION SCI