Incidence of glistenings with the latest generation of yellow-tinted hydrophobic acrylic intraocular lenses

被引:45
作者
Colin, Joseph [1 ]
Praud, Delphine [1 ,2 ]
Touboul, David [1 ]
Schweitzer, Cedric [1 ]
机构
[1] Ctr Hosp Univ Pellegrin, Serv Ophtalmol, F-33076 Bordeaux, France
[2] Ctr Invest Clin Epidemiol Clin, Bordeaux, France
关键词
IN-VITRO ANALYSIS; LIGHT-SCATTERING; VISUAL FUNCTION; SINGLE-PIECE; LONG-TERM;
D O I
10.1016/j.jcrs.2012.01.031
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the incidence of glistenings in a large series of consecutive eyes with a blue light-filtering hydrophobic acrylic intraocular lens (IOL) and the relationship with parameters such as age, sex, follow-up length, and corrected distance visual acuity (CDVA). SETTING: Service d'Ophtalmologie, Centre Hospitalier et Universitaire Bordeaux, Bordeaux, France. DESIGN: Cross-sectional study. METHODS: This study included 111 eyes of 74 patients (age range 33 to 86 years). All cases had cataract surgery with Acrysof SN60WF monofocal IOL implantation and had a routine postoperative examination between March 2011 and June 2011. The incidence and severity of glistenings were evaluated with the slitlamp. Also evaluated was their relationship with age, sex, follow-up length, IOL power, and CDVA. Glistenings were subjectively graded (0 = absent; 1 = moderate; 2 = dense). RESULTS: Glistenings occurred in 96 eyes (86.5%). Glistenings were of grade 1 severity in 45 eyes (40.5%) and of grade 2 severity in 51 eyes (45.9%). The follow-up was significantly longer in eyes with grade 2 glistenings (P <=.01). A limited, but significant, correlation was found between glistening severity and length of follow-up (r = 0.32, P<.01). Although there was a trend toward decreased visual acuities at higher glistening grades (r = 0.22, P=.01), there were no significant differences in CDVA between the glistening severity groups (P=.14). CONCLUSIONS: Glistenings were common in eyes with the blue light-filtering hydrophobic acrylic IOL and increased over time.
引用
收藏
页码:1140 / 1146
页数:7
相关论文
共 22 条
[1]   Glistenings in the AcrySof intraocular lens: Pilot study [J].
Christiansen, G ;
Durcan, FJ ;
Olson, RJ ;
Christiansen, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) :728-733
[2]   Glistenings in the Artiflex phakic intraocular lens [J].
Cisneros-Lanuza, Angel ;
Hurtado-Sarrio, Mercedes ;
Duch-Samper, Antonio ;
Gallego-Pinazo, Roberto ;
Menezo-Rozalen, Jose L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (08) :1405-1408
[3]   Glistenings on Intraocular Lenses in Healthy Eyes: Effects and Associations [J].
Colin, Joseph ;
Orignac, Isabelle .
JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (12) :869-875
[4]   Glistenings in a large series of hydrophobic acrylic intraocular lenses [J].
Colin, Joseph ;
Orignac, Isabelle ;
Touboul, David .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (12) :2121-2126
[5]   Clinical performance of Alcon SA30AL and SA60AT single-piece acrylic intraocular lenses [J].
Davison, JA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1112-1123
[6]   Visual significance of glistenings seen in the AcrySof intraocular lens [J].
Dhaliwal, DK ;
Mamalis, N ;
Olson, RJ ;
Crandall, AS ;
Zimmerman, P ;
Alldredge, OC ;
Durcan, FJ ;
Omar, O .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (04) :452-457
[7]   Optical and atomic force microscopy of an explanted AcrySof intraocular lens with glistenings [J].
Dogru, M ;
Tetsumoto, K ;
Tagami, Y ;
Kato, K ;
Nakamae, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (04) :571-575
[8]   In vitro comparison of glistening formation among hydrophobic acrylic intraocular lenses [J].
Gregori, NZ ;
Spencer, TS ;
Mamalis, N ;
Olson, RJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1262-1268
[9]   Effects on visual function of glistenings and folding marks in AcrySof intraocular lenses [J].
Gunenc, U ;
Oner, FH ;
Tongal, S ;
Ferliel, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (10) :1611-1614
[10]   Glistening formation in an AcrySof lens initiated by spinodal decomposition of the polymer network by temperature change [J].
Kato, K ;
Nishida, M ;
Yamane, H ;
Nakamae, K ;
Tagami, Y ;
Tetsumoto, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (09) :1493-1498