Modification of a commercial gas chromatography isotope ratio mass spectrometer for on-line carbon isotope dilution: Evaluation of its analytical characteristics for the quantification of organic compounds

被引:4
作者
Alonso Sobrado, Laura [1 ]
Robledo Fernandez, Mario [1 ]
Cueto Diaz, Sergio [1 ]
Ruiz Encinar, Jorge [1 ]
Garcia Alonso, J. Ignacio [1 ]
机构
[1] Univ Oviedo, Dept Phys & Analyt Chem, E-33006 Oviedo, Spain
关键词
Gas chromatography; Combustion; Mass spectrometry; On-line isotope dilution; Compound-independent calibration; OPTIMIZATION; INJECTION;
D O I
10.1016/j.chroma.2015.09.071
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe the instrumental modification of a commercial gas chromatography isotope ratio mass spectrometer (GC-IRMS) and its application for on-line carbon isotope dilution. The main modification consisted in the addition of a constant flow of enriched (CO2)-C-13 diluted in helium after the chromatographic column through the splitter holder located inside the chromatographic oven of the instrument. In addition, and in contrast to the conventional mode of operation of GC-IRMS instruments where the signal at m/z 45 is amplified 100-fold with respect to the signal at m/z 44, the same signal amplification was used in both Faraday cups at m/z 44 and 45. Under these conditions isotope ratio precision for the ratio 44/45 was around 0.05% RSD (n=50). The evaluation of the instrument was performed with mixtures of organic compounds including 11 n-alkanes, 16 PAHs, 12 PCBs and 3 benzothiophenes. It was observed that compounds of very different boiling points could be analysed without discrimination in the injector when a Programmable Temperature Vaporizer (PTV) injector was employed. Moreover, the presence of heteroatoms (Cl or S) in the structure of the organic compounds did not affect their combustion efficiency and therefore the trueness of the results. Quantitative results obtained for all the analytes assayed were excellent in terms of precision (<3% RSD) and accuracy (average relative error <= 4%) and what is more important using a single and simple generic internal standard for quantification. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:99 / 108
页数:10
相关论文
共 23 条
[1]   Some unique properties of gas chromatography coupled with atomic-emission detection [J].
Andersson, JT .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 373 (06) :344-355
[2]  
[Anonymous], 2004, MODERN PRACTICE GAS, DOI DOI 10.1002/0471651141
[3]  
[Anonymous], 2005, EFSA J, V3, P284, DOI DOI 10.2903/J.EFSA.2005.263
[4]  
[Anonymous], 1997, 61619 EN INT EL COMM
[5]   Determination of the elemental composition of hydrocarbon mixtures by gas chromatography with atomic emission detection: Increasing the accuracy [J].
Chernetsova, ES ;
Revelskii, AI ;
Durst, D ;
Revelskii, IA .
JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 60 (09) :855-859
[6]   Gas Chromatography-Combustion-Mass Spectrometry with Postcolumn Isotope Dilution for Compound-Independent Quantification: Its Potential to Assess HS-SPME Procedures [J].
Cueto Diaz, Sergio ;
Ruiz Encinar, Jorge ;
Sanz-Medel, Alfredo ;
Garcia Alonso, J. Ignacio .
ANALYTICAL CHEMISTRY, 2010, 82 (16) :6862-6869
[7]   A Quantitative Universal Detection System for Organic Compounds in Gas Chromatography with Isotopically Enriched 13CO2 [J].
Cueto Diaz, Sergio ;
Ruiz Encinar, Jorge ;
Sanz-Medel, Alfredo ;
Garcia Alonso, J. Ignacio .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2009, 48 (14) :2561-2564
[8]  
Dettmer-Wilde K., 2014, PRACTICAL GAS CHROMA
[9]  
EPA US, 2014, TOX PRIOR POLL
[10]   A multivariate study of the programmed temperature vaporization injection-gas chromatographic-mass spectrometric determination of polycyclic aromatic hydrocarbons -: Application to marine sediments analysis [J].
Fernandez-Gonzalez, V. ;
Concha-Grana, E. ;
Muniategui-Lorenzo, S. ;
Lopez-Mahia, P. ;
Prada-Rodriguez, D. .
TALANTA, 2008, 74 (05) :1096-1103