Screening of the most relevant parameters for method development in ultra-high,performance hydrophilic interaction chromatography

被引:69
作者
Periat, Aurelie [1 ]
Debrus, Benjamin [1 ]
Rudaz, Serge [1 ]
Guillarme, Davy [1 ]
机构
[1] Univ Lausanne, Univ Geneva, Sch Pharmaceut Sci, CH-1211 Geneva 4, Switzerland
关键词
HILIC; Guidelines for method development; Principal component analysis; Column chemistry; Retention and selectivity behaviour; UHPLC; INTERACTION LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; REVERSED-PHASE; STATIONARY PHASES; SELECTIVITY; SEPARATION; SILICA; RETENTION; MODE; RICH;
D O I
10.1016/j.chroma.2013.01.037
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The goal of the present work was to provide some guidelines for method development in hydrophilic interaction chromatography (HILIC). For this purpose, a training set of 82 representative pharmaceutical compounds possessing diverse polarity and including acidic, basic and neutral properties was analyzed. All these drugs were injected on five short HILIC columns packed with sub-2 mu m particles and dedicated for UHPLC (ultra-high performance liquid chromatography) operation. Four different pH conditions ranging from pH 3 to 6 were tested at two ionic strengths (10 and 50 mmol/L) and finally, the reference organic modifier in HILIC, namely acetonitrile was modified with small amounts of methanol or isopropanol. From these experiments and using multivariate data analysis, it is clear that the stationary phase was the most relevant parameters for tuning selectivity in HILIC, since the types of interactions (i.e. dipole-dipole, hydrogen bonding and ion exchange) with analytes strongly vary between columns. Among the selected phases, the diol phase was the less interesting one, in terms of selectivity and peak shape. The zwitterionic phase was attractive, as it allowed a better retention of acidic compounds. Finally, the bare silica phase was the most versatile HILIC column packed with sub-2 mu m particles in terms of retention, peak shape and selectivity. Mobile phase pH was the other important parameter to achieve an appropriate selectivity and retention, even if it remains always difficult to assess precisely the mobile phase pH, analyte pK(a) and silanols pK(a), when working with more 70% acetonitrile. Finally, buffer ionic strength and organic modifier nature could be considered as secondary parameters for HILIC method development. In conclusion, screening four different columns packed with sub-2 mu m particles at two mobile phase pH, using a fast gradient seems to be a good generic approach for initial HILIC method development. The total time for such a screening was estimated at similar to 1 h, including reequilibration. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 83
页数:12
相关论文
共 29 条
  • [1] HYDROPHILIC-INTERACTION CHROMATOGRAPHY FOR THE SEPARATION OF PEPTIDES, NUCLEIC-ACIDS AND OTHER POLAR COMPOUNDS
    ALPERT, AJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 499 : 177 - 196
  • [2] Bicker W., 2008, J SEP SCI, V2971, P31
  • [3] Knowledge discovery in metabolomics: An overview of MS data handling
    Boccard, Julien
    Veuthey, Jean-Luc
    Rudaz, Serge
    [J]. JOURNAL OF SEPARATION SCIENCE, 2010, 33 (03) : 290 - 304
  • [4] Evaluation of various HILIC materials for the fast separation of polar compounds
    Chauve, Benedicte
    Guillarme, Davy
    Cleon, Philippe
    Veuthey, Jean-Luc
    [J]. JOURNAL OF SEPARATION SCIENCE, 2010, 33 (6-7) : 752 - 764
  • [5] Approach to hydrophilic interaction chromatography column selection: Application to neurotransmitters analysis
    Chirita, Raluca-Ioana
    West, Caroline
    Finaru, Adriana-Luminita
    Elfakir, Claire
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (18) : 3091 - 3104
  • [6] Development and validation of a hydrophilic interaction chromatography-mass spectrometry assay for taurine and methionine in matrices rich in carbohydrates
    de Person, M
    Hazotte, A
    Elfakir, C
    Lafosse, M
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1081 (02) : 174 - 181
  • [7] HILIC methods in pharmaceutical analysis
    Dejaegher, Bieke
    Heyden, Yvan Vander
    [J]. JOURNAL OF SEPARATION SCIENCE, 2010, 33 (6-7) : 698 - 715
  • [8] Determination of N-desethylamodiaquine by hydrophilic interaction liquid chromatography with tandem mass spectrometry:: Application to in vitro drug metabolism studies
    Dravid, Prajakta V.
    Frye, Reginald F.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 863 (01): : 129 - 134
  • [9] The impact of extra-column band broadening on the chromatographic efficiency of 5 cm long narrow-bore very efficient columns
    Fekete, Szabolcs
    Fekete, Jeno
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (31) : 5286 - 5291
  • [10] Effects of extra-column band spreading, liquid chromatography system operating pressure, and column temperature on the performance of sub-2-μm porous particles
    Fountain, Kenneth J.
    Neue, Uwe D.
    Grumbach, Eric S.
    Diehl, Diane M.