We have applied the synthetic-aperture method to linear-scanning microwave-induced thermoacoustic tomography in biological tissues. A nonfocused ultrasonic transducer was used to receive thermoacoustic signals, to which the delay-and-sum algorithm was applied for image reconstruction. We greatly improved the lateral resolution of images and acquired a clear view of the circular boundaries of buried cylindrical objects, which could not be obtained in conventional linear-scanning microwave-induced thermoacoustic tomography based on focused transducers. Two microwave sources, which had frequencies of 9 and 3 GHz, respectively, were used in the experiments for comparison. The 3 GHz system had a much larger imaging depth but a lower signal-noise ratio than the 9 GHz system in near-surface imaging. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2427 / 2431
页数:5
相关论文
共 13 条
[1]  
[Anonymous], 1986, Medical Applications of Microwave Imaging
[2]   NUMERICAL ALGORITHM FOR DIELECTRIC-PERMITTIVITY MICROWAVE IMAGING OF INHOMOGENEOUS BIOLOGICAL BODIES [J].
CAORSI, S ;
FRATTONI, A ;
GRAGNANI, GL ;
NORTINO, E ;
PASTORINO, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (06) :NS37-NS44
[3]  
CHAUDHARY SS, 1984, INDIAN J BIOCHEM BIO, V21, P76
[4]   MICROWAVE IMAGING OF TISSUE BLOOD CONTENT CHANGES [J].
HAWLEY, MS ;
BROQUETAS, A ;
JOFRE, L ;
BOLOMEY, JC ;
GABORIAUD, G .
JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (03) :197-202
[5]   Photoacoustic blood cell detection and imaging of blood vessels in phantom tissue [J].
Hoelen, CGA ;
Pongers, R ;
Hamhuis, G ;
de Mul, FFM ;
Greve, J .
OPTICAL AND IMAGING TECHNIQUES FOR BIOMONITORING III, PROCEEDINGS OF, 1998, 3196 :142-153
[6]   Three-dimensional photoacoustic imaging of blood vessels in tissue [J].
Hoelen, CGA ;
de Mul, FFM ;
Pongers, R ;
Dekker, A .
OPTICS LETTERS, 1998, 23 (08) :648-650
[7]   THE MEASURED ELECTRICAL-PROPERTIES OF NORMAL AND MALIGNANT HUMAN TISSUES FROM 50 TO 900 MHZ [J].
JOINES, WT ;
ZHANG, Y ;
LI, CX ;
JIRTLE, RL .
MEDICAL PHYSICS, 1994, 21 (04) :547-550
[8]   Backward mode detection of laser-induced wide-band ultrasonic transients with optoacoustic transducer [J].
Karabutov, AA ;
Savateeva, EV ;
Podymova, NB ;
Oraevsky, AA .
JOURNAL OF APPLIED PHYSICS, 2000, 87 (04) :2003-2014
[9]   PHOTOACOUSTIC ULTRASOUND (PAUS) - RECONSTRUCTION TOMOGRAPHY [J].
KRUGER, RA ;
LIU, PY ;
FANG, YR ;
APPLEDORN, CR .
MEDICAL PHYSICS, 1995, 22 (10) :1605-1609
[10]   Scanning thermoacoustic tomography in biological tissue [J].
Ku, G ;
Wang, LHV .
MEDICAL PHYSICS, 2000, 27 (05) :1195-1202