Robust Phased Array Non-Contact Vital Signs Monitoring in an Office Cubicle Setting

被引:0
|
作者
Hall, T. [1 ]
Tsay, J. [1 ]
Dominguez, G. [1 ]
Boothby, A. [1 ]
Das, V. [1 ]
Lopez, J. [1 ]
Nguyen, T. [2 ]
Banister, R. E. [2 ]
Lie, D. Y. C. [2 ]
机构
[1] Texas Tech Univ, Dept Elect & Comp Engn, Lubbock, TX 79409 USA
[2] Texas Tech Univ, Hlth Sci Ctr, Lubbock, TX 79409 USA
关键词
active antenna; beam-steering; Doppler Radar; e-health; homecare; noncontact vital signs monitoring; phased array; telemedicine; vital signs monitoring; remote monitoring;
D O I
暂无
中图分类号
TU [建筑科学];
学科分类号
0813 ;
摘要
Doppler-based non-contact vital signs (NCVS) sensor systems can monitor the heart and respiration rates without touching the patient, but the accuracy of the NCVS sensing can be degraded considerably by background clutters and movements artifacts from the monitored subject or other motions in the measurement environment. We have, therefore, developed a high directivity phased-array antenna NCVS system that can significantly increase the sensing accuracy and the monitoring range in a typical office cubicle setting. Depending on the position of the beam spot illuminated on the chest, our custom-made beam-steerable phased array NCVS system at 2.4GHz has demonstrated the accuracy of the heart rate monitoring to be within +/- 2 bpm (beat-per-minute) for 92.5%-99.2% of the time at 0.75m. As the monitoring range is extended to 1-1.5m with beam directly illuminating on the center of the chest, the heart rate monitoring accuracy remains within +/- 5 bpm above 78% of time. With more improvement, this phased array NCVS system should benefit patients considerably for telemedicine and remote monitoring, as the patient's vital signs can be monitored comfortably and continuously in both home and office settings, eventually capable of detecting potential sleep apnea and other respiratory or cardiovascular symptoms in real-time.
引用
收藏
页数:4
相关论文
共 50 条
  • [31] Non-Contact Automatic Vital Signs Monitoring of Infants in a Neonatal Intensive Care Unit Based on Neural Networks
    Khanam, Fatema-Tuz-Zohra
    Perera, Asanka G.
    Al-Naji, Ali
    Gibson, Kim
    Chahl, Javaan
    JOURNAL OF IMAGING, 2021, 7 (08)
  • [32] Simultaneous measurements of multiple vital signs using non-contact frequency modulated continuous wave radar monitoring
    Al Ahmad, Mahmoud
    Olule, Lillian Joyce Among
    ALEXANDRIA ENGINEERING JOURNAL, 2023, 71 : 609 - 617
  • [33] Sparsity-Based Multi-Person Non-Contact Vital Signs Monitoring via FMCW Radar
    Eder, Yonathan
    Eldar, Yonina C.
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2023, 27 (06) : 2806 - 2817
  • [34] Non-Contact Monitoring of Human Vital Signs Using FMCW Millimeter Wave Radar in the 120 GHz Band
    Lv, Wenjie
    He, Wangdong
    Lin, Xipeng
    Miao, Jungang
    SENSORS, 2021, 21 (08)
  • [35] Microphone Array for Non-contact Monitoring of Rolling Bearings
    Kudriashov, Volodymyr
    Ivanov, Vladislav
    Alexiev, Kiril
    Koprinkova-Hristova, Petia
    ADVANCED COMPUTING IN INDUSTRIAL MATHEMATICS, 2017, 681 : 103 - 114
  • [36] A Novel Microwave Measurement Technique for Non-contact Vital Sign Monitoring
    Sinharay, Arijit
    Das, Rajat
    Seth, Sayan
    2018 IEEE SENSORS, 2018, : 1515 - 1518
  • [37] Estimating vital signs through non-contact video-based approaches: A Survey
    Sinhal, Ruchika
    Singh, Kavita
    Shankar, Anuraj
    2017 INTERNATIONAL CONFERENCE ON RECENT INNOVATIONS IN SIGNAL PROCESSING AND EMBEDDED SYSTEMS (RISE), 2017, : 139 - 141
  • [38] 'No Touch' Vitals: A Pilot Study of Non-contact Vital Signs Acquisition in Exercising Volunteers
    Capraro, Geoffrey
    Etebari, Cameron
    Luchette, Katherine
    Mercurio, Laura
    Merck, Derek
    Kirenko, Ihor
    van Zon, Kees
    Bartula, Marek
    Rocque, Mukul
    Kobayashi, Leo
    2018 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS): ADVANCED SYSTEMS FOR ENHANCING HUMAN HEALTH, 2018, : 631 - 634
  • [39] A non-contact lie detector using Radar Vital Signs Monitor (RVSM) technology
    Geisheimer, J
    Greneker, EF
    IEEE AEROSPACE AND ELECTRONIC SYSTEMS MAGAZINE, 2001, 16 (08) : 10 - 14
  • [40] A Printed Dipole Array Antenna for Non-contact Monitoring System
    Liang, Qiancheng
    Mpanda, Ramadhani Selemani
    Wang, Xin
    Shi, Jingjing
    Xu, Lisheng
    2018 CROSS STRAIT QUAD-REGIONAL RADIO SCIENCE AND WIRELESS TECHNOLOGY CONFERENCE (CSQRWC), 2018,