Navigation system for robot-assisted intra-articular lower-limb fracture surgery

被引:45
作者
Dagnino, Giulio [1 ]
Georgilas, Ioannis [1 ]
Kohler, Paul [1 ]
Morad, Samir [1 ]
Atkins, Roger [2 ]
Dogramadzi, Sanja [1 ]
机构
[1] Univ West England, Bristol Robot Lab, Coldharbour Lane, Bristol BS16 1QY, Avon, England
[2] Bristol Royal Infirm & Gen Hosp, Upper Maudlin St, Bristol BS2 8HW, Avon, England
基金
美国国家卫生研究院;
关键词
Medical robotics; Fracture surgery; Computer-assisted surgery; Fracture reduction planning; Image guidance; 3D medical imaging; DISTAL FEMUR; REDUCTION;
D O I
10.1007/s11548-016-1418-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the surgical treatment for lower-leg intra-articular fractures, the fragments have to be positioned and aligned to reconstruct the fractured bone as precisely as possible, to allow the joint to function correctly again. Standard procedures use 2D radiographs to estimate the desired reduction position of bone fragments. However, optimal correction in a 3D space requires 3D imaging. This paper introduces a new navigation system that uses pre-operative planning based on 3D CT data and intra-operative 3D guidance to virtually reduce lower-limb intra-articular fractures. Physical reduction in the fractures is then performed by our robotic system based on the virtual reduction. 3D models of bone fragments are segmented from CT scan. Fragments are pre-operatively visualized on the screen and virtually manipulated by the surgeon through a dedicated GUI to achieve the virtual reduction in the fracture. Intra-operatively, the actual position of the bone fragments is provided by an optical tracker enabling real-time 3D guidance. The motion commands for the robot connected to the bone fragment are generated, and the fracture physically reduced based on the surgeon's virtual reduction. To test the system, four femur models were fractured to obtain four different distal femur fracture types. Each one of them was subsequently reduced 20 times by a surgeon using our system. The navigation system allowed an orthopaedic surgeon to virtually reduce the fracture with a maximum residual positioning error of (translational) and (rotational). Correspondent physical reductions resulted in an accuracy of 1.03 +/- A 0.2 mm and , when the robot reduced the fracture. Experimental outcome demonstrates the accuracy and effectiveness of the proposed navigation system, presenting a fracture reduction accuracy of about 1 mm and , and meeting the clinical requirements for distal femur fracture reduction procedures.
引用
收藏
页码:1831 / 1843
页数:13
相关论文
共 28 条
[1]  
Bellemans J., 2005, Total Knee Arthroplasty: A Guide To Get Better Performance, DOI [10.1007/3-540-27658-0_52, DOI 10.1007/3-540-27658-0_52]
[2]   Computer-assisted fracture reduction: a new approach for repositioning femoral fractures and planning reduction paths [J].
Buschbaum, Jan ;
Fremd, Rainer ;
Pohlemann, Tim ;
Kristen, Alexander .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2015, 10 (02) :149-159
[3]  
Dagnino G, 2015, EMBC 2015 C
[4]  
Dagnino G, 2015, CAOS 2015 C
[5]   Vision-based real-time position control of a semi-automated system for robot-assisted joint fracture surgery [J].
Dagnino, Giulio ;
Georgilas, Ioannis ;
Tarassoli, Payam ;
Atkins, Roger ;
Dogramadzi, Sanja .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (03) :437-455
[6]   Computer-Assisted Planning and Navigation for Corrective Distal Radius Osteotomy, Based on Pre- and Intraoperative Imaging [J].
Dobbe, J. G. G. ;
Strackee, S. D. ;
Schreurs, A. W. ;
Jonges, R. ;
Carelsen, B. ;
Vroemen, J. C. ;
Grimbergen, C. A. ;
Streekstra, G. J. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (01) :182-190
[7]   Recovery of knee function following fracture of the tibial plateau [J].
Gaston, P ;
Will, EM ;
Keating, JF .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (09) :1233-1236
[8]  
Gebhard F, 2015, DISTAL FEMUR FRACTUR
[9]  
Georgilas I, 2015, EMBC 2015 C, DOI [10.1109/EMBC.2015.7319491, DOI 10.1109/EMBC.2015.7319491]
[10]   Design of a parallel long bone fracture reduction robot with planning treatment tool [J].
Graham, A. E. ;
Xie, S. Q. ;
Aw, K. C. ;
Xu, W. L. ;
Mukherjee, S. .
2006 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS, VOLS 1-12, 2006, :1255-1260