Cone-beam image reconstruction from equi-angular sampling using spherical harmonics

被引:1
作者
Taguchi, K [1 ]
Zeng, GSL [1 ]
Gullberg, GT [1 ]
机构
[1] Toshiba Co Ltd, Ctr Res & Dev, Med Syst Co, Otawara, Tochigi 3248550, Japan
来源
MEDICAL IMAGING: 2001: IMAGE PROCESSING, PTS 1-3 | 2001年 / 4322卷
关键词
image reconstruction; cone-beam; spherical harmonics; short-object problem; equi-angular sampling;
D O I
10.1117/12.431166
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
All of exact cone-beam reconstruction algorithms(6-8) for so called "long-object problem" do not use equi-angular sampling but use equi-space sampling. However, cylindrical detectors (equi-angular sampling in xy and equi-spatial in z) have advantage in their compact design. Therefore, toward the long-object problem with equi-angular sampling, the purpose of this study is to develop a cone-beam reconstruction algorithm using equi-angular sampling for short-object problem. A novel implementation of Grangeat's algorithm(9) using equi-angular sampling has been developed for short-object problem with and without detector truncation. First, both cone-beam projection g(psi) (theta,phi) and the first derivative of plane integral (3D Radon transform) (p) over tilde (<(<psi>)over bar>) (theta,phi) are described using spherical harmonics with equi-angular sampling. Then, using Grangeat's formula, relationship between coefficients of spherical harmonics for g(<(<psi>)over bar>) (theta,phi) and (p) over tilde (<(<psi>)over bar>) (theta,phi) are found. Finally, a method has been developed to obtain (p) over tilde (<(<psi>)over bar>) (theta,phi) from cone-beam projection data in which the object is partially scanned. Images are reconstructed using the 3D Radon backprojection with rebinning. Computer simulations were performed in order to verify this approach: Isolated (axially bounded) objects were scanned both with circular and helical orbits. When the orbit of the cone vertex does not satisfy Tuy's data sufficiency conditions, strong oblique shadows and blurring in the axial direction were shown in reconstructed coronal images. On the other hand, if the trajectory satisfies Tuy's data sufficiency condition, the proposed algorithm provides an exact reconstruction. In conclusion, a novel implementation of the Grangeat's algorithm for cone-beam image reconstruction using equi-angular sampling has been developed.
引用
收藏
页码:867 / 876
页数:4
相关论文
共 50 条
  • [21] Three dimensional image reconstruction from cone-beam projections along arc-arc trajectory
    Emil Sidky
    中国体视学与图像分析, 2005, (03) : 129 - 133
  • [22] Efficient reconstruction from truncated data in circular cone-beam CT
    Wang Xian-Chao
    Yan Bin
    Liu Hong-Kui
    Li Lei
    Wei Xing
    Hu Guo-En
    ACTA PHYSICA SINICA, 2013, 62 (09)
  • [23] Image reconstruction on virtual PI-lines using filtered-backprojection in circular cone-beam CT
    Yu, LF
    Xia, D
    Zou, Y
    Pan, XC
    Medical Imaging 2005: Image Processing, Pt 1-3, 2005, 5747 : 2094 - 2099
  • [24] Derivative-Hilbert-Backprojection based image reconstruction from truncated projections in helical cone-beam CT
    张峰
    闫镔
    李磊
    席晓琦
    江桦
    魏星
    张翔
    崔进鲜
    Nuclear Science and Techniques, 2015, 26 (02) : 49 - 57
  • [25] Derivative-Hilbert-Backprojection based image reconstruction from truncated projections in helical cone-beam CT
    Zhang Feng
    Yan Bin
    Li Lei
    Xi Xiao-Qi
    Jiang Hua
    Wei Xing
    Zhang Xiang
    Cui Jin-Xian
    NUCLEAR SCIENCE AND TECHNIQUES, 2015, 26 (02)
  • [26] Arc based cone-beam reconstruction algorithm using an equal weighting scheme
    Nett, Brian E.
    Zhuang, Ting-Liang
    Leng, Shuai
    Chen, Guang-Hong
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2007, 15 (01) : 19 - 48
  • [27] Filtered backprojection formula for exact image reconstruction from cone-beam data along a general scanning curve
    Ye, YB
    Wang, G
    MEDICAL PHYSICS, 2005, 32 (01) : 42 - 48
  • [28] Analytic image reconstruction from partial data for a single-scan cone-beam CT with scatter correction
    Min, Jonghwan
    Pua, Rizza
    Kim, Insoo
    Han, Bumsoo
    Cho, Seungryong
    MEDICAL PHYSICS, 2015, 42 (11) : 6625 - 6640
  • [29] RECONSTRUCTION OF 3D TUBULAR STRUCTURES FROM CONE-BEAM PROJECTIONS
    Li, Jia
    Cohen, Laurent D.
    2011 8TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, 2011, : 1162 - 1166
  • [30] Reconstruction from Truncated Projections in Cone-beam CT using an Efficient 1D Filtering
    Xia, Yan
    Maier, Andreas
    Hofmann, Hannes G.
    Dennerlein, Frank
    Mueller, Kerstin
    Hornegger, Joachim
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668