Targeted analysis with benchtop quadrupole-orbitrap hybrid mass spectrometer: Application to determination of synthetic hormones in animal urine

被引:57
作者
Kumar, Praveen [1 ]
Rubies, Antoni [2 ]
Centrich, Francesc [2 ,3 ]
Granados, Merce [1 ]
Cortes-Francisco, Nuria [4 ]
Caixach, Josep [4 ]
Companyo, Ramon [1 ]
机构
[1] Univ Barcelona, Dept Quim Analit, E-08028 Barcelona, Spain
[2] Lab Agencia Salut Publ Barcelona, Barcelona, Spain
[3] CIBER Epidemiol & Publ Hlth CIBERESP, Madrid, Spain
[4] IDAEA CSIC, Mass Spectrometry Lab Organ Pollutants, Barcelona, Spain
关键词
Q Exactive; Synthetic hormones; Animal urine; Food safety control; HIGH-RESOLUTION; LIQUID; CONFIRMATION;
D O I
10.1016/j.aca.2013.04.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sensitive and unequivocal determination of analytes/contaminants in complex matrices is a challenge in the field of food safety control. In this study, various acquisition modes (Full MS/AIF, Full MS+tMS/MS, Full MS/dd MS/MS and tSIM/ddMS/MS) and parameters of a quadrupole-orbitrap hybrid mass spectrometer (Q Exactive) were studied in detail. One of the main conclusions has been that, reducing the scan range for Full MS (using the quadrupole) and targeted modes give higher signal-to-noise (S/N) ratios and thereby better detection limits for analytes in matrix. The use of Q Exactive in a complex case, for the confirmatory analysis of hormones in animal urine is presented. A targeted SIM data dependent MS/MS (tSIM/ddMS/MS) acquisition method for determination of eight synthetic hormones (trenbolone, 17 alpha ethinylestradiol, zeranol, stanozolol, dienestrol, diethylstilbestrol, hexestrol, taleranol) and a naturally occurring hormone (zearalenone) in animal urine were optimized to have sensitive precursors from targeted SIM mode and trigger MS/MS scans over the entire chromatograph peak. The method was validated according to EC/657/2002. CC alpha (decision limit) for the analytes ranged between 0.11 mu g L-1 and 0.69 mu g L-1 and CC beta (detection capability) ranged between 0.29 mu g L-1 and 0.90 mu g L-1. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 20 条
[1]  
[Anonymous], 2002, Off. J. Eur. Comm, VL221, P8
[2]  
[Anonymous], 2012, PROP DRAFT GUID PERF
[3]  
[Anonymous], AMANI PAPERS
[4]   New determination of the electron's mass -: art. no. 011603 [J].
Beier, T ;
Häffner, H ;
Hermanspahn, N ;
Karshenboim, SG ;
Kluge, HJ ;
Quint, W ;
Stahl, S ;
Verdú, J ;
Werth, G .
PHYSICAL REVIEW LETTERS, 2002, 88 (01) :4
[5]   Preventing false negatives with high-resolutionmass spectrometry: the benzophenone case [J].
Gallart-Ayala, H. ;
Nunez, O. ;
Moyano, E. ;
Galceran, M. T. ;
Martins, C. P. B. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (20) :3161-3166
[6]   Post-interface signal suppression, a phenomenon observed in a single-stage Orbitrap mass spectrometer coupled to an electrospray interfaced liquid chromatograph [J].
Kaufmann, A. ;
Widmer, M. ;
Maden, K. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2010, 24 (14) :2162-2170
[7]   Are liquid chromatography/electrospray tandem quadrupole fragmentation ratios unequivocal confirmation criteria? [J].
Kaufmann, Anton ;
Butcher, Patrick ;
Maiden, Kathryn ;
Widmer, Mirjam ;
Giles, Kevin ;
Uria, Diana .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (07) :985-998
[8]   Full Scan MS in Comprehensive Qualitative and Quantitative Residue Analysis in Food and Feed Matrices: How Much Resolving Power is Required? [J].
Kellmann, Markus ;
Muenster, Helmut ;
Zomer, Paul ;
Mol, Hans .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2009, 20 (08) :1464-1476
[9]   Optimized Fast and Sensitive Acquisition Methods for Shotgun Proteomics on a Quadrupole Orbitrap Mass Spectrometer [J].
Kelstrup, Christian D. ;
Young, Clifford ;
Lavallee, Richard ;
Nielsen, Michael L. ;
Olsen, Jesper V. .
JOURNAL OF PROTEOME RESEARCH, 2012, 11 (06) :3487-3497
[10]  
Kind Tobias, 2010, Bioanal Rev, V2, P23, DOI 10.1007/s12566-010-0015-9