Analysis and Spectral Assignments of Mixed Actinide Oxide Samples Using Laser-Induced Breakdown Spectroscopy (LIBS)

被引:46
作者
Barefield, James E., II [1 ]
Judge, Elizabeth J. [1 ]
Berg, John M. [1 ]
Willson, Stephen P. [1 ]
Le, Loan A. [1 ]
Lopez, Leon N. [1 ]
机构
[1] Los Alamos Natl Lab, Los Alamos, NM 87545 USA
关键词
Laser-induced breakdown spectroscopy; LIES; Mixed actinide oxides; Analysis and spectral assignments; EMISSION-SPECTROSCOPY; URANIUM;
D O I
10.1366/12-06830
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, we report for the first time the identification and assignments of complex atomic emission spectra of mixed actinide oxides using laser-induced plasma spectroscopy or laser-induced breakdown spectroscopy (LIES). Preliminary results of LIES measurements on samples of uranium dioxide (UO2)/plutonium dioxide (PuO2) and UO2/PuO2/americium dioxide (AmO2)/neptunium dioxide (NpO2) simulated fuel pellets (or mixed actinide oxide samples) are reported and discussed. We have identified and assigned >800 atomic emission lines for a UO2/PuO2/AmO2/NpO2 fuel pellet thus far. The identification and assignments of spectral emission lines for U, Pu, and Am are consistent with wavelength data from the literature. However, only a few emission lines have been assigned with a high degree of confidence for Np compared with atomic emission data from the literature. We also indicate where atomic emission lines for Cm would most likely appear in the spectral regions shown. Finally, we demonstrate that a LIES system with a resolving power of approximately 20 000 is adequate for analyzing complex mixtures of actinide elements within the same sample.
引用
收藏
页码:433 / 440
页数:8
相关论文
共 18 条
[1]  
Barefield II J.E., 1987, ATOMIC EMISSION SPEC
[2]  
Blaise J., 1984, ANL8395
[3]   Detection of Uranium Using Laser-Induced Breakdown Spectroscopy [J].
Chinni, Rosemarie C. ;
Cremers, David A. ;
Radziemski, Leon J. ;
Bostian, Melissa ;
Navarro-Northrup, Claudia .
APPLIED SPECTROSCOPY, 2009, 63 (11) :1238-1250
[4]   Monitoring Uranium, Hydrogen, and Lithium and Their Isotopes Using a Compact Laser-Induced Breakdown Spectroscopy (LIBS) Probe and High-Resolution Spectrometer [J].
Cremers, David A. ;
Beddingfield, Alan ;
Smithwick, Robert ;
Chinni, Rosemarie C. ;
Jones, C. Randy ;
Beardsley, Burt ;
Karch, Larry .
APPLIED SPECTROSCOPY, 2012, 66 (03) :250-261
[5]   PRELIMINARY TERM ANALYSIS OF AM-I AND AM-II SPECTRA [J].
FRED, M ;
TOMKINS, FS .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1957, 47 (12) :1076-1087
[6]   ICP AES ACTINIDE DETECTION LIMITS [J].
HUFF, EA ;
BOWERS, DL .
APPLIED SPECTROSCOPY, 1990, 44 (04) :728-729
[7]  
Judge E.J., 2011, 1107097 LOS AL NAT L, P1
[8]  
Judge EdwinA., 2012, 10 of New Documents Illustrating Early Christianity, P1
[9]   Quantitative determination of uranium and europium in glass matrix by laser-induced breakdown spectroscopy [J].
Jung, E. C. ;
Lee, D. H. ;
Yun, J. -I. ;
Kim, J. G. ;
Yeon, J. W. ;
Song, K. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2011, 66 (9-10) :761-764
[10]  
Kurucz R., 1995 ATOMIC LINE DAT