In vitro estimations of in vivo jet nebulizer efficiency using actual and simulated tidal breathing patterns

被引:24
作者
Bosco, AP
Rhem, RG
Dolovich, MB
机构
[1] McMaster Univ, Fac Hlth Sci, Dept Med, Hamilton, ON L8N 3Z5, Canada
[2] McMaster Univ, Fac Hlth Sci, Dept Nucl Med, Hamilton, ON L8N 3Z5, Canada
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2005年 / 18卷 / 04期
关键词
nebulizers; aerosols; breath simulation; corticosteroids; in vitro model; cascade impaction; and laser diffraction;
D O I
10.1089/jam.2005.18.427
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
In vivo aerosol delivery efficiency was estimated in vitro for two jet nebulizers using a breath monitor (Breathe!; Pari GmbH, Germany) and breath simulator (COMPAS (R); Pari GmbH) to reproduce subject tidal breathing patterns. The AeroEclipse (Trudell Medical International, Canada), a breath-actuated nebulizer, and the LC Star (Pari GmbH), a breath-enhanced nebulizer, were filled with levalbuterol HCI solution (Sepracor, USA) and operated with compressed O-2 at 8 lpm. Tidal breathing patterns of 20 adult subjects were digitally recorded with the Breathe! Breath Monitor. Subjects then breathed tidally from each nebulizer separately for 1 minute and to nebulizer dryness. Levalbuterol aerosol collected on filters placed between the nebulizer and mouth was chemically assayed to determine the inspired mass (IM), wasted mass (WM) and total emitted mass (TM). Measurements were repeated using the COMPAS (R) Breath Simulator to simulate each subject's tidal breathing pattern. IM, WM, and TM measurements using actual versus simulated tidal breathing were highly comparable for each nebulizer, except the IM (p < 0.05) from LC Star measured at nebulizer dryness. Breath simulation was an inaccurate tool for estimating the time to nebulizer dryness as simulated measurements to nebulizer dryness took significantly longer than measurements preformed with actual tidal breathing (p < 0.001). While breath simulation provides an accurate in vitro tool for estimating in vivo aerosol delivery, it should not completely replace in vivo measurements until inherent limitations in simulator operation can be overcome to provide a more clinically realistic simulation.
引用
收藏
页码:427 / 438
页数:12
相关论文
共 29 条
[1]   A phase I study of aerosolized administration of tgAAVCF to cystic fibrosis subjects with mild lung disease [J].
Aitken, ML ;
Moss, RB ;
Waltz, DA ;
Dovey, ME ;
Tonelli, MR ;
McNamara, SC ;
Gibson, RL ;
Ramsey, BW ;
Carter, BJ ;
Reynolds, TC .
HUMAN GENE THERAPY, 2001, 12 (15) :1907-1916
[2]  
ARCHER A, 1998, EUR J RESP J, V12, P68
[3]   The output of budesonide from nebulizers [J].
Barry, PW ;
O'Callaghan, C .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1998, 102 (02) :321-322
[4]   Drug output from nebulizers is dependent on the method of measurement [J].
Barry, PW ;
O'Callaghan, C .
EUROPEAN RESPIRATORY JOURNAL, 1998, 12 (02) :463-466
[5]  
Callaghan CO, 1997, THORAX, V52, P31
[6]   Do sinusoidal models of respiration accurately reflect the respiratory events of patients breathing on nebulizers? [J].
Coates, AL ;
Macneish, CF ;
Allen, PD ;
Ho, SL ;
Lands, LC .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1999, 12 (04) :265-273
[7]  
COCHRANE GM, 1997, EUR RESPIR REV, V7, P383
[8]   A review of issues relating to nebulizer standards [J].
Dennis, JH .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1998, 11 :S73-S79
[9]   DRUG NEBULIZER DESIGN AND PERFORMANCE - BREATH ENHANCED JET VS CONSTANT OUTPUT JET VS ULTRASONIC [J].
DENNIS, JH .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1995, 8 (03) :277-280
[10]  
DENYER J, 1997, J AEROSOL MED, V10, P99