A Comparison of Precision of Fit, Rotational Freedom, and Torque Loss with Copy-Milled Zirconia and Prefabricated Titanium Abutments

被引:39
作者
Alikhasi, Marzieh [1 ,2 ]
Monzavi, Abbas [3 ,4 ]
Bassir, Seyed Hossein [1 ]
Naini, Roshanak Baghai [5 ]
Khosronedjad, Nasim [6 ]
Keshavarz, Sareh [6 ]
机构
[1] Univ Tehran Med Sci, Sch Dent, Dent Res Ctr, Tehran, Iran
[2] Univ Tehran Med Sci, Sch Dent, Dept Prosthodont & Implants, Tehran, Iran
[3] Univ Tehran Med Sci, Sch Dent, Dent Laser Res Ctr, Tehran, Iran
[4] Univ Tehran Med Sci, Sch Dent, Dept Prosthodont, Tehran, Iran
[5] Univ So Calif, Lab Immunoregulat & Tissue Engn, Div Periodontol Diagnost Sci & Dent Hyg, Los Angeles, CA USA
[6] Tabriz Univ Med Sci, Tabriz, Iran
关键词
copy milling; dental implant-abutment interface; fit; rotational freedom; torque; zirconia; IMPLANT SOFT-TISSUE; MISFIT; RESTORATIONS; ALUMINA; CROWNS;
D O I
10.11607/jomi.2937
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The present study sought to compare copy-milled zirconia abutments with standard prefabricated titanium abutments in terms of horizontal and vertical misfit, rotational freedom, and torque loss. Materials and Methods: Twenty regular-platform implants (Branemark, Nobel Biocare) with a standard external-hexagon implant-abutment interface were divided into two groups. In one group, prefabricated titanium abutments (Easy Abutment, Nobel Biocare) were used; in the other group, custom-made zirconia abutments were fabricated using the Zirkonzahn milling system. Measurements of geometry, rotational freedom, and marginal discrepancy were performed using a video measuring machine. To measure torque loss, each abutment was torqued to 32 Ncm and then loosened with a digital torque wrench. The torque loss values were reported as a percentage of the applied torque. Results: Rotational freedom and horizontal misfit were significantly higher with the customized abutments. There were no statistically significant differences in vertical misfit or in mean percentage of torque loss between the two groups. Conclusions: Within the limitations of this study, it could be concluded that copy-milled zirconia abutments do not fit as accurately as prefabricated titanium abutments. However, the less precise fit of customized zirconia abutments does not seem to affect their ability to maintain the applied torque.
引用
收藏
页码:996 / 1002
页数:7
相关论文
共 40 条
[1]   Fracture Resistance of Yttria-Stabilized Zirconia Dental Implant Abutments [J].
Adatia, Nimet D. ;
Bayne, Stephen C. ;
Cooper, Lyndon F. ;
Thompson, Jeffery Y. .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2009, 18 (01) :17-22
[2]   Comparison of Fit Accuracy between Procera® Custom Abutments and Three Implant Systems [J].
Alves da Cunha, Tiago de Morais ;
Correia de Araujo, Roberto Paulo ;
Barbosa da Rocha, Paulo Vicente ;
Pazos Amoedo, Rosa Maria .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 (06) :890-895
[3]  
Andersson B, 1998, INT J PROSTHODONT, V11, P212
[4]   Marginal Adaptation of All-Ceramic Crowns on Implant Abutments [J].
Att, Wael ;
Hoischen, Tomas ;
Gerds, Thomas ;
Strub, Joerg Rudolf .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2008, 10 (04) :218-225
[5]  
Barbosa Gustavo Augusto Seabra, 2008, Braz. Dent. J., V19, P358, DOI 10.1590/S0103-64402008000400013
[6]  
Binon P P, 1996, Implant Dent, V5, P235, DOI 10.1097/00008505-199600540-00002
[7]  
Binon PP, 1996, INT J PROSTHODONT, V9, P149
[8]  
Binon PP, 1996, INT J PROSTHODONT, V9, P511
[9]  
Binon PP, 2000, INT J ORAL MAX IMPL, V15, P76
[10]   Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation [J].
Butz, F ;
Heydecke, G ;
Okutan, M ;
Strub, JR .
JOURNAL OF ORAL REHABILITATION, 2005, 32 (11) :838-843