The accuracy of a glucose-oxidase-based point-of-care glucometer in premature infants

被引:3
作者
Tsao, Lon-Yen [1 ,4 ]
Chang, Ming-Yeh [2 ]
Hsiao, Chien-Chou [1 ,3 ,4 ]
机构
[1] Changhua Christian Hosp, Dept Neonatol, Changhua 500, Taiwan
[2] Changhua Christian Hosp, Dept Pediat Neurol, Changhua 500, Taiwan
[3] Chung Shan Med Univ, Sch Med, Taichung, Taiwan
[4] Changhua Christian Hosp, Clin Nutr Support Serv Team, Changhua 500, Taiwan
来源
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION | 2013年 / 98卷 / 06期
关键词
Neonatology; Paediatric Practice; Intensive Care; Measurement; BLOOD-GLUCOSE;
D O I
10.1136/archdischild-2012-303613
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective The purpose of this study was to evaluate the accuracy of a novel commercial glucose-oxidase (GO)-based glucometer in measuring glucose concentrations in arterial blood samples collected from premature infants. Patients and methods A commercial glucometer, which is a plasma separation GO-based glucometer, was employed to compare the accuracy of the commercial blood glucometer with that of a standard glucose analyser for monitoring blood glucose levels in premature infants. Arterial blood samples were collected and analysed, including samples adjusted for specific glucose (3.6mmol/L, 8.3mmol/L and 16.7mmol/L) and PaO2 levels (range, 40-400mmHg). Results In total, 159 samples were collected. Bland-Altman analysis showed good correlation between the commercial glucometer and standard glucose analyser. Error-grid analysis revealed that all of the results were within zone A (ie, the clinically accurate estimate zone). The biases between the two systems were low at different PaO2 levels and haematocrits. Finally, the influence of different PaO2 levels was within acceptable ranges. Conclusions The GO-based glucometer evaluated in this study provides accurate results even when measured at high PaO2 and different haematocrits and is suitable for measuring arterial glucose levels in premature infants.
引用
收藏
页码:F545 / F548
页数:4
相关论文
共 48 条
[31]   Large-scale performance evaluation of Accu-Chek inform II point-of-care glucose meters [J].
Jeong, Tae-Dong ;
Cho, Eun-Jung ;
Ko, Dae-Hyun ;
Lee, Woochang ;
Chun, Sail ;
Hong, Ki-Sook ;
Min, Won-Ki .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2016, 76 (08) :657-663
[32]   Impact of High-Dose Intravenous Vitamin C for Treatment of Sepsis on Point-of-Care Blood Glucose Readings [J].
Howell, Anna Peyton ;
Parrett, Jenna L. ;
Malcom, Daniel R. .
JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2021, 15 (02) :309-316
[33]   A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood [J].
Wei, Huiping ;
Lan, Fang ;
He, Qitian ;
Li, Haiwei ;
Zhang, Fuyong ;
Qin, Xue ;
Li, Shan .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (03)
[34]   Point-of-care ultrasound for neonatal central catheter positioning: impact on X-rays and line tip position accuracy [J].
Rossi, Serena ;
Jogeesvaran, K. Haran ;
Matu, Eugen ;
Khan, Hammad ;
Grande, Elisabetta ;
Meau-Petit, Virginie .
EUROPEAN JOURNAL OF PEDIATRICS, 2022, 181 (05) :2097-2108
[35]   Point-of-care ultrasound for neonatal central catheter positioning: impact on X-rays and line tip position accuracy [J].
Serena Rossi ;
K. Haran Jogeesvaran ;
Eugen Matu ;
Hammad Khan ;
Elisabetta Grande ;
Virginie Meau-Petit .
European Journal of Pediatrics, 2022, 181 :2097-2108
[36]   Nanomaterial-assisted wearable glucose biosensors for noninvasive real-time monitoring: Pioneering point-of-care and beyond [J].
Safarkhani, Moein ;
Aldhaher, Abdullah ;
Heidari, Golnaz ;
Zare, Ehsan Nazarzadeh ;
Warkiani, Majid Ebrahimi ;
Akhavan, Omid ;
Huh, YunSuk ;
Rabiee, Navid .
NANO MATERIALS SCIENCE, 2024, 6 (03) :263-283
[37]   Elevated prehospital point-of-care glucose is associated with worse neurologic outcome after out-of-hospital cardiac arrest [J].
Abramson, Tiffany M. ;
Bosson, Nichole ;
Whitfield, Denise ;
Gausche-Hill, Marianne ;
Niemann, James T. .
RESUSCITATION PLUS, 2022, 9
[38]   Glucose and Lactate Reference Intervals for White Sturgeon and Evaluation of Two Point-of-Care Devices in Sturgeon Infected with Veronaea botryosa [J].
Hagen, Katharina B. ;
Lima, Kelly Marie ;
Ang, June ;
Montealegre-Golcher, Federico ;
Alonso, Flavio H. ;
Soto, Esteban .
JOURNAL OF AQUATIC ANIMAL HEALTH, 2022, 34 (01) :28-37
[39]   Accuracy and reliability of the i-STAT point-of-care device for the determination of haemoglobin concentration before and after major blood loss [J].
Ng, W. L. ;
Short, T. G. ;
Gunn, K. N. ;
Fuge, G. S. ;
Slon, B. .
ANAESTHESIA AND INTENSIVE CARE, 2014, 42 (04) :495-499
[40]   Integration of Solution-Based Assays onto Lateral Flow Device for One-Step Quantitative Point-of-Care Diagnostics Using Personal Glucose Meter [J].
Zhang, JingJing ;
Shen, Zhe ;
Xiang, Yu ;
Lu, Yi .
ACS SENSORS, 2016, 1 (09) :1091-1096