The accuracy of a glucose-oxidase-based point-of-care glucometer in premature infants

被引:3
|
作者
Tsao, Lon-Yen [1 ,4 ]
Chang, Ming-Yeh [2 ]
Hsiao, Chien-Chou [1 ,3 ,4 ]
机构
[1] Changhua Christian Hosp, Dept Neonatol, Changhua 500, Taiwan
[2] Changhua Christian Hosp, Dept Pediat Neurol, Changhua 500, Taiwan
[3] Chung Shan Med Univ, Sch Med, Taichung, Taiwan
[4] Changhua Christian Hosp, Clin Nutr Support Serv Team, Changhua 500, Taiwan
来源
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION | 2013年 / 98卷 / 06期
关键词
Neonatology; Paediatric Practice; Intensive Care; Measurement; BLOOD-GLUCOSE;
D O I
10.1136/archdischild-2012-303613
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective The purpose of this study was to evaluate the accuracy of a novel commercial glucose-oxidase (GO)-based glucometer in measuring glucose concentrations in arterial blood samples collected from premature infants. Patients and methods A commercial glucometer, which is a plasma separation GO-based glucometer, was employed to compare the accuracy of the commercial blood glucometer with that of a standard glucose analyser for monitoring blood glucose levels in premature infants. Arterial blood samples were collected and analysed, including samples adjusted for specific glucose (3.6mmol/L, 8.3mmol/L and 16.7mmol/L) and PaO2 levels (range, 40-400mmHg). Results In total, 159 samples were collected. Bland-Altman analysis showed good correlation between the commercial glucometer and standard glucose analyser. Error-grid analysis revealed that all of the results were within zone A (ie, the clinically accurate estimate zone). The biases between the two systems were low at different PaO2 levels and haematocrits. Finally, the influence of different PaO2 levels was within acceptable ranges. Conclusions The GO-based glucometer evaluated in this study provides accurate results even when measured at high PaO2 and different haematocrits and is suitable for measuring arterial glucose levels in premature infants.
引用
收藏
页码:F545 / F548
页数:4
相关论文
共 48 条
  • [11] Evaluation of a point-of-care blood glucose monitor in healthy goats
    Quandt, Jane E.
    Barletta, Michele
    Cornell, Karen K.
    Giguere, Steeve
    Hofmeister, Erik H.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2018, 28 (01) : 45 - 53
  • [12] Clinical Reliability of Point-of-Care Glucose Testing in Critically Ill Patients
    Pilackas, Kylie
    El-Oshar, Seraj
    Carter, Chris
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2020, 14 (01): : 65 - 69
  • [13] The application of personal glucose meters as universal point-of-care diagnostic tools
    Lisi, Fabio
    Peterson, Joshua R.
    Gooding, J. Justin
    BIOSENSORS & BIOELECTRONICS, 2020, 148 (148)
  • [14] Evaluation of a point-of-care glucose meter for general use in complex tertiary care facilities
    Chan, Pak Cheung R.
    Rozmanc, Mary
    Selden-Long, Isolde
    Kwan, Jason
    CLINICAL BIOCHEMISTRY, 2009, 42 (10-11) : 1104 - 1112
  • [15] Point-of-care testing for measuring haemolymph glucose in invertebrates is not a valid method
    Principe, Silas C.
    Augusto, Alessandra
    Costa, Tania M.
    CONSERVATION PHYSIOLOGY, 2019, 7
  • [16] Reliability of point-of-care testing for glucose measurement in critically ill adults
    Kanji, S
    Buffie, J
    Hutton, B
    Bunting, PS
    Singh, A
    McDonald, K
    Fergusson, D
    McIntyre, LA
    Hebert, PC
    CRITICAL CARE MEDICINE, 2005, 33 (12) : 2778 - 2785
  • [17] Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit
    Tendl, Kristina A.
    Christoph, Juergen
    Bohn, Adele
    Herkner, Kurt R.
    Pollak, Arnold
    Prusa, Andrea-Romana
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (09) : 1747 - 1754
  • [18] Comparing the accuracy of point-of-care with laboratory (capillary, venous, and arterial) blood glucose levels in critically ill patients with and without shock
    Abdulaziz Alshaer
    Basma A. Badgheish
    Zahra Hashim Alsadah
    Khalid Sewify
    Sarah Alghazal
    Sarah Alzahrani
    Abeer Qadi
    Reham Alqahtani
    Ghadeer Abdullah Farsani
    Amal Shilash
    BMC Research Notes, 15
  • [19] Comparing the accuracy of point-of-care with laboratory (capillary, venous, and arterial) blood glucose levels in critically ill patients with and without shock
    Alshaer, Abdulaziz
    Badgheish, Basma A.
    Alsadah, Zahra Hashim
    Sewify, Khalid
    Alghazal, Sarah
    Alzahrani, Sarah
    Qadi, Abeer
    Alqahtani, Reham
    Farsani, Ghadeer Abdullah
    Shilash, Amal
    BMC RESEARCH NOTES, 2022, 15 (01)
  • [20] A New Digital Point-of-Care Tool With Advanced Blood Glucose Measuring Technology
    Gessler, Simon
    Porsch, Ulrich
    Klonoff, David C.
    Marquant, Michael
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2023, 17 (05): : 1321 - 1325