The accuracy of a glucose-oxidase-based point-of-care glucometer in premature infants

被引:3
|
作者
Tsao, Lon-Yen [1 ,4 ]
Chang, Ming-Yeh [2 ]
Hsiao, Chien-Chou [1 ,3 ,4 ]
机构
[1] Changhua Christian Hosp, Dept Neonatol, Changhua 500, Taiwan
[2] Changhua Christian Hosp, Dept Pediat Neurol, Changhua 500, Taiwan
[3] Chung Shan Med Univ, Sch Med, Taichung, Taiwan
[4] Changhua Christian Hosp, Clin Nutr Support Serv Team, Changhua 500, Taiwan
来源
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION | 2013年 / 98卷 / 06期
关键词
Neonatology; Paediatric Practice; Intensive Care; Measurement; BLOOD-GLUCOSE;
D O I
10.1136/archdischild-2012-303613
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective The purpose of this study was to evaluate the accuracy of a novel commercial glucose-oxidase (GO)-based glucometer in measuring glucose concentrations in arterial blood samples collected from premature infants. Patients and methods A commercial glucometer, which is a plasma separation GO-based glucometer, was employed to compare the accuracy of the commercial blood glucometer with that of a standard glucose analyser for monitoring blood glucose levels in premature infants. Arterial blood samples were collected and analysed, including samples adjusted for specific glucose (3.6mmol/L, 8.3mmol/L and 16.7mmol/L) and PaO2 levels (range, 40-400mmHg). Results In total, 159 samples were collected. Bland-Altman analysis showed good correlation between the commercial glucometer and standard glucose analyser. Error-grid analysis revealed that all of the results were within zone A (ie, the clinically accurate estimate zone). The biases between the two systems were low at different PaO2 levels and haematocrits. Finally, the influence of different PaO2 levels was within acceptable ranges. Conclusions The GO-based glucometer evaluated in this study provides accurate results even when measured at high PaO2 and different haematocrits and is suitable for measuring arterial glucose levels in premature infants.
引用
收藏
页码:F545 / F548
页数:4
相关论文
共 48 条
  • [1] Comparison of accuracy of glucose-oxidase-based and glucose-dehydrogenase-based point-of-care glucometers
    A Roman
    A Janier-Dubry
    C Hanicq
    P Flament
    F Vertongen
    E Stevens
    Critical Care, 11 (Suppl 2):
  • [2] Accuracy of point-of-care blood glucose measurements in the medical ICU
    Y Liu
    D Wu
    X Song
    Y Meng
    L Weng
    B Du
    Critical Care, 12 (Suppl 2):
  • [3] Precision and accuracy of a point of care glucometer for detection of hypoglycaemia in horses
    Hughes, Kristopher
    Moore, Claire
    Woods, Sophie
    Wilkes, Edwina
    VETERINARY JOURNAL, 2024, 305
  • [4] Critical Care Glucose Point-of-Care Testing
    Narla, S. N.
    Jones, M.
    Hermayer, K. L.
    Zhu, Y.
    ADVANCES IN CLINICAL CHEMISTRY, VOL 76, 2016, 76 : 97 - 121
  • [5] Point-of-Care Glucose and Ketone Monitoring
    Chong, Siew Kim
    Reineke, Erica L.
    TOPICS IN COMPANION ANIMAL MEDICINE, 2016, 31 (01) : 18 - 26
  • [6] Precision and Accuracy Comparison of Point-of-Care and Laboratory Glucose Concentrations in Cardiothoracic Surgery Patients
    Denfeld, Quin E.
    Goodell, Teresa T.
    Stafford, Kelly N.
    Kazmierczak, Steve
    JOURNAL OF CARDIOVASCULAR NURSING, 2011, 26 (06) : 512 - 518
  • [7] Accuracy of Two Professional-Use Systems for Point-of-Care Testing of Blood Glucose
    Pleus, S.
    Jendrike, N.
    Kamecke, U.
    Baumstark, A.
    Liebing, C.
    Freckmann, G.
    DIABETES STOFFWECHSEL UND HERZ, 2020, 29 (02): : 85 - 95
  • [8] Measurement accuracy of two professional-use systems for point-of-care testing of blood glucose
    Baumstark, Annette
    Jendrike, Nina
    Kamecke, Ulrike
    Liebing, Christina
    Pleus, Stefan
    Freckmann, Guido
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (03) : 445 - 455
  • [9] Evaluation of Four Point-Of-Care Glucose Meters in Alpacas
    Beemer, O.
    Byers, S.
    Bohn, A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2013, 27 (04) : 990 - 995
  • [10] Evaluation of a continuous glucose monitoring system compared with an in-house standard laboratory assay and a handheld point-of-care glucometer in critically ill neonatal foals
    Hug, Sophie A.
    Riond, Barbara
    Schwarzwald, Colin C.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2013, 23 (04) : 408 - 415