Biomechanical Evaluation of a Novel Porous-Structure Implant: Finite Element Study

被引:18
作者
Inglam, Samroeng [1 ]
Chantarapanich, Nattapon [2 ]
Suebnukarn, Siriwan [1 ]
Vatanapatimaku, Natapoom [3 ]
Sucharitpwatskul, Sedthawatt [3 ]
Sitthiseripratip, Kriskrai [3 ]
机构
[1] Thammasat Univ, Fac Dent, Pathum Thani, Thailand
[2] Prince Songkla Univ, Inst Biomed Engn, Fac Med, Hat Yai, Thailand
[3] Natl Met & Mat Technol Ctr, Pathum Thani, Thailand
关键词
dental implant; engineered porous-structure implant; finite element analysis; implant displacement; strain; stress; DENTAL IMPLANT; STRESS-DISTRIBUTION; BONE STRESS; TITANIUM; DESIGN; FABRICATION;
D O I
10.11607/jomi.1666
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The biomechanical performance of a novel engineered porous-structure implant (EPSI) with various porosities and a conventional solid-structure implant (CSSI) was investigated and compared. Materials and Methods: The three-dimensional finite element method was applied to titanium dental implant models placed in a block of bone that included both cortical and medullary bone. Five different pore sizes and porosities of the EPSI (58% porosity [PSI-58], 62% porosity [PSI-62], 71% porosity [PSI-71], 75% porosity [PSI-75], and 79% porosity [PSI-79]), were compared with the CSSI. Equivalent von Mises (EQV) stress, strain energy density, and displacement were examined for each implant design. Results: The maximum EQV stresses exhibited in cortical bone of the EPSI models were lower than those of the CSSI model. Higher EPSI porosity tended to increase the EQV stress. The EPSI appeared to share the load with the cortical bone, as evidenced by lower strain energy density in the cortical bone of EPSI models. High values for displacement were observed at the coronal part of the implant in all models. Slight differences in maximum displacement values were seen between EPSI and CSSI models. Conclusion: The EPSI effectively reduced the maximum EQV stress in the cortical bone and enhanced the load-sharing capacity. A significant amount of energy was absorbed by the implant instead of being transferred to the surrounding cortical bone. Varying the porosity of an implant had less effect on implant displacement.
引用
收藏
页码:E48 / E56
页数:9
相关论文
共 41 条
[1]   Effect of dental implant cross-sectional design on cortical bone structure using finite element analysis [J].
Abu-Hammad, Osama ;
Khraisat, Ameen ;
Dar-Odeh, Najla ;
El-Maaytah, Mohammed .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2007, 9 (04) :217-221
[2]  
Benn D K, 1992, Implant Dent, V1, P79
[3]   Novel production method and in-vitro cell compatibility of porous Ti-6Al-4V alloy disk for hard tissue engineering [J].
Bhattarai, Shanta Raj ;
Khalil, Khalil Abdel-Razek ;
Dewidar, Montasser ;
Hwang, Pyoung Han ;
Yi, Ho Keun ;
Kim, Hak Yong .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2008, 86A (02) :289-299
[4]   3-DIMENSIONAL STRESS-DISTRIBUTION AROUND A DENTAL IMPLANT AT DIFFERENT STAGES OF INTERFACE DEVELOPMENT [J].
BORCHERS, L ;
REICHART, P .
JOURNAL OF DENTAL RESEARCH, 1983, 62 (02) :155-159
[5]   Formation of highly adherent nano-porous alumina on Ti-based substrates:: a novel bone implant coating [J].
Briggs, EP ;
Walpole, AR ;
Wilshaw, PR ;
Karlsson, M ;
Pålsgård, E .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (09) :1021-1029
[6]  
CAMERON HU, 1976, J BIOMED MATER RES, V10, P295
[7]   TRABECULAR BONE-DENSITY AND LOADING HISTORY - REGULATION OF CONNECTIVE-TISSUE BIOLOGY BY MECHANICAL ENERGY [J].
CARTER, DR ;
FYHRIE, DP ;
WHALEN, RT .
JOURNAL OF BIOMECHANICS, 1987, 20 (08) :785-+
[8]  
Clelland N L, 1991, Int J Oral Maxillofac Implants, V6, P391
[9]   A 3-DIMENSIONAL FINITE-ELEMENT ANALYSIS OF A POROUS ROOTED CO-CR-MO ALLOY DENTAL IMPLANT [J].
COOK, SD ;
WEINSTEIN, AM ;
KLAWITTER, JJ .
JOURNAL OF DENTAL RESEARCH, 1982, 61 (01) :25-29
[10]   BONE STRESS-ADAPTATION MODELS [J].
COWIN, SC .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :528-533