A study to determine the differences between the displayed dose values for two full-field digital mammography units and values calculated using a range of Monte-Carlo-based techniques: a phantom study

被引:12
作者
Borg, M. [1 ]
Badr, I. [2 ]
Royle, G. J. [3 ]
机构
[1] Mater Dei Hosp, Med Imaging Dept, Msida, Malta
[2] St Georges Healthcare NHS Trust, Radiol Protect Ctr, London SW17 0QT, England
[3] UCL, Dept Med Phys & Bioengn, London WC1E 6BT, England
关键词
D O I
10.1093/rpd/ncs167
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Modern full-field digital mammography (FFDM) units display the mean glandular dose (MGD) and the entrance or incident air kerma (K) to the breast following each exposure. Information on how these values are calculated is limited and knowing how displayed MGD values compare and correlate to conventional Monte-Carlo-based methods is useful. From measurements done on polymethyl methacrylate (PMMA) phantoms, it has been shown that displayed and calculated MGD values are similar for thin to medium thicknesses and appear to differ with larger PMMA thicknesses. As a result, a multiple linear regression analysis on the data was performed to generate models by which displayed MGD values on the two FFDM units included in the study may be converted to the Monte-Carlo values calculated by conventional methods. These models should be a useful tool for medical physicists requiring MGD data from FFDM units included in this paper and should reduce the survey time spent on dose calculations.
引用
收藏
页码:217 / 228
页数:12
相关论文
共 13 条
  • [11] Van Engen R., 2006, EUROPEAN GUIDELINES, V4th
  • [12] SPECTRAL DEPENDENCE OF GLANDULAR TISSUE DOSE IN SCREEN-FILM MAMMOGRAPHY
    WU, XZ
    BARNES, GT
    TUCKER, DM
    [J]. RADIOLOGY, 1991, 179 (01) : 143 - 148
  • [13] NORMALIZED AVERAGE GLANDULAR DOSE IN MOLYBDENUM TARGET RHODIUM FILTER AND RHODIUM TARGET RHODIUM FILTER MAMMOGRAPHY
    WU, XZ
    GINGOLD, EL
    BARNES, GT
    TUCKER, DM
    [J]. RADIOLOGY, 1994, 193 (01) : 83 - 89