Validation and reliability testing of a new, fully integrated gait analysis insole

被引:75
作者
Braun, Benedikt Johannes [1 ]
Veith, Nils Thomas [1 ]
Hell, Rebecca [1 ]
Doebele, Stefan [2 ]
Roland, Michael [3 ]
Rollmann, Mika [1 ]
Holstein, Joerg [1 ]
Pohlemann, Tim [1 ]
机构
[1] Univ Saarland, Dept Trauma Hand & Reconstruct Surg, D-66421 Homburg, Germany
[2] Univ Tubingen, Dept Trauma Surg, BG Trauma Ctr, Tubingen, Germany
[3] Univ Saarland, Chair Appl Mech, D-66123 Saarbrucken, Germany
来源
JOURNAL OF FOOT AND ANKLE RESEARCH | 2015年 / 8卷
关键词
Gait analysis; Integrated insole system; Validation; Reliability; TIBIAL FRACTURES; SYSTEM; QUANTIFICATION; PARAMETERS; WALKING;
D O I
10.1186/s13047-015-0111-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: A new tool (OpenGo, Moticon GmbH) was introduced to continuously measure kinetic and temporospatial gait parameters independently through an insole over up to 4 weeks. The goal of this study was to investigate the validity and reliability of this new insole system in a group of healthy individuals. Methods: Gait data were collected from 12 healthy individuals on a treadmill at two different speeds. In total, six trials of three minutes each were performed by every participant. Validation was performed with the FDM-S System (Zebris). Complete sensor data were used for a within test reliability analysis of over 10000 steps. Intraclass correlation was calculated for different gait parameters and analysis of variance performed. Results: Intraclass correlation for the validation was >0.796 for temporospatial and kinetic gait parameters. No statistical difference was seen between the insole and force plate measurements (difference between means: 36.3 +/- 27.19 N; p = 0.19 and 0.027 +/- 0.028 s; p = 0.36). Intraclass correlation for the reliability was >0.994 for all parameters measured. Conclusion: The system is feasible for clinical trials that require step by step as well as grouped analysis of gait over a long period of time. Comparable validity and reliability to a stationary analysis tool has been shown.
引用
收藏
页数:7
相关论文
共 29 条
  • [1] Atallah L, 2011, P 2011 INT C BOD SEN, P29
  • [2] Gait analysis using a shoe-integrated wireless sensor system
    Bamberg, Stacy J. Morris
    Benbasat, Ari Y.
    Scarborough, Donna Moxley
    Krebs, David E.
    Paradiso, Joseph A.
    [J]. IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2008, 12 (04): : 413 - 423
  • [3] Benocci M., 2009, PERV COMP TECHN HEAL
  • [4] Comparison of four methods of calculating the symmetry of spatial-temporal parameters of gait
    Blazkiewicz, Michalina
    Wiszomirska, Ida
    Wit, Andrzej
    [J]. ACTA OF BIOENGINEERING AND BIOMECHANICS, 2014, 16 (01) : 29 - 35
  • [5] Braun BJ, 2015, MED HYPOTHESES
  • [6] Gait analysis in the therapeutic environment
    Coutts, F
    [J]. MANUAL THERAPY, 1999, 4 (01) : 2 - 10
  • [7] Least-action principle in gait
    Fan, Y. F.
    Loan, M.
    Fan, Y. B.
    Li, Z. Y.
    Luo, D. L.
    [J]. EPL, 2009, 87 (05)
  • [8] Feng Y, 2011, INF AUT ICIA 2011 IE, P329
  • [9] DEVIATIONS IN GAIT METRICS IN PATIENTS WITH CHRONIC ANKLE INSTABILITY
    Gigi, R.
    Mor, A.
    Haim, A.
    Luger, E.
    Melamed, E.
    Beer, Y.
    Nof, M.
    Nyska, M.
    Elbaz, A.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2014, 22 : S123 - S123
  • [10] Godfrey A, 2014, IEEE ENG MED BIO, P5892, DOI 10.1109/EMBC.2014.6944969