Evaluation of a biplanar diode array dosimeter for quality assurance of step-and-shoot IMRT

被引:60
作者
Feygelman, Vladimir [1 ]
Forster, Kenneth [1 ]
Opp, Daniel [1 ,2 ]
Nilsson, Gorgen [3 ]
机构
[1] H Lee Moffitt Canc Ctr & Res Inst, Div Radiat Oncol, Tampa, FL 33612 USA
[2] Univ S Florida, Dept Phys, Tampa, FL 33620 USA
[3] ScandiDos AB, Uppsala, Sweden
关键词
radiation therapy; intensity-modulated radiation therapy; dose distribution comparison; three-dimensional dosimetry; diode array dosimeter; MODULATED RADIATION-THERAPY; ION-CHAMBER ARRAY; KODAK EDR2 FILM; DOSE VERIFICATION; IONIZATION-CHAMBER; RADIOCHROMIC FILM; RADIOGRAPHIC FILM; POLYMER GEL; RADIOTHERAPY; QA;
D O I
10.1120/jacmp.v10i4.3080
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper, we describe and characterize a novel biplanar diode array, and demonstrate its applicability to dosimetric QA of step-and-shoot IMRT. It is the first commercially available device of its kind specifically designed for performing measurements at varying gantry angles. The detector consists of a cylindrical PMMA phantom with two orthogonal detector boards. There are a total of 1069 p-type 1 mm wide diode detectors covering the measurement area of 20 x 20 cm(2) in each of the measurement planes. The orthogonal detector arrays ensure that the dose modulation information is not lost regardless of the beam incidence angle. For absolute calibration, the dose to the reference detector is calculated at the appropriate SSD and radiological depth by the treatment planning system and is scaled by the measured accelerator output. The directly measured rotational response on the central axis shows the maximum variation of approximately +/- 3% in the narrow +/- 1 degrees angular intervals centered on the detector boards. This variation is reduced to less than +/- 2% outside of the four similarly centered +/- 5 degrees angular intervals. For all detectors, the difference between the measured and the calculated dose for a plan with 12 equally spaced beams is -0.2 +/- 0.9%. Of eleven IMRT plans, ten passed the gamma (3%, 3 mm) criterion at or above 95%, while one passed at 92%. The biplanar diode array is a useful tool for IMRT QA, allowing for essentially instantaneous online analysis of absolute dose errors in 3D.
引用
收藏
页码:64 / 78
页数:15
相关论文
共 31 条
[1]   Three-dimensional dose verification for intensity-modulated radiation therapy in the radiological physics centre head-and-neck phantom using optical computed tomography scans of ferrous xylenol-orange gel dosimeters [J].
Babic, Steven ;
Battista, Jerry ;
Jordan, Kevin .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (04) :1281-1291
[2]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[3]   Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system [J].
Cadman, P ;
Bassalow, R ;
Sidhu, NPS ;
Ibbott, G ;
Nelson, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :3001-3010
[4]   Dosimetric accuracy of Kodak EDR2 film for IMRT verifications [J].
Childress, NL ;
Salehpour, M ;
Dong, L ;
Bloch, C ;
White, RA ;
Rosen, II .
MEDICAL PHYSICS, 2005, 32 (02) :539-548
[5]   Introducing gel dosimetry in a clinical environment: Customization of polymer gel composition and magnetic resonance imaging parameters used for 3D dose verifications in radiosurgery and intensity modulated radiotherapy [J].
Crescenti, Remo A. ;
Scheib, Stefan G. ;
Schneider, Uwe ;
Gianolini, Stefano .
MEDICAL PHYSICS, 2007, 34 (04) :1286-1297
[6]   MAGIC-type polymer gel for three-dimensional dosimetry:: Intensity-modulated radiation therapy verification [J].
Gustavsson, H ;
Karlsson, A ;
Bäck, SÅJ ;
Olsson, LE ;
Haraldsson, P ;
Engström, P ;
Nyström, H .
MEDICAL PHYSICS, 2003, 30 (06) :1264-1271
[7]   Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine [J].
Herzen, J. ;
Todorovic, M. ;
Cremers, F. ;
Platz, V. ;
Albers, D. ;
Bartels, A. ;
Schmidt, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (04) :1197-1208
[8]   Initial evaluation of commercial optical CT-based 3D gel dosimeter [J].
Islam, KTS ;
Dempsey, JF ;
Ranade, MK ;
Maryanski, MJ ;
Low, DA .
MEDICAL PHYSICS, 2003, 30 (08) :2159-2168
[9]   A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J].
Jursinic, PA ;
Nelms, BE .
MEDICAL PHYSICS, 2003, 30 (05) :870-879
[10]   Evaluation of a 2D diode array for IMRT quality assurance [J].
Létourneau, D ;
Gulam, M ;
Yan, D ;
Oldham, M ;
Wong, JW .
RADIOTHERAPY AND ONCOLOGY, 2004, 70 (02) :199-206