A comparison of lithium dilution cardiac output measurements made using central and antecubital venous injection of lithium chloride

被引:20
作者
Jonas, MM
Kelly, FE
Linton, RAF [1 ]
Band, DM
O'Brien, TK
Linton, NWF
机构
[1] St Thomas Hosp, Rayne Inst, London SE1 7EH, England
[2] Southampton Gen Hosp, Shackleton Dept Anaesthet, Southampton SO16 7EH, Hants, England
关键词
cardiac output; monitoring; thermodilution; critical care; hemodynamics; heart; lithium;
D O I
10.1023/A:1009914714769
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective. We have previously described an indicator dilution technique of measuring cardiac output in which lithium chloride is injected as a bolus via a central venous catheter and cardiac output derived from the arterial lithium dilution curve recorded from a lithium-selective electrode, which we have developed for this purpose. It would be an advantage if the lithium could be injected via the basilic vein (in the antecubital fossa) in those patients who do not need central venous catheterisation for other reasons. We have therefore compared cardiac output measurements made using these two routes of lithium chloride administration. Methods. Lithium dilution cardiac output was measured 10 times in each of 10 patients, injecting the lithium chloride alternately via the basilic or central venous catheter. Results. The mean difference was 0.8 +/- 5.2% (SD) (range -8.5 to +7.0%) over a range of cardiac output of 4.5-13 l/min. Conclusions. Injection of lithium chloride via the basilic vein in the antecubital fossa allows accurate lithium dilution cardiac output measurements to be made in patients who do not have central venous catheters in place.
引用
收藏
页码:525 / 528
页数:4
相关论文
共 9 条
  • [1] The shape of indicator dilution curves used for cardiac output measurement in man
    Band, DM
    Linton, RAF
    OBrien, TK
    Jonas, MM
    Linton, NWF
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1997, 498 (01): : 225 - 229
  • [2] The effectiveness of right heart catheterization in the initial care of critically ill patients
    Connors, AF
    Speroff, T
    Dawson, NV
    Thomas, C
    Harrell, FE
    Wagner, D
    Desbiens, N
    Goldman, L
    Wu, AW
    Califf, RM
    Fulkerson, WJ
    Vidaillet, H
    Broste, S
    Bellamy, P
    Lynn, J
    Knaus, WA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11): : 889 - 897
  • [3] ERRORS IN MEASUREMENT OF CARDIAC OUTPUT WITH DYE-DILUTION CURVES IN SHOCK
    HOMER, LD
    MOSS, GS
    HERMAN, CM
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1969, 27 (01) : 101 - &
  • [4] Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output
    Kurita, T
    Morita, K
    Kato, S
    Kikura, M
    Horie, M
    Ikeda, K
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1997, 79 (06) : 770 - 775
  • [5] Lithium dilution cardiac output measurement: A comparison with thermodilution
    Linton, R
    Band, D
    OBrien, T
    Jonas, M
    Leach, R
    [J]. CRITICAL CARE MEDICINE, 1997, 25 (11) : 1796 - 1800
  • [6] A NEW METHOD OF MEASURING CARDIAC-OUTPUT IN MAN USING LITHIUM DILUTION
    LINTON, RAF
    BAND, DM
    HAIRE, KM
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (02) : 262 - 266
  • [7] A new method of analysing indicator dilution curves
    Linton, RAF
    Linton, NWF
    Band, DM
    [J]. CARDIOVASCULAR RESEARCH, 1995, 30 (06) : 930 - 938
  • [8] ERRORS IN THE MEASUREMENT OF CARDIAC-OUTPUT BY THERMODILUTION
    NISHIKAWA, T
    DOHI, S
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (02): : 142 - 153
  • [9] LIMITATIONS OF INDICATOR-DILUTION METHODS IN EXPERIMENTAL SHOCK
    ORIOL, A
    SEKELJ, P
    MCGREGOR, M
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1967, 23 (04) : 605 - &